Literature DB >> 29080274

Low-dose CT perfusion with projection view sharing.

Thomas Martin1, John Hoffman2, Jeff R Alger1, Michael McNitt-Gray2, Danny Jj Wang1,3.   

Abstract

PURPOSE: CT Perfusion (CTP) is a widely used clinical imaging modality. However, CTP typically involves the use of substantial radiation dose (CTDIvol ≥~200 mGy). The purpose of this study is to present a low-dose CTP technique using a projection view-sharing reconstruction algorithm originally developed for dynamic MRI - "K-space Weighted Image Contrast" (KWIC).
METHODS: The KWIC reconstruction is based on an angle-bisection scheme. In KWIC, a Fourier transform was performed along each projection to form a "k-space"-like CT data space, based on the central-slice theorem. As a projection view-sharing technique, KWIC preserves the spatiotemporal resolution of undersampled CTP data by progressively increasing the number of projection views shared for more distant regions of "k-space". KWIC reconstruction was evaluated on a digital FORBILD head phantom with numerically simulated time-varying objects. The numerically simulated scans were undersampled using the angle-bisection scheme to achieve 50%, 25%, and 12.5% of the original dose (288, 144, and 72 projections, respectively). The area-under-the-curve (AUC), time-to-peak (TTP), and full width half maximum (FWHM) were measured in KWIC recons and compared to fully sampled filtered back projection (FBP) reconstructions. KWIC reconstruction and dose reduction was also implemented for three clinical CTP cases (45 s, 1156 projections per turn, 1 s/turn, CTDIvol 217 mGy). Quantitative perfusion metrics were computed and compared between KWIC reconstructed CTP data and those of standard FBP reconstruction.
RESULTS: The AUC, TTP, and FWHM in the numerical simzulations were unaffected by the undersampling/dose reduction (down to 12.5% dose) with KWIC reconstruction compared to the fully sampled FBP reconstruction. The normalized root-mean-square-error (NRMSE) of the AUC in the FORBILD head phantom is 0.04, 0.05, and 0.07 for 50%, 25%, and 12.5% KWIC, respectively, as compared to FBP reconstruction. The cerebral blood flow (CBF) and cerebral blood volume had no significant difference between FBP and 50%, 25%, and 12.5% KWIC reconstructions (P > 0.05).
CONCLUSIONS: This study demonstrates that KWIC preserves perfusion metrics for CTP with substantially reduced dose. Clinical implementation will require further investigation into methods of rapid switching of a CT x-ray source.
© 2017 American Association of Physicists in Medicine.

Entities:  

Keywords:  CT perfusion; filtered back projection (FBP); k-space weighted image contrast (KWIC); projection view sharing

Mesh:

Year:  2017        PMID: 29080274      PMCID: PMC5774228          DOI: 10.1002/mp.12640

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  36 in total

1.  Resampling of data between arbitrary grids using convolution interpolation.

Authors:  V Rasche; R Proksa; R Sinkus; P Börnert; H Eggers
Journal:  IEEE Trans Med Imaging       Date:  1999-05       Impact factor: 10.048

2.  k-space weighted image contrast (KWIC) for contrast manipulation in projection reconstruction MRI.

Authors:  H K Song; L Dougherty
Journal:  Magn Reson Med       Date:  2000-12       Impact factor: 4.668

Review 3.  Radiation dose reduction in perfusion CT imaging of the brain: A review of the literature.

Authors:  Ahmed E Othman; Saif Afat; Marc A Brockmann; Omid Nikoubashman; Carolin Brockmann; Konstantin Nikolaou; Martin Wiesmann
Journal:  J Neuroradiol       Date:  2015-12-10       Impact factor: 3.447

4.  Radiation dose reduction in time-resolved CT angiography using highly constrained back projection reconstruction.

Authors:  Mark Supanich; Yinghua Tao; Brian Nett; Kari Pulfer; Jiang Hsieh; Patrick Turski; Charles Mistretta; Howard Rowley; Guang-Hong Chen
Journal:  Phys Med Biol       Date:  2009-06-30       Impact factor: 3.609

5.  Combination of a low-tube-voltage technique with hybrid iterative reconstruction (iDose) algorithm at coronary computed tomographic angiography.

Authors:  Yoshinori Funama; Katsuyuki Taguchi; Daisuke Utsunomiya; Seitaro Oda; Yumi Yanaga; Yasuyuki Yamashita; Kazuo Awai
Journal:  J Comput Assist Tomogr       Date:  2011 Jul-Aug       Impact factor: 1.826

6.  Respiratory motion-compensated radial dynamic contrast-enhanced (DCE)-MRI of chest and abdominal lesions.

Authors:  Wei Lin; Junyu Guo; Mark A Rosen; Hee Kwon Song
Journal:  Magn Reson Med       Date:  2008-11       Impact factor: 4.668

7.  Use of a computer-controlled motion phantom to investigate the temporal and spatial fidelity of HYPR processing.

Authors:  Lauren Keith; Mahdi Rahimi; James Holmes; Jean Brittain; Frank Korosec
Journal:  Magn Reson Med       Date:  2014-02       Impact factor: 4.668

8.  A trial of imaging selection and endovascular treatment for ischemic stroke.

Authors:  Chelsea S Kidwell; Reza Jahan; Jeffrey Gornbein; Jeffry R Alger; Val Nenov; Zahra Ajani; Lei Feng; Brett C Meyer; Scott Olson; Lee H Schwamm; Albert J Yoo; Randolph S Marshall; Philip M Meyers; Dileep R Yavagal; Max Wintermark; Judy Guzy; Sidney Starkman; Jeffrey L Saver
Journal:  N Engl J Med       Date:  2013-02-08       Impact factor: 91.245

9.  Golden-angle radial sparse parallel MRI: combination of compressed sensing, parallel imaging, and golden-angle radial sampling for fast and flexible dynamic volumetric MRI.

Authors:  Li Feng; Robert Grimm; Kai Tobias Block; Hersh Chandarana; Sungheon Kim; Jian Xu; Leon Axel; Daniel K Sodickson; Ricardo Otazo
Journal:  Magn Reson Med       Date:  2013-10-18       Impact factor: 4.668

10.  FDA investigates the safety of brain perfusion CT.

Authors:  M Wintermark; M H Lev
Journal:  AJNR Am J Neuroradiol       Date:  2009-11-05       Impact factor: 4.966

View more
  3 in total

1.  Early Diagnosis of Acute Ischemic Stroke by Brain Computed Tomography Perfusion Imaging Combined with Head and Neck Computed Tomography Angiography on Deep Learning Algorithm.

Authors:  Yi Yang; Jinjun Yang; Jiao Feng; Yi Wang
Journal:  Contrast Media Mol Imaging       Date:  2022-05-09       Impact factor: 3.009

2.  The effect of patellofemoral pain syndrome on patellofemoral joint kinematics under upright weight-bearing conditions.

Authors:  Jae-Suk Yang; Michael Fredericson; Jang-Hwan Choi
Journal:  PLoS One       Date:  2020-09-30       Impact factor: 3.240

3.  Low Dose CT Perfusion With K-Space Weighted Image Average (KWIA).

Authors:  Chenyang Zhao; Thomas Martin; Xingfeng Shao; Jeffry R Alger; Vinay Duddalwar; Danny J J Wang
Journal:  IEEE Trans Med Imaging       Date:  2020-11-30       Impact factor: 10.048

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.