| Literature DB >> 29079631 |
Andrew R Chapman1, Nicholas L Mills1.
Abstract
Entities:
Keywords: acute coronary syndromes; acute myocardial infarction
Mesh:
Substances:
Year: 2017 PMID: 29079631 PMCID: PMC5890635 DOI: 10.1136/heartjnl-2017-312269
Source DB: PubMed Journal: Heart ISSN: 1355-6037 Impact factor: 5.994
99th centile=the upper reference limit as determined in a healthy reference range population
| 99th centile | 10% coefficient of variation* | Limit of detection | |
| Roche Elecsys high-sensitivity cardiac troponin T | 14 ng/L | 13 ng/L | 5 ng/L |
| Abbott ARCHITECTSTAT high-sensitivity cardiac troponin I | 16 ng/L (females) | 4.7 ng/L | 2 ng/L |
Limit of detection, the lowest concentration which can be reliably distinguished from a sample with no troponin present.
*Lowest concentration where coefficient of variation is <10% (measure of dispersion of replicate sample results around the mean (SD/mean)).