| Literature DB >> 29077730 |
Gloria Roberti1, Gianluca Manni1,2, Ivano Riva1, Gabor Holló3, Luciano Quaranta4, Luca Agnifili5, Michele Figus6, Sara Giammaria7, Davide Rastelli2, Francesco Oddone1.
Abstract
PURPOSE: To compare the detection rate of central visual field defect (CVFD) between the 30-degree Octopus G1 program (Dynamic strategy) and the HFA 10-2 SITA-Standard test in early glaucoma eyes not showing any CVFD on the HFA 24-2 SITA-Standard test.Entities:
Mesh:
Year: 2017 PMID: 29077730 PMCID: PMC5659771 DOI: 10.1371/journal.pone.0186793
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Demographic data of study population.
| N = 43 (mean ± SD) | |
|---|---|
| 65.2±10.1 (range: 38–85) | |
| 22/19 | |
| 9.8±0.4 | |
| 68.8±11.8 | |
| 14.7±2.7 |
SD = standard deviation; M = male; F = female; BCVA = best corrected visual acuity; RNFL = retinal nerve fiber layer measured with optical coherence tomography; IOP = intraocular pressure.
Visual field results with the HFA 24–2 and HFA 10–2 test and the Octopus G1 program.
| HFA 24–2 (n = 41) | HFA 10–2 (n = 41) | Octopus G1 (n = 41) | |
|---|---|---|---|
| 0 | 33 (80.4%) | 23 (56.0%) | |
| 5.5±0.7 | 5.6±0.8 | 5.2±1.1 | |
| -3.26±2.6 | -1.67±2.1 | 4.81±2.8 | |
| 3.84±2.4 | 2.20±1.9 | 4.58±1.7 | |
| 11.95±5.5 | N/A | 19.68±10.6 |
PSD = pattern standard deviation; HFA = Humphrey field analyzer; sLV = square loss variance; CVFD = central visual field defect: MS = mean sensitivity; N/A = not applicable.
* In Humphrey visual field analysis, the abnormal mean deviation values are negative numbers; in Octopus perimetry the abnormal mean defect values are positive numbers
Fig 1Pattern deviation plots of the Humphrey Field Analyzer (HFA) 24–2 test and the HFA 10–2 test, and the corrected probability plot of the Octopus G1 program of the same early glaucoma eye from the study.
The pattern deviation plot of the HFA 24–2 test shows no depressed test point within the central 10° area (the area surrounded by the cross shape line) but the sensitivity is reduced in the more peripheral area of the visual field. The pattern deviation plot of the HFA 10–2 test shows a cluster of depressed test point locations. The peripheral retinal sensitivity is not tested. The corrected probability plot of the Octopus G1 program shows a cluster of reduced sensitivity both in the central 10° area (the region within the inner black circle) and in the periphery.
Fig 2Central 10 degrees of the visual field with the locations of HFA 24–2, HFA 10–2 and Octopus G1 plotted in the same grid.
The inner circle delimits the central 10 degrees area with test locations of the three tests (HFA 24–2, HFA 10–2 and Octopus G1) marked with different symbols. Cross = HFA 24–2; square = HFA 10–2; triangle = Octopus G1.
Agreement of the classification of the central 10-degree visual field between the HFA 10–2 test and the Octopus G1 test.
| Octopus G1 | ||||
|---|---|---|---|---|
| Normal | Abnormal | Total | ||
| 8 | ||||
| 33 | ||||
| 18 | 23 | 41 | ||