| Literature DB >> 29066495 |
Mahesheema Ali1,2, Deepthi Rajapakshe2, Liyun Cao3, Sridevi Devaraj4,2.
Abstract
Recent studies have reported that biotin interferes with certain immunoassays. In this study, we evaluated the analytical interference of biotin on immunoassays that use streptavidin-biotin in our pediatric hospital. We tested the effect of different concentrations of biotin (1.5-200 ng/ml) on TSH, Prolactin, Ferritin, CK-MB, β-hCG, Troponin I, LH, FSH, Cortisol, Anti-HAV antibody (IgG and IgM), assays on Ortho Clinical Diagnostic Vitros 5600 Analyzer. Biotin (up to 200 ng/mL) did not significantly affect Troponin I and HAV assays. Biotin (up to 12.5 ng/ml) resulted in <10% bias in CK-MB, β-hCG, AFP, Cortisol, Ferritin assays and biotin >6.25 ng/mL significantly affected TSH (>20% bias) assay. Prolactin was significantly affected even at low levels (Biotin 1.5 ng/mL). Thus, we recommend educating physicians about biotin interference in common immunoassays and adding an electronic disclaimer.Entities:
Keywords: Biotin; Immunoassay; Interference; Vitros
Mesh:
Substances:
Year: 2017 PMID: 29066495
Source DB: PubMed Journal: Ann Clin Lab Sci ISSN: 0091-7370 Impact factor: 1.256