Inadvertent positioning of a pacing lead in the left atrium (LA) or left ventricle (LV) is a rare but well-recognized complication of pacemaker or defibrillator implantation. Traditional management involves systemic anticoagulation to reduce the risk of thromboembolic complications. Although international consensus guidelines (Heart Rhythm Society, 2009) do not recommend routine percutaneous lead extraction in asymptomatic patients (class IIIC indication), extraction can be considered “if the clinical scenario is compelling.” Lead extraction from the LA carries a risk of cerebral embolism, and therefore techniques to reduce this risk may be of clinical benefit.
Case report
A woman underwent dual-chamber implantable cardioverter-defibrillator implantation for long QT syndrome 1 and recurrent syncope at the age of 36. At her second generator change 13 years later, a new atrial lead was implanted owing to failure of the original lead. It was difficult to find an atrial site with satisfactory parameters, but eventually a suitable location was achieved.After the procedure, the patient felt generally unwell with fatigue and muscle and joint pains, although these were attributed to a new diagnosis of systemic lupus erythematosus. She had an episode of transient unilateral paresthesia and although computed tomography of the brain was normal, plain chest radiographs showed a posterior position of the atrial lead (Figure 1) and transthoracic echocardiography confirmed that the atrial lead was in the LA (Figure 2), traversing a patent foramen ovale (PFO).
Figure 1
Posteroanterior (left) and lateral (right) plain chest radiographs demonstrating the posterior position of the left atrial lead (blue arrow), the redundant right atrial lead (red arrow), and the right ventricular lead (yellow arrow).
Figure 2
A: Sentinel Cerebral Protection Device in left common carotid (white arrow) and brachiocephalic (yellow arrow) arteries and pigtail catheter in aortic arch (red arrow). B: Transthoracic echocardiogram demonstrating pacing lead in the left atrium (arrow). C: Extracted lead tip with adherent atrial tissue and thrombus. D: Thrombus retrieved from filter basket.
Posteroanterior (left) and lateral (right) plain chest radiographs demonstrating the posterior position of the left atrial lead (blue arrow), the redundant right atrial lead (red arrow), and the right ventricular lead (yellow arrow).A: Sentinel Cerebral Protection Device in left common carotid (white arrow) and brachiocephalic (yellow arrow) arteries and pigtail catheter in aortic arch (red arrow). B: Transthoracic echocardiogram demonstrating pacing lead in the left atrium (arrow). C: Extracted lead tip with adherent atrial tissue and thrombus. D: Thrombus retrieved from filter basket.Anticoagulation with rivaroxaban was commenced and she was referred for consideration of lead extraction. Transesophageal echocardiography (TEE) confirmed that the lead was in the LA but did not demonstrate any adherent vegetation or thrombus. Both surgical and percutaneous approaches were discussed with the patient, including the risk of systemic embolization from the lead. After multidisciplinary discussion, a decision to proceed with percutaneous extraction was agreed.The extraction procedure was performed under general anesthesia with intraoperative TEE. An arch aortogram was performed and a Sentinel Cerebral Protection Device (Claret Medical, Santa Rosa, CA) was positioned in the aortic arch via the right radial artery. The proximal filter was deployed in the brachiocephalic artery and the distal filter was deployed in the left common carotid artery (Figure 2). Systemic unfractionated heparin was administered to maintain an activated clotting time of greater than 250 seconds throughout the procedure.A guidewire was placed in the superior vena cava via the right femoral vein to allow rapid deployment of a Bridge Occlusion Balloon (Spectranetics, Colorado Springs, CO) in the event of a superior vena cava tear. The generator pocket was opened and the generator and leads were freed. The LA lead, which had been in place for 2 years, was cut to allow the passage of a Lead Locking Device (Spectranetics). The lead was extracted without complication using a 14 French laser sheath (Spectranetics; 80 Hz laser at the clavicular level only) and a new atrial lead was implanted in the right atrial appendage, with the position confirmed on TEE. The Sentinel Cerebral Protection Device was removed and examination of the filter basket revealed a small thrombus (Figure 2). There was no postprocedural hematoma, and so the wound was closed with absorbable sutures and managed with simple dressings. Anticoagulation with rivaroxaban was recommenced the following day.The patient developed a postprocedural lower respiratory tract infection requiring antibiotics, but otherwise she made an uneventful recovery.
Discussion
In this report, we present a novel approach for cerebral protection during the percutaneous extraction of a left-sided pacemaker lead.Lead malposition in the LV and, less commonly, the LA is a serious complication of cardiac device implantation and usually occurs via an atrial septal defect or PFO, or via inadvertent arterial access. The incidence is unknown and difficult to assess, and complications include thromboembolism and left-sided endocarditis. Though the 2009 Heart Rhythm Society expert statement does not recommend lead extraction in asymptomatic patients with anomalously placed LV leads (there are no guidelines regarding LA leads), in patients with thromboembolism or endocarditis surgical and percutaneous extraction are clinically indicated and well described.Echocardiography may not always detect thrombus or vegetation adherent to the lead, and in view of the potentially catastrophic cerebrovascular consequences during left-sided percutaneous extraction procedures, protection of the cerebral circulation is desirable. Although there are previous reports of the use of a Cordis Angioguard (Cordis, Miami Lakes, FL) device for this indication,2, 3 the Angioguard device requires quadruple arterial access and 4 separate cerebral protection devices to be placed in the left and right common carotid arteries and the left and right vertebral arteries. In contrast, the Sentinel Cerebral Protection Device can be deployed to the brachiocephalic artery and left common carotid artery via a single 6 French right radial or brachial artery sheath, reducing procedural time and risk of femoral arterial complications.The Sentinel Cerebral Protection Device has been used previously during transcatheter aortic valve implantation, where it has been shown to reduce postprocedural magnetic resonance imaging–detected cerebral lesions and neurocognitive deterioration, and also during thoracic endovascular aneurysm repair and left atrial appendage occlusion. However, to the best of our knowledge, this is the first reported use of the device for a left-sided pacemaker lead extraction.There are several limitations of this approach. The Sentinel Cerebral Protection Device is available in 1 size only and is suitable for most, but not all, vascular anatomies. It does not protect the left vertebral artery territory, nor does it protect the peripheral circulation from embolization. It is important to minimize peripheral embolization by avoiding oversizing of sheaths (to reduce the risk of air embolization) and by not crossing the PFO with the extraction sheath. Systemic anticoagulation during the procedure is required to reduce the risk of thrombus formation on the device, and this increases the risk of pocket hematoma formation. Furthermore, although this approach is likely to reduce intraprocedural embolization, postprocedural embolization is still possible and patients with endocardial leads and a PFO have been shown to be at increased risk of cerebrovascular accident/transient ischemic attack, even without left-sided leads.
Conclusion
Cerebral thromboembolism is a potentially devastating consequence of percutaneous extraction of inadvertent left-sided pacing and defibrillator leads. Protection of the cerebral circulation during these procedures, although unproven, is likely to reduce the risk of this complication and has been demonstrated to be beneficial during transcatheter aortic valve implantation procedures. This is the first reported use in a lead extraction of a simple protection device that can be delivered rapidly via a single right radial or brachial arterial access.Lead malposition in the left ventricle and, less commonly, the left atrium is a serious complication of cardiac device implantation and usually occurs via an atrial septal defect or patent foramen ovale, or via inadvertent arterial access. The incidence is unknown and difficult to assess, and complications include thromboembolism and left-sided endocarditis.International consensus guidelines recommend anticoagulation in asymptomatic patients with anomalously placed left ventricular leads (there are no guidelines regarding left atrial leads). In patients with thromboembolism or endocarditis, surgical and percutaneous extraction are clinically indicated and well described.Cerebral thromboembolism is a potentially devastating consequence of percutaneous extraction of inadvertent left-sided pacing and defibrillator leads. Protection of the cerebral circulation during these procedures, although unproven, is likely to reduce the risk of this complication and has been demonstrated to be beneficial during transcatheter aortic valve implantation procedures.
Authors: Nicolas M Van Mieghem; Lennart van Gils; Habib Ahmad; Floortje van Kesteren; Hendrik W van der Werf; Guus Brueren; Michiel Storm; Mattie Lenzen; Joost Daemen; Ad F M van den Heuvel; Pim Tonino; Jan Baan; Peter J Koudstaal; Marguerite E I Schipper; Aad van der Lugt; Peter P T de Jaegere Journal: EuroIntervention Date: 2016-07-20 Impact factor: 6.534
Authors: Bruce L Wilkoff; Charles J Love; Charles L Byrd; Maria Grazia Bongiorni; Roger G Carrillo; George H Crossley; Laurence M Epstein; Richard A Friedman; Charles E H Kennergren; Przemyslaw Mitkowski; Raymond H M Schaerf; Oussama M Wazni Journal: Heart Rhythm Date: 2009-05-22 Impact factor: 6.343
Authors: J Jenkins Thompson; Kelly M McDonnell; John F Reavey-Cantwell; Kenneth A Ellenbogen; Jayanthi N Koneru Journal: Circ Arrhythm Electrophysiol Date: 2014-12
Authors: John N Bahadorani; Amir A Schricker; Victor G Pretorius; Ulrika Birgersdotter-Green; Arturo Dominguez; Ehtisham Mahmud Journal: Catheter Cardiovasc Interv Date: 2015-02-12 Impact factor: 2.692
Authors: Christopher V DeSimone; Paul A Friedman; Amit Noheria; Nikhil A Patel; Daniel C DeSimone; Sami Bdeir; Christopher A Aakre; Vaibhav R Vaidya; Joshua P Slusser; David O Hodge; Michael J Ackerman; Alejandro A Rabinstein; Samuel J Asirvatham Journal: Circulation Date: 2013-08-14 Impact factor: 29.690