Woo Jung Lee1, Lance T Uradomo2, Yang Zhang2, William Twaddell2, Peter Darwin2.
Abstract
BACKGROUND AND AIMS: EUS-guided liver biopsy is an emerging method of liver tissue acquisition which is safe and had been shown to produce excellent histological yield. There is limited data comparing the diagnostic yield of different FNA needles. We aimed to compare the diagnostic performance of four commercially available 19-gauge FNA needles.Entities:
Year: 2017 PMID: 29056843 PMCID: PMC5615982 DOI: 10.1155/2017/1497831
Source DB: PubMed Journal: Diagn Ther Endosc ISSN: 1026-714X
Figure 1Commercially available FNA needles with different needles tips. From left to right: TruCore percutaneous needles, Olympus EZ Shot 2, Covidien SharkCore, Boston Scientific Expect Slimline, and Cook Echotip Procore.
Figure 2This diagram represents schema of each FNA needles biopsy; different numbers of throws were grouped separately, and biopsies were obtained from right to left lobes of liver from livers 1 and 2, totaling 12 separate specimen jars for each needle. Only 1 specimen was obtained from each lobe of liver for percutaneous reference needle, totaling 4 specimen bottles.
Figure 3(a) Picture showing intact specimen and its length; (b) picture of specimen showing two complete portal tracts (CPT).
Figure 4Comparison of CPT obtained from each FNA needle using one-way ANOVA analysis with Bonferroni correction.
Mean complete portal tracts by needle type and number of throws.
| Number of throws | |||
|---|---|---|---|
| Three | Six | Nine | |
| Cook | 2.75 | 3.75 | 3.5 |
| Boston Scientific | 3 | 4 | 5 |
| Olympus | 1.5 | 5.5 | 6.25 |
| Covidien | 9.5 | 5.75 | 11.25 |
|
|
|
| |
Total specimen length by needle type.
| Total specimen length (cm) | |
|---|---|
| Cook | 4.73 |
| Boston Scientific | 2.98 |
| Olympus | 4.26 |
| Covidien | 5.07 |
| Trucut | 1.55 |
Total specimen length by needle type and number of throws.
| Number of throws | |||
|---|---|---|---|
| Three | Six | Nine | |
| Cook | 2.25 | 3.43 | 3.25 |
| Boston Scientific | 3.82 | 5 | 5.35 |
| Olympus | 2.55 | 4.97 | 5.25 |
| Covidien | 4.28 | 5.63 | 5.33 |
|
|
|
| |