Effect of central corneal thickness on intraocular pressure and comparison of Topcon CT-80 non-contact tonometry with Goldmann applanation tonometry.

Tarannum Mansoori1, Nagalla Balakrishna2.   

Abstract

BACKGROUND: To compare intraocular pressure (IOP) measurements obtained with the Topcon CT-80 non-contact tonometer (NCT) and Goldmann applanation tonometer (GAT), in different ranges of IOP in normal and glaucoma subjects, and to assess the influence of central corneal thickness (CCT) on the IOP measurements in Asian Indian eyes.
METHODS: Four hundred and two eyes of 402 subjects (193 newly diagnosed primary open angle glaucoma [POAG] and 209 normal) were enrolled for this prospective study. For each eye, IOP was measured with GAT by a glaucoma specialist and NCT by a trained optometrist. The IOP values were compared among the tonometers in the three different IOP ranges (≤ 12 mmHg, 13-20 mmHg and ≥ 21 mmHg) using Bland-Altman graphs. Correlation between GAT and NCT was assessed by Pearson correlation co-efficient. CCT was measured with ultrasound pachymetry and its correlation with GAT and NCT was analysed using linear regression analysis.
RESULTS: The mean paired difference of IOP between NCT and GAT was 1.556 ± 2.69 mmHg (r = 0.26, p = 0.006) at IOP range of ≤ 12 mmHg, -1.665 ± 2.6 mmHg (r = 0.51, p < 0.0001) in IOP range of 13-20 mmHg and -2.202 ± 3.44 mmHg (r = 0.82, p < 0.0001) in the IOP range of ≥ 21 mmHg. Linear regression analysis showed a mean IOP variation of 0.27 mmHg per 10 μm change in CCT for NCT (p < 0.0001) and IOP change of 0.19 mmHg per 10 μm change in CCT for GAT (p = 0.01).
CONCLUSION: In this study of normotensive and POAG subjects, the Topcon CT-80 NCT showed an overestimation of IOP at the lower range and underestimation of IOP in normal and higher ranges of IOP. Clinicians should keep in mind that CCT influences IOP measurement with both types of tonometer and that the IOP readings obtained with these tonometers are not interchangeable.
© 2017 Optometry Australia.

Entities:  

Keywords:  Goldmann applanation tonometer; Topcon CT-80 NCT; central corneal thickness

Mesh:

Year:  2017        PMID: 29044716     DOI: 10.1111/cxo.12620

Source DB:  PubMed          Journal:  Clin Exp Optom        ISSN: 0816-4622            Impact factor:   2.742


  6 in total

1.  Ocular biometric parameters are associated with non-contact tonometry measured intraocular pressure in non-pathologic myopic patients.

Authors:  Yazhen Ma; Yanping Ma; Chenli Feng; Minqian Shen; Yuanzhi Yuan
Journal:  Int Ophthalmol       Date:  2019-12-02       Impact factor: 2.031

2.  Serum Calcium Level as a Useful Surrogate for Risk of Elevated Intraocular Pressure.

Authors:  Yu-Min Chang; Jiann-Torng Chen; Ming-Cheng Tai; Wei-Liang Chen; Ying-Jen Chen
Journal:  J Clin Med       Date:  2021-04-23       Impact factor: 4.241

3.  Comparability of three intraocular pressure measurement: iCare pro rebound, non-contact and Goldmann applanation tonometry in different IOP group.

Authors:  Min Chen; Lina Zhang; Jia Xu; Xinyi Chen; Yuxiang Gu; Yuping Ren; Kaijun Wang
Journal:  BMC Ophthalmol       Date:  2019-11-14       Impact factor: 2.209

4.  Multiple scan averaging to yield accurate quantitative analysis of optical coherence tomography angiograms.

Authors:  Hafi M Khan; Alex Gentle; James A Armitage; Chi-Ho To; Andrew K C Lam
Journal:  Sci Rep       Date:  2020-04-10       Impact factor: 4.379

5.  The Association of Central Corneal Thickness and Intraocular Pressure Measures by Non-Contact Tonometry and Goldmann Applanation Tonometry among Glaucoma Patients.

Authors:  Samuel Kyei; Frank Assiamah; Michael Agyemang Kwarteng; Cynthia Pakyennu Gboglu
Journal:  Ethiop J Health Sci       Date:  2020-11

6.  Comparison of Non-contact Tonometry and Goldmann Applanation Tonometry Measurements in Non-pathologic High Myopia.

Authors:  Peiyuan Wang; Yunhe Song; Fengbin Lin; Zhenyu Wang; Xinbo Gao; Weijing Cheng; Meiling Chen; Yuying Peng; Yuhong Liu; Xiulan Zhang; Shida Chen
Journal:  Front Med (Lausanne)       Date:  2022-03-03
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.