| Literature DB >> 29026283 |
Wilson Takashi Hida1,2,3,4, Patrick Frenzel Tzelikis1,2,3, César Vilar1,2, Mario Augusto Pereira Dias Chaves1,2,5, Antônio Francisco Pimenta Motta1,2,3, Pedro Carlos Carricondo1,2,3, Bruna Vieira Ventura1,2,4,6, Renato Ambrosio1,2,3, Walton Nosé1,2,4, Milton Ruiz Alves1,2,3.
Abstract
PURPOSE: The purpose of this study was to compare intraoperative parameters between femtosecond laser-assisted cataract surgery (FLACS) and conventional phacoemulsification using Centurion® Vision System with Active Fluidics.Entities:
Keywords: FLACS; aspiration; ultrasound
Year: 2017 PMID: 29026283 PMCID: PMC5627560 DOI: 10.2147/OPTH.S136136
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Age, LOCS III and PNS between FLACS and conventional surgery using Centurion Vision System with Active Fluidics Phaco Platform
| FLACS
| Conventional
| ||
|---|---|---|---|
| Average±SD | Average±SD | ||
| Age | 67.9±8.7 | 69.1±8.9 | 0.193 – ns |
| LOCS III | 2.02±0.44 | 1.98±0.72 | 0.061 – ns |
| PNS | 1.60±0.80 | 1.52±0.92 | 0.371 – ns |
Abbreviations: LOCS, Lens Opacity Classification System; PNS, Pentacam® Nucleus Staging; FLACS, femtosecond laser-assisted cataract surgery; ns, not significant.
Figure 1Comparison of CDE between FLACS and conventional surgery using Centurion Vision System with Active Fluidics Phaco Platform.
Notes: P-value is from a paired t-test. *P<0.0001.
Abbreviations: CDE, cumulative dissipated energy; FLACS, femtosecond laser-assisted cataract surgery.
Figure 2Comparison of torsional time (s) between FLACS and conventional surgery using Centurion Vision System with Active Fluidics Phaco Platform.
Notes: P-value is from a paired t-test. *P<0.0001.
Abbreviation: FLACS, femtosecond laser-assisted cataract surgery.
Figure 3Comparison of phacoemulsification time (s) between FLACS and conventional surgery using Centurion Vision System with Active Fluidics Phaco Platform.
Notes: P-value is from a paired t-test. **P=0.0138.
Abbreviation: FLACS, femtosecond laser-assisted cataract surgery.
Figure 4Comparison of fluid usage (BSS), case time (s) and ASP time (s) between FLACS and conventional surgery using Centurion Vision System with Active Fluidics Phaco Platform.
Notes: P-value is from a paired t-test. *P<0.0001.
Abbreviations: BSS, balanced salt solution; ASP, aspiration; FLACS, femtosecond laser-assisted cataract surgery.