| Literature DB >> 29017580 |
Hui Zhang1, Xiaoxiao Zhu2, Genwang Pei3, Xianshang Zeng4, Nan Zhang5, Ping Xu6, Deng Chen7, Weiguang Yu4, Xinchao Zhang8.
Abstract
BACKGROUND: The study aims to compare the long-term functional and radiographic outcomes of two devices for the treatment of primary intertrochanteric fractures (IFs, OTA 3.1A2.1-A2.3) in elderly patients with osteoporosis.Entities:
Keywords: Harris Hip Score; InterTAN nail; Intertrochanteric fracture; Proximal femoral nail anti-rotation
Mesh:
Year: 2017 PMID: 29017580 PMCID: PMC5634834 DOI: 10.1186/s13018-017-0648-2
Source DB: PubMed Journal: J Orthop Surg Res ISSN: 1749-799X Impact factor: 2.359
Patient demographics and outcomes
| Variable | ITa ( | PFNAb ( |
|
|---|---|---|---|
| Sex, no. M/F | 64/80 | 53/86 | 0.281*c |
| Age (years) | 0.839*d | ||
| 65–69 | 40 | 36 | |
| 70–79 | 76 | 74 | |
| 80–89 | 22 | 20 | |
| 90–92 | 6 | 9 | |
| BMI, kg/m2 | 26.1 ± 2.1 | 25.7 ± 1.9 | 0.130*e |
| BMD | −3.7 ± 0.3 | −3.7 ± 0.4 | 0.179*e |
| Medical complications | 9 | 7 | 0.658*c |
| Side, no. Left/right | 81/63 | 72/67 | 0.453*c |
| OTA fracture type, no. | 0.679*d | ||
| 31A2.1 | 44 | 36 | |
| 31A2.2 | 72 | 75 | |
| 31A2.3 | 28 | 28 | |
| ASA classification | 0.501*d | ||
| 1 | 23 | 25 | |
| 2 | 65 | 57 | |
| 3 | 49 | 54 | |
| 4 | 7 | 3 | |
| Injury operation interval | 0.702*d | ||
| < 24 h | 22 | 23 | |
| 24–48 h | 77 | 79 | |
| 48–72 h | 29 | 27 | |
| > 72 h | 16 | 10 | |
| Follow-up (months) | 38.7 ± 3.55 | 39.1 ± 3.43 | 0.408*e |
IT InterTAN nail, PFNA proximal femoral nail anti-rotation, ASA American Society of Anesthesiologists, BMI body mass index, BMD bone mineral density
*No statistically significant values
aSmith & Nephew, Memphis, Tennessee
bSynthes, Solothurn, Switzerland
cAnalyzed using chi-square test
dAnalyzed using the Mann-Whitney test
eAnalyzed using independent samples t-test
Fig. 1Flow diagram demonstrating methods for identification of studies to assess elderly osteoporosis patients with IFs (OTA 3.1A2.1–A2.3) fixed by an InterTAN nail (IT) or a proximal femoral nail anti-rotation (PFNA) device
Fig. 2The bar chart clearly shows the changes of the number of patients being lost to follow-up at each time point. The “xx” on the bar chart represents “final follow-up”
The number of patients being lost to follow-up at each time point
| 1 | 3 | 6 | 12 | 15 | 18 | 24 | 27 | 30 | 36 | F | |
|---|---|---|---|---|---|---|---|---|---|---|---|
| IT | 144 | 141 | 140 | 138 | 138 | 137 | 136 | 133 | 130 | 127 | 124 |
| PFNA | 139 | 136 | 134 | 132 | 130 | 130 | 128 | 124 | 120 | 118 | 115 |
| N | 0 | 6 | 3 | 4 | 2 | 1 | 3 | 7 | 7 | 5 | 6 |
N the number of patients being lost to follow-up, F final follow-up
Operation variables
| Variable | ITa ( | PFNAb ( |
|
|---|---|---|---|
| Reduction results, no. | 0.552*c | ||
| Anatomical | 82 | 84 | |
| Acceptable | 62 | 55 | |
| Poor | 0 | 0 | |
| Implant position, no. | 0.948*c | ||
| Optimal | 101 | 97 | |
| Suboptimal | 43 | 42 | |
| TAD, mm | 19.9 ± 1.25 | 20.2 ± 1.36 | 0.144*d |
| Length of stay (days) | 11.9 ± 1.62 | 11.67 ± 1.88 | 0.21*d |
| Operation time (min) | 67.2 ± 3.98 | 68.9 ± 4.27 | 0.848*d |
| Blood loss (ml) | 180.7 ± 23.03 | 185.1 ± 18.96 | 0.078*d |
TAD tip-apex distance, IT InterTAN nail, PFNA proximal femoral nail anti-rotation
*No statistically significant values
aSmith & Nephew, Memphis, Tennessee
bSynthes, Solothurn, Switzerland
cAnalyzed using the Mann-Whitney test
dAnalyzed using independent samples t-test
Long-term follow-up outcomes
| Variable | ITa ( | PFNAb ( |
|
|---|---|---|---|
| Screw cut-out | 3 | 11 | 0.024*c |
| Femoral shaft fracture# | 3 | 12 | 0.014*c |
| Delayed union | 4 | 5 | 0.709c |
| Malunion | 2 | 0 | 0.498d |
| Non-union | 2 | 0 | 0.498d |
| Varus collapse of the femoral head | 1 | 8 | 0.020*c |
| Lower limb shortening(> 1.5 cm) | 2 | 2 | 0.972c |
| Heterotopic ossification | 3 | 1 | 0.326c |
| Avascular necrosis of the femoral head | 1 | 1 | 1.000d |
| Postoperative thigh pain | 5 | 13 | 0.043*c |
| Lower extremity venous thrombosis | 7 | 6 | 0.797c |
| HHS at the last follow-up | 72.2 ± 7.27 | 72.4 ± 7.20 | 0.842e |
| Mortality | 19 | 22 | 0.529c |
IT InterTAN nail, PFNA proximal femoral nail anti-rotation, HHS Harris Hip Score
*Statistically significant values
#Displaced shaft fractures distal or around the tip of the main nail
aSmith & Nephew, Memphis, Tennessee
bSynthes, Solothurn, Switzerland
cAnalyzed using the Pearson chi-squared test
dAnalyzed using Fisher’s exact test
eAnalyzed using independent samples t-test