Use of Volatile Anesthetic Agent in Extracorporeal Circuit as a Cause of Break in Polycarbonate Connector-Lessons Learnt.

Deepak Gowda1, K Rashmi1, Naveen Pandarinathan1, Neelam Desai1.   

Abstract

Mishaps, near misses, and lethal incidents are known to occur during cardiopulmonary bypass. We share one such rare case of break in polycarbonate connector because of the use of isoflurane in extracorporeal circuit and its successful management.

Entities:  

Keywords:  atrial septal defect; cardiopulmonary bypass; isoflurane; polycarbonate

Mesh:

Substances:

Year:  2017        PMID: 28979044      PMCID: PMC5621584     

Source DB:  PubMed          Journal:  J Extra Corpor Technol        ISSN: 0022-1058


  9 in total

1.  A retrospective study on perfusion incidents and safety devices.

Authors:  B L Mejak; A Stammers; E Rauch; S Vang; T Viessman
Journal:  Perfusion       Date:  2000-01       Impact factor: 1.972

2.  Isoflurane damage to a Draeger Primus anaesthetic machine.

Authors:  T Vemmer; R Srikiran; B Liban
Journal:  Anaesthesia       Date:  2004-03       Impact factor: 6.955

3.  Isoflurane damage to a Draeger Primus water trap.

Authors:  C Abraham; S K Deshpande
Journal:  Eur J Anaesthesiol       Date:  2006-06       Impact factor: 4.330

4.  Ohmeda Tec 5 isoflurane vaporiser and damage to the water trap on the Draeger Primus anaesthesia machine.

Authors:  S Reddy; S M Bishop; J A Moore
Journal:  Anaesthesia       Date:  2008-08       Impact factor: 6.955

5.  Isoflurane-induced failure of the Bentley-10 oxygenator.

Authors:  D E Maltry; G W Eggers
Journal:  Anesthesiology       Date:  1987-01       Impact factor: 7.892

6.  Near catastrophic oxygenator failure.

Authors:  S Cooper; R Levin
Journal:  Anesthesiology       Date:  1987-01       Impact factor: 7.892

Review 7.  Leaching of the plasticizer di(2-ethylhexyl)phthalate (DEHP) from plastic containers and the question of human exposure.

Authors:  Hanno C Erythropel; Milan Maric; Jim A Nicell; Richard L Leask; Viviane Yargeau
Journal:  Appl Microbiol Biotechnol       Date:  2014-11-07       Impact factor: 4.813

8.  Isoflurane cracks the polycarbonate connector of extra-corporeal circuit -A case report-.

Authors:  Hyung Sun Lim; Sung Hun Cho; Deok Kyu Kim; Sang Kyi Lee; Seong Hoon Ko
Journal:  Korean J Anesthesiol       Date:  2010-03-29

9.  Development of a pump-oxygenator to replace the heart and lungs; an apparatus applicable to human patients, and application to one case.

Authors:  C DENNIS; D S SPRENG; G E NELSON; K E KARLSON; R M NELSON; J V THOMAS; W P EDER; R L VARCO
Journal:  Ann Surg       Date:  1951-10       Impact factor: 12.969

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.