Literature DB >> 28960123

Assessment of platelet function on the routine coagulation analyzer Sysmex CS-2000i.

Corinne Frère1, Katsushi Kobayashi2, Claire Dunois2, Jean Amiral2, Pierre-Emmanuel Morange1, Marie-Christine Alessi1.   

Abstract

BACKGROUND: Light transmission aggregometry (LTA) is considered as the gold standard for testing platelet function in the setting of both platelet disorders suspicion and response to antiplatelet therapy evaluation. LTA requires however specialized equipment, substantial blood sample volumes, is technically challenging and time-consuming. AIM: To evaluate an automated platelet aggregation method performed on a routine coagulation analyzer Sysmex CS-2000i.
METHODS: 46 patients presenting a bleeding syndrome and 62 patients with acute coronary syndrome receiving dual antiplatelet therapy were studied in total. Platelet aggregations were performed on CS-2000i equipped with a dedicated software and on APACT-4004 (Elitech, France) as the reference instrument. Aggregation was measured by monitoring the changes in light absorbance occurring in response to ADP 2.5, 5 and 10µM, collagen 3.3 µg/mL, epinephrin 10µM, ristocetin 1.25 mg/mL and arachidonic acid 0.5 mg/mL in platelet rich plasma (PRP). PRP were tested simultaneously on both CS-2000i and APACT-4004 devices. Platelet stirred speed were 800 rpm for both instruments.
RESULTS: Significant correlations were observed between CS-2000i and LTA after all stimulations (p< 0.001). Patients presenting a bleeding syndrome had similar aggregation profiles with both methods. A single patient presented a severe platelet disorder (Glanzmann Thrombasthenia) and its PRP showed defective aggregation in response to all agonists except ristocetin with both instruments. Finally, the inter-agreement rates for CS-2000i and APACT-4004 to detect low responders to thienopyridines or aspirine were strong (weighted kappa> 0.70).
CONCLUSION: Platelet aggregation on the routine coagulation analyzer CS-2000i is an easily accessible, handy, reliable, standardized, and rapid tool to assess platelet function which allows to skirt most of the LTA limitations.

Entities:  

Keywords: 

Mesh:

Substances:

Year:  2017        PMID: 28960123     DOI: 10.1080/09537104.2017.1353683

Source DB:  PubMed          Journal:  Platelets        ISSN: 0953-7104            Impact factor:   3.862


  5 in total

1.  Platelet Aggregation Testing on a Routine Coagulation Analyzer: A Method Comparison Study.

Authors:  Jan Stratmann; Lida Karmal; Birga Zwinge; Wolfgang Miesbach
Journal:  Clin Appl Thromb Hemost       Date:  2019 Jan-Dec       Impact factor: 2.389

2.  Predicting bleeding risk in a Chinese immune thrombocytopenia (ITP) population: development and assessment of a new predictive nomogram.

Authors:  Mingjing Wang; Weiyi Liu; Yonggang Xu; Hongzhi Wang; Xiaoqing Guo; Xiaoqing Ding; Richeng Quan; Haiyan Chen; Shirong Zhu; Teng Fan; Yujin Li; Xuebin Zhang; Yan Sun; Xiaomei Hu
Journal:  Sci Rep       Date:  2020-09-18       Impact factor: 4.379

3.  Antiplatelet Activity of Coumarins: In Vitro Assays on COX-1.

Authors:  Cristina Zaragozá; Francisco Zaragozá; Irene Gayo-Abeleira; Lucinda Villaescusa
Journal:  Molecules       Date:  2021-05-19       Impact factor: 4.411

4.  The clinical efficacy of using autologous platelet-rich plasma in total hip arthroplasty: A retrospective comparative study.

Authors:  Xi Wang; Jianbin Ma; Zhiyuan Wang; Lin Xiao
Journal:  Medicine (Baltimore)       Date:  2018-10       Impact factor: 1.817

5.  A multicenter study to evaluate automated platelet aggregometry on Sysmex CS-series coagulation analyzers-preliminary findings.

Authors:  Sean Platton; Áine McCormick; Musfira Bukht; David Gurney; Ian Holding; Gary W Moore
Journal:  Res Pract Thromb Haemost       Date:  2018-08-30
  5 in total