| Literature DB >> 28955687 |
Asep Santoso1, Ik-Sun Choi2, Kyung-Soon Park2, Taek-Rim Yoon2.
Abstract
Proximal femoral cephalomedullary nailing is one of the treatments of choice for intertrochanteric fracture. Zimmer Natural Nail® (ZNN; Zimmer) is one of the alternatives. We report two cases of broken tail portion of the ZNN lag screw during the removal procedure. This may be the first reported cases in scientific literature with this pattern of failure. We report these cases to highlight one of the possible problems that could occur during removal of this implant which can possibly prolong the operation time significantly if the required device is not prepared.Entities:
Keywords: Breakage; Hip fractures; Lag screw; Zimmer Natural Nail
Year: 2017 PMID: 28955687 PMCID: PMC5612981 DOI: 10.5371/hp.2017.29.3.199
Source DB: PubMed Journal: Hip Pelvis ISSN: 2287-3260