Ljubisa Borota1, Andreas Patz2. 1. 1 Department of Surgical Sciences, Uppsala University, Uppsala, Sweden. 2. 2 Toshiba Medical Systems Europe, Zoetermeer, the Netherlands.
Abstract
Aim of the study A new functionality that enables vertical mobility of the lateral arm of a biplane angiographic machine is referred to as the flexible lateral isocenter. The aim of this study was to analyze the impact of the flexible lateral isocenter on the air-kerma rate under experimental conditions. Material and methods An anthropomorphic head-and-chest phantom with anteroposterior (AP) diameter of the chest varying from 22 cm to 30 cm simulated human bodies of different body constitutions. The angulation of the AP arm in the sagittal plane varied from 35 degrees to 55 degrees for each AP diameter. The air-kerma rate (mGy/min) values were read from the system dose display in two settings for each angle: flexible lateral isocenter and fixed lateral isocenter. Results The air-kerma rate was significantly lower for all AP diameters of the chest of the phantom when the flexible lateral isocenter was used: (a) For 22 cm, the p value was 0.028; (b) For 25 cm, the p value was 0.0169; (c) For 28 cm, the p value was 0.01005 and (d) For 30 cm, the p value was 0.01703. Conclusion Our results show that the flexible lateral isocenter contributes significantly to the reduction of the air-kerma rate, and thus to a safer environment in terms of dose lowering both for patients and staff.
Aim of the study A new functionality that enables vertical mobility of the lateral arm of a biplane angiographic machine is referred to as the flexible lateral isocenter. The aim of this study was to analyze the impact of the flexible lateral isocenter on the air-kerma rate under experimental conditions. Material and methods An anthropomorphic head-and-chest phantom with anteroposterior (AP) diameter of the chest varying from 22 cm to 30 cm simulated human bodies of different body constitutions. The angulation of the AP arm in the sagittal plane varied from 35 degrees to 55 degrees for each AP diameter. The air-kerma rate (mGy/min) values were read from the system dose display in two settings for each angle: flexible lateral isocenter and fixed lateral isocenter. Results The air-kerma rate was significantly lower for all AP diameters of the chest of the phantom when the flexible lateral isocenter was used: (a) For 22 cm, the p value was 0.028; (b) For 25 cm, the p value was 0.0169; (c) For 28 cm, the p value was 0.01005 and (d) For 30 cm, the p value was 0.01703. Conclusion Our results show that the flexible lateral isocenter contributes significantly to the reduction of the air-kerma rate, and thus to a safer environment in terms of dose lowering both for patients and staff.
The treatment of cerebral aneurysms, arteriovenous malformations and fistulas is
always very challenging for many reasons. An inappropriate approach to the target,
suboptimal catheterization of vascular structures, with subsequent erroneous
injection of various embolic materials; or deployment of implants may lead to the
unsatisfactory treatment of vascular pathology or even to vascular catastrophes. The
main prerequisite for optimal treatment is optimal visualization of the target in
two projections.[1-4] All state-of-the-art
angiographic machines possess two arms intended for simultaneous visualization of
the target in two planes, anteroposterior (AP) and lateral. Extreme angulations of
both arms are often necessary to obtain optimal visualization of a vascular
structure. The table and the patient on it, as well as all anesthetic and
neurosurgical instruments necessary for a neurointervention, limit angulations of
both arms. This problem can partially be solved by changing the position of the
table, since the position of the table is adjustable in all three planes, regardless
of the model of angiographic machine used. An appropriate combination of table
position, angulations of both arms, as well as distances between detector panels and
X-ray tubes in both projections is therefore currently being used for optimal
visualization of the target.A new functionality, described in this study, that enables mobility of the lateral
arm in the vertical direction additionally contributes to the adjustability of the
lateral arm and thus to better visualization of vascular structures regardless of
their position, size or shape. This functionality is referred to as the flexible
lateral isocenter (FLIC). Thanks to this novel functionality, it is also possible to
obtain the desired projections with the minimal possible distance between the
detector plate and X-ray tube.The aim of this study was to show the impact of this new functionality on the
radiation dose, since one of the factors on which the dose is dependent is the
distance between the detector panel and X-ray tube.
Material and methods
The biplane machine used in this study was the Toshiba Infinix-i/BP (Toshiba Medical
Systems, Shimoishigami, Otawara-shi, Tochigi, Japan). The technical characteristics
of the fluoroscopy used for the measurements were detector input dose of 0.45 µGy/s
for a reference field of view (FOV) of 20 × 20 cm, which was used for all
measurements. The X-ray factors were voltage, 80 kV; range of current, 50–200 mA;
and filtration, 0.3 mm copper. Current (mA) and time (ms) are not constant in
fluoroscopy as they vary according to the automatic brightness control (ABC)
response.FLIC enables adjustment of the lateral arm ±7 cm in the vertical direction.The air-kerma rate (mGy/min) values were read from the system dose display.The head of an anthropomorphic head-and-chest phantom with AP diameters of the chest
measuring 22 cm, 25 cm, 28 cm and 30 cm for each dose measurement was used as a
target. The AP diameter of the thorax was extended by adding 1 or 2 cm thick
Plexiglas plates. The distance between the detector plate of the AP arm and the
thorax was always 2cm. The caudo-cranial angulation of the AP arm varied from 35
degrees to 55 degrees, with a 5-degree increment between measurements.To achieve the needed caudo-cranial angulation, to keep lateral projection of the
target, and to avoid collision between the AP arm and the chest of the phantom, the
distance between the X-ray tube and detector plate increased for each angulation
when the non-flexible isocenter (non-FLIC) was used (Figure 1). When the FLIC was used, the same
aim was achieved by vertical adjustment of the lateral arm and a marginal increase
of the distance between X-ray tube and detector plate (Figure 2). For both FLIC and non-FLIC setups
the exposed anatomical area and volume (head of the head-and-chest phantom) were the
same for each measurement. This ensured the air-kerma values for both setups were
comparable and dependent only on the distance between the X-ray tube and detector
plate.
Figure 1.
Use of the non-flexible lateral isocenter. To get the desired working
anteroposterior projection and to keep the lateral projection by keeping
the height of the table unchanged, the source-to-image distance has been
extended (blue arrow).
Figure 2.
Use of the flexible lateral isocenter. The same angulation as in Figure 1 with kept
lateral view is achieved using the flexible lateral isocenter that
compensates for table lowering (red arrows). The source-to-image
distance is shorter compared to the non-flexible lateral isocenter (blue
arrow).
Use of the non-flexible lateral isocenter. To get the desired working
anteroposterior projection and to keep the lateral projection by keeping
the height of the table unchanged, the source-to-image distance has been
extended (blue arrow).Use of the flexible lateral isocenter. The same angulation as in Figure 1 with kept
lateral view is achieved using the flexible lateral isocenter that
compensates for table lowering (red arrows). The source-to-image
distance is shorter compared to the non-flexible lateral isocenter (blue
arrow).The Excel T test was used for statistical analysis of the results.
For the difference between air-kerma values, a p value <0.05 was
considered significant.
Results
(a) Thorax AP diameter: 22 cm
When the non-FLIC was used, the distance between the X-ray tube and detector
plate increased from 97 cm to 118 cm as the angulation in caudo-cranial
direction increased from 35 degrees to 55 degrees. Thus the lateral projection
was not lost regardless of angulation, and the detector plate did not collide
with the thorax of the phantom. The increase of distance caused an increase of
dose from 16.3 mGy/min to 39.7 mGy/min (Figure 3). When the FLIC was used, the
lowering of the table was compensated for by vertical adjustment of the lateral
arm and a marginal increase of distance between the X-ray tube and detector
plate: from 95 cm for 35 degrees to 105 cm for 55 degrees. This increase in
distance caused an increase in dose from 14 mGy/min to 19.2 mGy/min (Figure 3).
Figure 3.
Air-kerma rate is significantly lower for thorax anteroposterior
diameter of 22 cm when the flexible lateral isocenter was used,
p value = 0.0280.
Air-kerma rate is significantly lower for thorax anteroposterior
diameter of 22 cm when the flexible lateral isocenter was used,
p value = 0.0280.
(b) Thorax AP diameter: 25 cm
When the FLIC was used, the increase of angulation from 35 degrees to 55 degrees
was compensated for by vertical adjustment of the lateral arm and an increase in
the distance between the X-ray tube and detector plate from 95 cm to 108 cm,
which was followed by a dose increase from 15.9 mGy/min to 29.8 mGy/min (Figure 4). For the same
angulations, the distance between the X-ray tube and detector plate increased
from 101 cm to 123 cm when the non-FLIC was used. This increase of distance
caused an increase of the dose from 17.9 mGy/min to 40.6 mGy/min (Figure 4).
Figure 4.
Air-kerma rate is significantly lower for thorax anteroposterior
diameter of 25 cm when the flexible lateral isocenter was used,
p value = 0.0169.
Air-kerma rate is significantly lower for thorax anteroposterior
diameter of 25 cm when the flexible lateral isocenter was used,
p value = 0.0169.
(c) Thorax AP diameter: 28 cm and 30 cm
For these two AP diameters, the angulation of 55 degrees could not be reached
when the non-FLIC was used. For an AP thorax diameter of 28 cm, the distance
between the X-ray tube and detector plate increased from 103 cm to 125 cm and an
angulation of 54 degrees was reached. This was followed by a dose increase from
18.7 mGy/min to 44.2 mGy/min (Figure 5). Similarly, for the AP thorax diameter of 30 cm, the
distance between the X-ray tube and detector plate increased from 106 cm to
124 cm and an angulation of 50 degrees was reached. This increase in distance
led to a dose increase from 19.9 mGy/min to 40.2 mGy/min (Figure 6). The angulation of 55 degrees
was reached for both AP thorax diameters when the FLIC was used. For these two
diameters, the distance between the X-ray tube and detector plate increased from
95 cm to 113 cm and from 96 cm to 116 cm, respectively. This was followed by a
dose increase from 15.6 mGy/min to 34.9 mGy/min and 16.8 mGy/min to
41.5 mGy/min, respectively (Figures 5 and 6).
Figure 5.
Air-kerma rate is significantly lower for thorax anteroposterior
diameter of 28 cm when the flexible lateral isocenter was used,
p value = 0.01005. When the non-flexible
lateral isocenter was used, the maximum achieved angulation was 54
degrees.
Figure 6.
Air-kerma rate is significantly lower for thorax anteroposterior
diameter of 30 cm when the flexible lateral isocenter was used,
p value = 0.01703. When the non-flexible
lateral isocenter was used, the maximum achieved angulation was 50
degrees.
Air-kerma rate is significantly lower for thorax anteroposterior
diameter of 28 cm when the flexible lateral isocenter was used,
p value = 0.01005. When the non-flexible
lateral isocenter was used, the maximum achieved angulation was 54
degrees.Air-kerma rate is significantly lower for thorax anteroposterior
diameter of 30 cm when the flexible lateral isocenter was used,
p value = 0.01703. When the non-flexible
lateral isocenter was used, the maximum achieved angulation was 50
degrees.The Excel T Test showed that the dose when the FLIC was used was
significantly lower (0.001 < p value < 0.05) than the
dose when the non-FLIC was used for each thorax diameter (Figures 3–6):(e) For 22 cm, the p value was 0.028;(f) For 25 cm, the p value was 0.0169;(g) For 28 cm, the p value was 0.01005 and(h) For 30 cm, the p value was 0.01703.
Discussion
Since the beginning of the era of endovascular treatment of cerebral aneurysms,
arteriovenous malformations and fistulas, optimal visualization of these targets has
been the main prerequisite for a successful treatment and the object of numerous
studies.[1-3] The
visualization of these vascular structures has evolved from monoplane and biplane
fluoroscopy to three-dimensional (3D) rotational angiography.[4,5] Moreover, the evolution of
computerized tomographic angiography as well as magnetic resonance tomographic
angiography enabled almost noninvasive visualization of the angioarchitecture of
cerebral aneurysms, arteriovenous malformations and fistulas without exposing
patients to the risks related to trans-catheter angiography.[6-8] The preoperative, 3D
visualization of cerebral vascular structures is the first and very important step
in the planning of the treatment of these conditions. Since the consequences of
complications that sometimes occur during endovascular intervention are often
disastrous, an adequate visualization of the vascular target during treatment is
essential for the appropriate choice of devices and techniques. The biplane systems
have become an obligatory part of the equipment of an angiographic theater intended
for neurointerventional procedures. The superiority of a biplane system compared to
a monoplane system cannot be scientifically proven based on existing data in the
scientific literature, but this system can definitely improve operators confidence,
which should provide a better treatment in terms of reduced risk of complications
and better radiological and clinical result of the intervention.[9]All state-of-the-art biplane machines possess two arms intended for visualization of
vascular structures in two planes, which are, in the neutral position, perpendicular
to each other. During the intervention, the angulation of these arms is changed in
order to obtain the optimal visualization of a vascular target. Both arms are
constructed so that the angle between the axis of the X-rays and the detector plate
is 90 degrees, and this axis always hits the intersection of the diagonals of the
detector plate. This intersection, which is the geometric center of the detector
plate, is referred to as the isocenter. The distance between the X-ray source and
the detector plate is flexible and referred to as the source-to-image distance
(SID). Beside the angulations, the only possible mobility of the lateral isocenter
of all state-of-the-art biplane machines, except the Toshiba Infinix-i/BP, is the
mobility in the horizontal direction. The FLIC, which enables additional mobility in
the vertical direction of the lateral isocenter, is integrated in all commercially
available Infinix-i/BP machines.The amplitude of the angulation of both arms is very high regardless of the model of
the bi-plane system because of the specific demands for optimal visualization of
cerebral vascular structures. A factor that limits the full capacity of the
angulation is the table that carries the body of the patient connected by tubes and
cables with anesthetic and neurosurgical systems for monitoring and maintenance of
vital functions. The amplitude of vertical movement of the FLIC is 14 cm, which
corresponds to approximately 75% of a mean AP diameter of the head of an adult human.[10] This means that the FLIC markedly expands the capacity of angulation of both
arms and in this way contributes to easier achievement of the optimal visualization
of vascular targets with challenging angioarchitecture. Our experimental work has
shown that the FLIC enables such visualization with a marginal increase of SID, at
least in the AP projection.Since the beginning of the era of neurointervention, X-ray equipment has evolved
tremendously in terms of improved image quality and decreased fluoroscopic and
angiographic doses. This development is based on technical improvements of X-ray
generators, hardware and software for acquisition of data and generation of images,
as well as on the improvements in radiation-protection systems integrated into the
biplane machines.[11-15] The biggest challenge to the
improvement of dose-saving systems is that the image quality must be kept on an
optimal level regardless of the dose reduction.[16] The most frequent and the easiest way of reducing the dose is optimization of
existing parameters of fluoroscopy and angiography.[16-19] Spot fluoroscopy, a
qualitatively new type of asymmetric collimation, represents a breakthrough in the
technology of dose-reduction systems.[20]The FLIC is also a qualitatively new technical solution that originally was aimed at
improvement of the amplitude of angulation of both arms. Thanks to the FLIC, even
extreme angulations can be achieved without or with marginal extension of the SID.
Thus the FLIC also prevents the dose increase caused by the SID increase, which is
inevitable if a non-FLIC system is used for achieving these angulations. This was
clearly shown by our experiment.This experimental study has certain limitations. The effect of the FLIC on the
amplitude of angulation has been described only briefly and not elaborated on in
detail since we assumed that this issue is comprehensible “per se.” The measurements
of the dose were carried out in only the sagittal plane because measurements in
several planes of both arms would involve a much more complex study design, with the
same or very similar results. Finally, the only parameter we measured in this
experiment was the air kerma rate.
Conclusion
The FLIC is a novel, original functionality intended for better exploitation of
existing technical capabilities of the arms of a biplane angiographic machine. Our
experimental work, in spite of certain methodological limitations, clearly shows
that the FLIC also contributes to a significant dose reduction, at least in the AP
plane. In other words, the FLIC is a functionality that effectively couples two
independent but equally important functionalities of each angiographic machine.
Authors: Maria T Karamessini; George C Kagadis; Theodore Petsas; Dimitrios Karnabatidis; Dimitrios Konstantinou; George C Sakellaropoulos; George C Nikiforidis; Dimitrios Siablis Journal: Eur J Radiol Date: 2004-03 Impact factor: 3.528
Authors: Michael Söderman; Maria Mauti; Sjirk Boon; Artur Omar; María Marteinsdóttir; Tommy Andersson; Staffan Holmin; Bart Hoornaert Journal: Neuroradiology Date: 2013-09-05 Impact factor: 2.804