S M O'Brien1,2, C Winter1, C A Burden1,2, M Boulvain3, T J Draycott1,4, J F Crofts1. 1. Department of Obstetrics & Gynaecology, Southmead Hospital, Bristol, UK. 2. School of Clinical Sciences, University of Bristol, Bristol, UK. 3. Department of Obstetrics & Gynaecology, Hôpitaux Universitaires de Genève, Geneva, Switzerland. 4. School of Social & Community Medicine, University of Bristol, Bristol, UK.
Abstract
OBJECTIVE: To determine the pressure and traction forces exerted on a model fetal head by the BD Odon Device, forceps and Kiwi ventouse during simulated births. DESIGN: Simulation study. SETTING: Simulated operative vaginal birth. POPULATION OR SAMPLE: Eighty-four simulated operative vaginal births. METHODS: A bespoke fetal mannequin with pressure sensors around the head and strain gauge across the neck was used to investigate pressure applied over the head, and traction across the neck during 84 simulated births using the BD Odon Device, non-rotational forceps and Kiwi ventouse. MAIN OUTCOME MEASURES: Peak pressure on the fetal face and lateral aspects of the head during correct use of the BD Odon Device and forceps. Peak pressure on orbits and neck during misplacement of the BD Odon Device and forceps. Peak traction force generated until instrument failure using the BD Odon Device, forceps and Kiwi ventouse. RESULTS: When correctly sited and using 80 kPa inflation pressure on the cuff, the BD Odon Device generated a lower peak pressure on the fetal head than forceps (83 versus 146 kPa). When instruments were purposefully misplaced over the orbits, the BD Odon Device generated a lower peak pressure on the orbits compared with forceps (70 versus 123 kPa). When purposefully misplaced over the neck, the BD Odon Device, compared with forceps, generated a greater peak pressure on the anterio-lateral aspect of the neck (56 versus 17 kPa) and a lower peak pressure on the posterior aspect of the neck (76 versus 93 kPa) than forceps. In cases of true cephalic disproportion, the BD Odon Device 'popped-off' at a lower traction force than did forceps (208 versus 270 N). CONCLUSIONS: In simulated assisted vaginal birth with correctly placed instruments, the peak pressure exerted on the fetal head by a BD Odon Device is lower than the pressure exerted by non-rotational forceps. In cases in which delivery of the fetal head is not possible due to cephalo-pelvic disproportion, lower traction forces could be applied using the BD Odon Device than with forceps before the procedure was abandoned due to device failure. TWEETABLE ABSTRACT: BD Odon Device exerts less pressure on a model fetal head than forceps, but more than Kiwi ventouse.
OBJECTIVE: To determine the pressure and traction forces exerted on a model fetal head by the BD Odon Device, forceps and Kiwi ventouse during simulated births. DESIGN: Simulation study. SETTING: Simulated operative vaginal birth. POPULATION OR SAMPLE: Eighty-four simulated operative vaginal births. METHODS: A bespoke fetal mannequin with pressure sensors around the head and strain gauge across the neck was used to investigate pressure applied over the head, and traction across the neck during 84 simulated births using the BD Odon Device, non-rotational forceps and Kiwi ventouse. MAIN OUTCOME MEASURES: Peak pressure on the fetal face and lateral aspects of the head during correct use of the BD Odon Device and forceps. Peak pressure on orbits and neck during misplacement of the BD Odon Device and forceps. Peak traction force generated until instrument failure using the BD Odon Device, forceps and Kiwi ventouse. RESULTS: When correctly sited and using 80 kPa inflation pressure on the cuff, the BD Odon Device generated a lower peak pressure on the fetal head than forceps (83 versus 146 kPa). When instruments were purposefully misplaced over the orbits, the BD Odon Device generated a lower peak pressure on the orbits compared with forceps (70 versus 123 kPa). When purposefully misplaced over the neck, the BD Odon Device, compared with forceps, generated a greater peak pressure on the anterio-lateral aspect of the neck (56 versus 17 kPa) and a lower peak pressure on the posterior aspect of the neck (76 versus 93 kPa) than forceps. In cases of true cephalic disproportion, the BD Odon Device 'popped-off' at a lower traction force than did forceps (208 versus 270 N). CONCLUSIONS: In simulated assisted vaginal birth with correctly placed instruments, the peak pressure exerted on the fetal head by a BD Odon Device is lower than the pressure exerted by non-rotational forceps. In cases in which delivery of the fetal head is not possible due to cephalo-pelvic disproportion, lower traction forces could be applied using the BD Odon Device than with forceps before the procedure was abandoned due to device failure. TWEETABLE ABSTRACT: BD Odon Device exerts less pressure on a model fetal head than forceps, but more than Kiwi ventouse.
Complications of operative vaginal birth (OVB) are related to pressure exerted by the
instrument on the fetal head (forceps exerting a positive pressure, vacuum exerting a
negative pressure) and the traction force required to complete the birth. Compared to
spontaneous vaginal birth, OVB is associated with higher rates of maternal perineal and anal
sphincter trauma and neonatal facial injury (forceps) and neonatal cephalohaematoma,
subgaleal and retinal haemorrhage (ventouse) (1,2). Furthermore, injuries are more
likely to occur if the use of OVB does not follow guidance concerning its safe use (3), including when an instrument is incorrectly
applied (4) or if there is prolonged traction
(5).The BD Odon Device (BD, Franklin Lakes, New Jersey, USA) is a new device being developed
for OVB. The BD Odon Device consists of an inflatable circular air cuff attached to a thin
circumferential polyethylene sleeve. A plastic applicator places the air cuff and sleeve
into the birth canal, past the widest diameter of the fetal head. The air cuff is inflated,
and the applicator removed. During maternal contractions the accoucheur applies traction to
the sleeve, to expedite the birth.The use of an air cuff positioned circumferentially around the fetal head as the
‘anchor point’ for traction has the potential to reduce fetal injury when
compared to forceps. Pressure applied to the fetal head during birth may be more evenly
distributed than pressure by forceps and therefore a lower risk of facial injury might be
expected. Similarly, the wider distribution of pressure may also reduce the risk of adverse
outcomes such as subgaleal haemorrhage and cephalohaematoma associated wit the use of
ventouse.We developed a simulation model to study the pressure and force applied across the fetal
head and neck during OVB. The model head was used to compare pressures on the fetal head in
births using (i) non-rotational forceps, (ii) Kiwi ventouse (Clinical Innovations, Salt Lake
City, Utah, USA) and (iii) the BD Odon Device. Aware that the risk of neonatal injury is
increased with incorrect use of a device, we also sought to simulate attempted OVB in true
cephalopelvic disproportion to measure the traction force that could be applied to a
non-deliverable fetus before the device failed (‘popped-off’) and to compare
the pressure exerted on vulnerable structures (fetal orbit and neck) when a BD Odon Device
or forceps were incorrectly applied.
Methods
Development of a fetal mannequin to measure dynamic pressure changes during simulated
operative vaginal birth
A bespoke fetal mannequin was designed and manufactured by a multi-professional team of
obstetricians, midwives, engineers and model makers. A PROMPT Flex® fetal mannequin
(Limbs & Things Ltd, Bristol, UK) was adapted. Pressure sensors (Tekscan®,
Boston, Massachusetts, USA) were mounted against a bespoke modelled fetal fetal skull and
neck. Three pressure sensors (Tekscan Pressure Mapping Sensor 5101: sensor pad dimensions
= 111.8mm x 1118mm, thickness 0.102mm; 1,936 sensels; sensel density = 15.5
sensels/cm2) covered the majority of the fetal skull including the entirety
of the face and the lateral aspects of the head. These locations are shown in Figure
1.An additional pressure sensor (Tekscan Pressure Mapping Sensor 6300: sensor pad
dimensions = 33.5mm x 264.2mm, thickness 0.102mm; 2,288 sensels; sensel density = 25.8
sensels/cm2) was placed around the fetal neck. The fetal neck was modified
with the addition of a silicone ‘collar’ to an anterio-posterior diameter of
58mm, equivalent to the 50th centile of fetal neck diameters at 40 weeks
gestation (6). Moulded silicone representing
features of the fetal face (nose, mouth, orbits and ears) and scalp skin (5mm thick) was
positioned over the pressure sensors to simulate a fetal head. The fetal mannequin had a
bi-parietal diameter (BPD) of 96mm, to simulate an average-sized term baby (BPD on
50th centile of 97mm) (7). A
calibration device (Tekscan PB15C) was used to equilibrate, calibrate and zero all
pressure sensors prior to each use.
Simulation of operative vaginal births
The bespoke pressure monitoring fetal mannequin was used with a standard PROMPT
Flex® maternal mannequin (Limbs & Things Ltd, Bristol, UK) to enable the
simulation of operative vaginal births. The traction across the fetal neck was measured
using the PROMPT Flex Force Monitoring fetal mannequin and associated software (Limbs and
Things Ltd, Bristol, UK).A series of simulated OVBs using the BD Odon Device, Kiwi ventouse and non-rotational
forceps were performed by a single operator (SO’B). The air cuff of the BD Odon
Device was inflated to 60kPa and 80kPa. This is the expected pressure range that will be
used in-vivo. The peak pressure over the face (right orbit, left orbit, nose and chin) and
lateral aspects of head were measured in 40 simulated births (cephalic presentation,
direct occipito-anterior, vertex 2cm below the ischial spines) in which either a BD Odon
Device (cuff inflation pressure 60kPa n = 10 or 80kPa n =10), non-rotational forceps
(n=10) or Kiwi ventouse (vacuum pressure 70N) (n=10) were correctly applied and used to
complete the birth of the fetal model in the standard manner.Peak pressure exerted on sensitive fetal structures (orbits and neck) were measured
throughout birth in an OA position at station 2cm below the ischial spines in 19
nonstandard scenarios: (i) with the BD Odon Device cuff placed purposefully over the orbit
and inflated to 60kPa (n = 3) (ii) with the BD Odon Device cuff placed purposefully over
the orbit and inflated to 80kPa (n = 3) (iii) with non-rotational forceps placed
purposefully over the orbit (n = 3) (iv) with the BD Odon Device cuff placed purposefully
around the neck and inflated to 60kPa (n = 5) (v) with the BD Odon Device cuff placed
purposefully around the neck and inflated to 80kPa (n = 5) and with non-rotational forceps
placed correctly on the fetal head (n = 5). Pressure data was initially captured and
analysed using the proprietary iScan® program (Tekscan, Boston, Massachusetts,
USA). The location of pressure sensors on the model fetal face and neck is shown in Figure
1.Twenty further scenarios were performed to evaluate the force at which a device would
detach or ‘pop-off’ the fetal head when the head was not deliverable. A
bespoke ‘pelvic shelf’ was produced to prevent decent and birth of the fetus
in order to simulate cephalo-pelvic disproportion. The traction force (N) exerted during
attempted non-rotational forceps (n=5), kiwi ventouse (n=5) and BD Odon Device with cuff
inflated to 60kPa (n = 5) and BD Odon Device with cuff inflated to 80kPa birth (n=5) on a
cephalic presentation, direct OA position with the vertex at the ischial spines was
measured using the integrated force monitoring device within the PROMPT Flex fetal
mannequin.Traction force data was captured by the PROMPT Flex® birthing simulator software
(Limbs &Things, Bristol, UK) at 20Hz and subsequently exported for analysis.Results are presented using descriptive statistical data due to the limited number of
repetitions within each scenario. Data are reported as mean values for each dataset with
full ranges of all values.
Results
Eighty-four simulated operative vaginal births were performed (Table 1).
Pressure over fetal face, lateral aspects of head, orbits, nose and mentum
Mean peak pressures over fetal face and lateral aspects of the head are shown in Table 2.
The mean peak pressure over the lateral aspects of the fetal head was greater using
non-rotational forceps (146kPa) compared to the BD Odon Device (109kPa at 60kPa air cuff
pressure and 83kPa at 80kPa air cuff pressure) and Kiwi ventouse (79kPa). The difference
in magnitude of these applied pressures over the lateral aspects of the head is
illustrated in Figure 2.Mean peak pressures over the fetal face was comparable between the simulated births
performed with non-rotational forceps (108kPa), Kiwi ventouse (96kPa) and BD Odon Device
with cuff inflation pressure of 60kPa (99kPa) and 80kPa (106kPa).Mean peak pressures over the orbits were greater using the BD Odon Device at both 60kPa
and 80kPa cuff inflation pressures (47kPa and 67kPa respectively) than non-rotational
forceps and Kiwi ventouse (24kPa and 19kPa respectively).Mean peak pressures over the nose were lower using the BD Odon Device at 80kPa compared
to all other scenarios, where the mean peak pressures were broadly comparable (Table
2).Mean peak pressures over the mentum were comparable using the BD Odon Device at 80kPa
inflation pressure (30kPa), non-rotational forceps (38kPa) and Kiwi ventouse (44kPa) and
higher in scenarios using the BD Odon Device at 60kPa inflation pressure (60kPa).
Pressures exerted when devices are incorrectly sited
Three simulated births were performed with the BD Odon Device air cuff purposefully
incorrectly sited over the left orbit and inflated to 60kPa, and a further three simulated
births with the air cuff inflated to 80kPa. Peak pressures over the left fetal orbit were
compared to three simulated births in which the non-rotational forceps were also
incorrectly sited to lie over the left fetal orbit. It was only possible to perform three
births for each scenario due to sensor degradation during these tests, so robust
statistical comparison is not possible. However, the measurements suggest that incorrectly
placed forceps generate substantially greater mean peak pressure over the fetal orbit
(123kPa) than an incorrectly positioned BD Odon Device inflated to 60kPa or 80kPa (60kPa
and 70kPa respectively) (Table 3).The air cuff of the BD Odon Device was purposefully placed around the fetal neck (50mm
below the fetal chin) and inflated to 60kPa and 80kPa. Five OVBs were performed with the
fetus in a direct OA position at each inflation pressure. A comparison of applied peak
pressure was made with five non-rotational forceps births (Table 4). Forceps tended to
exert a higher median peak pressure on the posterior aspect of the fetal neck (94kPa) when
compared to BD Odon Device with the cuff inflated to 60kPa (87kPa) and 80kPa (76kPa).
However, the median peak pressure applied to the anterio-lateral aspects of the fetal neck
(the likely location of the fetal carotid arteries) by an BD Odon Device at 60kPa (59kPa)
and 80kPa (56kPa) inflation was greater than that generated with non-rotational forceps
(17kPa).
Evaluation of the force at which a device detaches from the model fetal head when the
head is not deliverable
The maximum traction force applied before the device detached from the fetal head during
an obstructed OVB (in which it was not possible for the fetal head to be delivered) was
greater in non-rotational forceps (270N) compared to attempts using the BD Odon Device at
80kPa inflation pressure (208N), 60kPa inflation pressure (167N) and Kiwi ventouse (70N)
(Table 5).
Discussion
Main Findings
The BD Odon Device, when correctly sited, generates less peak pressure on a model fetal
face than correctly sited forceps, but higher peak pressure than Kiwi ventouse. The
mechanism of action of the Kiwi ventouse, whereby there is no instrument in contact with
the face or lateral aspect of the head, clearly explains the lower pressures for this
instrument. When incorrectly sited, the BD Odon Device generates less pressure on
vulnerable facial structures (the orbit) than forceps.When the BD Odon Device was purposefully placed around the neck (previous simulation work
has demonstrated that this is unlikely to occour), it generates more pressure over the
anterior, but less pressure over the posterior aspect of the neck than forceps.When used inappropriately in an obstructed birth and used forcefully until device
failure, the BD Odon Device generates substantially less traction than forceps but more
than Kiwi ventouse. However, as our simulator was unable to generate a chignon, the true
pop-off/failure force for a Kiwi ventouse is likely to be higher in clinical practice
– previous studies have demonstrated pop-off forces of between 110N to 130N (5).
Interpretation
The BD Odon Device generated lower levels of peak pressure over the lateral aspects of
the fetal head than forceps, but higher levels than Kiwi ventouse. This is biologically
plausible.Forceps have a much lower instrument surface area in contact with the fetal head (the
blades) than the BD Odon Device (the circumferential air cuff) hence identical traction
forces will result in lower pressure peak pressure exerted by the Odon Device when
compared to forceps. It is therefore plausible that the risk of neonatal injuries
specifically associated with high peak pressures, such as facial nerve palsy, scalp
injury, skull fracture and bruising (2) are
likely to be lower in OVBs using the BD Odon Device than those conduced using forceps. The
low pressure detected on the lateral aspects of the head during a Kiwi ventouse birth is
in-keeping with the birthing mechanism and lack of contact of the instrument with the
lateral aspects of the fetal head.Direct pressure to the orbit during birth can result in serious and permanent ophthalmic
injuries (8). The peak pressure generated by
the BD Odon Device, at both inflation pressures of 60kPa and 80kPa, when placed directly
over the orbit was substantially lower than that generated by forceps – this is
likely to correlate to lower rates of trauma to the face during birth if the BD Odon
Device is incorrectly sited compared to incorrectly sited forceps.The BD Odon Device generated lower peak pressure over the posterior aspect of the neck
compared to forceps. This may reflects the mechanism in which a baby in the OA position
extends it’s neck as it negotiates the pelvic curve. Pressure is exerted on the
posterior aspect of the neck as the fetus lies directly beneath the pubic symphysis,
acting as a locus around which the fetal head extends. However, when a baby is delivered
with the assistance of a BD Odon Device that has been purposefully misplaced around the
fetal neck the air cuff rests between the posterior aspect of the neck and the pubic
symphysis and appears to act as a cushion, redistributing pressure around the
circumference of the neck.We acknowledge that the simulated pressure readings can not be a true reflection of the
exact pressures exerted in clinical practice. However the relative degree and distribution
of pressures in vivo are likely to be similar to those we have observed in simulation.
This simulation study suggests that the BD Odon Device generates approximately half the
peak pressure generated by the forceps, with pressure distributed across a wider area i.e.
there is less point pressure.The clinical significance of the observed pressure on the anterior portion of the neck
when the BD Odon Device is purposefully misplaced is unclear. Animal studies and clinical
observation of 48 births in healthy volunteers in Argentina suggest the BD Odon Device is
extremely unlikely to be placed around the fetal neck. Reported mean systolic blood
pressure of term neonates is 72.6 (SD 9.0) mmHg (9) therefore even if a pressure of 59kPa was exerted on the fetal neck by a
misplaced Odon Device the systolic circulation through the carotid arteries should not be
occluded. In addition, the BD Odon Device is unlikely to exert this peak pressure at any
point other than during traction by the operator during a contraction.Greater traction forces used in OVBs correlate with higher rates of neonatal injury and
maternal anal sphincter damage (5). The BD Odon
Device generates less traction force before device failure than forceps. The incidence of
adverse outcomes related to inappropriate traction force applied using an BD Odon Device
is therefore likely to be less than those associated with forceps. Previous research has
demonstrated that traction forces of 110 to 130N are routine using Kiwi (5), suggesting that rates of adverse outcomes due
to high traction may be comparable between the BD Odon Device and the Kiwi ventouse. The
BD Odon Device does not generate negative pressure on the fetal head, reducing the
likelihood of adverse outcomes such as subgaleal or retinal haemorrhage and
cephalohaematoma being associated with the use of a BD Odon Device when compared to vacuum
assisted births.
Strengths and Limitations
This is the first study to attempt to quantify the pressures exerted on a baby’s
head and face during OVB and the methodology is necessarily pragmatic. However, our group
has extensive experience using simulation models to identify intrapartum forces (10) which have previously enabled the development
of validated and effective training tools (11,12).We used a modified version of the PROMPT Flex® Force Monitoring birthing simulator
(Limbs & Things®, Bristol, UK) with bespoke fetal heads incorporating
pressure sensors. The pressure sensors have previously been employed to quantify pressures
generated using forceps made from novel materials (13). We acknowledge that given the complexities of the birthing process, and the
inherent limitations of any simulation based modeling, our results are unlikely to be
quantifiably reproducible in-vivo. However, the results are likely to be internally
consistent and reflect the location and broad relationships in the pressures exerted by
the BD Odon Device, forceps and Kiwi ventouse.We have not been able to quantify negative pressures, or replicate the chignon associated
with ventouse births. However, due to its mechanism of action we are confident that the BD
Odon Device not generate any negative pressure on the fetal head and therefore will not
cause a chignon. It is therefore unlikely that the most serious outcomes generated by
negative pressure (subgaleal or retinal haemorrhage and cephalohaematoma), or those
associated with movement of the cup over an established chignon (scalp abrasion/avulsion)
will occur following births conducted using the BD Odon Device.
Conclusion
The BD Odon Device generates lower peak pressure than non-rotational forceps during
simulated birth and does not exert a negative pressure required to perform a vacuum assisted
birth. It is therefore likely that the BD Odon Device will be associated with lower adverse
outcomes related to both peak pressure (bruising, facial nerve palsy, skull fracture) and
negative pressure (subgaleal or retinal haemorrhage and cephalohaematoma, scalp
abrasion/avulsion) compared to currently available instruments (forceps and ventouse). This
study has generated sufficient data to suggest that the BD Odon Device is likely to be as
safe, if not safer, than forceps and ventouse. Clinical studies are now required to evaluate
the efficacy of the BD Odon Device.Click here for additional data file.Click here for additional data file.Click here for additional data file.Click here for additional data file.Click here for additional data file.Click here for additional data file.
Authors: David M Sherer; Margarita Sokolovski; Mudar Dalloul; Emil Dib; John C Pezzullo; Joseph A Osho; Ovadia Abulafia Journal: J Ultrasound Med Date: 2007-11 Impact factor: 2.153
Authors: I Gurol-Urganci; D A Cromwell; L C Edozien; T A Mahmood; E J Adams; D H Richmond; A Templeton; J H van der Meulen Journal: BJOG Date: 2013-07-03 Impact factor: 6.531
Authors: J F Crofts; E Lenguerrand; G L Bentham; S Tawfik; H A Claireaux; D Odd; R Fox; T J Draycott Journal: BJOG Date: 2015-02-17 Impact factor: 6.531
Authors: Matthew W Gillman; Janet W Rich-Edwards; Sheryl L Rifas-Shiman; Ellice S Lieberman; Ken P Kleinman; Steven E Lipshultz Journal: J Pediatr Date: 2004-02 Impact factor: 4.406
Authors: Timothy J Draycott; Joanna F Crofts; Jonathan P Ash; Louise V Wilson; Elaine Yard; Thabani Sibanda; Andrew Whitelaw Journal: Obstet Gynecol Date: 2008-07 Impact factor: 7.661
Authors: Ganga L Verma; Jessica J Spalding; Marc D Wilkinson; G Justus Hofmeyr; Valerie Vannevel; Fidelma O'Mahony Journal: Cochrane Database Syst Rev Date: 2021-09-24
Authors: Emily J Hotton; Erik Lenguerrand; Mary Alvarez; Stephen O'Brien; Tim J Draycott; Joanna F Crofts Journal: Am J Obstet Gynecol Date: 2020-12-13 Impact factor: 8.661
Authors: Javier A Schvartzman; Hugo Krupitzki; Mario Merialdi; Ana Pilar Betrán; Jennifer Requejo; My Huong Nguyen; Effy Vayena; Angel E Fiorillo; Enrique C Gadow; Francisco M Vizcaino; Felicitas von Petery; Victoria Marroquin; María Luisa Cafferata; Agustina Mazzoni; Valerie Vannevel; Robert C Pattinson; A Metin Gülmezoglu; Fernando Althabe; Mercedes Bonet Journal: Reprod Health Date: 2018-03-12 Impact factor: 3.223
Authors: Stephen O'Brien; Emily J Hotton; Erik Lenguerrand; Julia Wade; Cathy Winter; Tim J Draycott; Joanna F Crofts Journal: Trials Date: 2019-03-05 Impact factor: 2.279
Authors: Joe Viana; Tone Breines Simonsen; Hildegunn E Faraas; Nina Schmidt; Fredrik A Dahl; Kari Flo Journal: BMC Health Serv Res Date: 2020-02-14 Impact factor: 2.655
Authors: Emily J Hotton; Mary Alvarez; Erik Lenguerrand; Julia Wade; Natalie S Blencowe; Tim J Draycott; Joanna F Crofts Journal: Pilot Feasibility Stud Date: 2021-03-19
Authors: Emily J Hotton; Natalie S Blencowe; Erik Lenguerrand; Tim J Draycott; Joanna F Crofts; Julia Wade Journal: BMJ Open Date: 2021-12-15 Impact factor: 2.692
Authors: Emily J Hotton; Natalie S Blencowe; Nichola Bale; Erik Lenguerrand; Tim J Draycott; Joanna F Crofts; Julia Wade Journal: BMJ Open Date: 2022-08-04 Impact factor: 3.006