Change in Acetabular Cup Orientation From Supine to Standing Position and Its Effect on Wear of Highly Crosslinked Polyethylene.

Matthew G Teeter1, Prateek Goyal2, Xunhua Yuan3, James L Howard2, Brent A Lanting2.   

Abstract

BACKGROUND: The purpose of this study is to measure acetabular cup position and wear of the highly crosslinked polyethylene liner in the supine and standing position for patients at a minimum of 10 years after the operation.
METHODS: A total of 38 patients were recruited at a mean of 12.5 years after the operation. All patients received a single acetabular cup design with a highly crosslinked liner and a 28-mm cobalt-chromium femoral head. Patients underwent supine and standing radiostereometric examinations in which the X-ray sources and detectors were positioned to obtain an anterior-posterior and cross-table lateral radiograph. Acetabular cup position and the three-dimensional wear rate were measured from the radiographs, and outcome scores were recorded for each patient.
RESULTS: Anteversion significantly increased (P < .0001) a mean of 12° from supine (15.1° ± 10.4°) to standing (27.2° ± 10.5°) position. Inclination also significantly increased (P = .001) a mean of 2° from supine (44.4° ± 6.8°) to standing (46.3° ± 7.7°) position. There was no difference (P = .093) in wear rate between supine (0.067 ± 0.070 mm/y) and standing (0.073 ± 0.074 mm/y) positions. There were no correlations between cup orientation and wear rate in either position.
CONCLUSION: Highly crosslinked polyethylene is a forgiving bearing material. Although adherence to the traditional acetabular position target zone is recommended, ensuring hip stability and consideration of the patient's functional position are also important objectives to consider for the acetabular position.
Copyright © 2017 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  acetabular cup position; anteversion angle; inclination angle; polyethylene wear; radiostereometric analysis; total hip replacement

Mesh:

Substances:

Year:  2017        PMID: 28917617     DOI: 10.1016/j.arth.2017.08.016

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  7 in total

Review 1.  Radiographic assessment of the cup orientation after total hip arthroplasty: a literature review.

Authors:  Jing-Xin Zhao; Xiu-Yun Su; Zhe Zhao; Ruo-Xiu Xiao; Li-Cheng Zhang; Pei-Fu Tang
Journal:  Ann Transl Med       Date:  2020-02

2.  [Semi-automated measurement and analysis of three-dimensional acetabular orientation in asymptomatic population and patients of developmental dysplasia of the hip].

Authors:  Xu Jiang; Henghui Zhang; Xumin Hu; Kai Xie; Tianyou Kan; Bo Li; Chi Zhang; Songtao Ai; Liangbin Gao; Mengning Yan; Liao Wang
Journal:  Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi       Date:  2021-12-15

3.  Does accelerometer-based portable navigation provide more accurate and precise cup orientation without prosthetic impingement than conventional total hip arthroplasty? A randomized controlled study.

Authors:  Masato Kiyohara; Satoshi Hamai; Kyohei Shiomoto; Satoru Harada; Tetsunari Harada; Goro Motomura; Satoshi Ikemura; Masanori Fujii; Shinya Kawahara; Yasuharu Nakashima
Journal:  Int J Comput Assist Radiol Surg       Date:  2022-03-26       Impact factor: 2.924

4.  Assessment of Acetabular Component Anteversion after Total Hip Arthroplasty: Comparison of Anteroposterior and Cross-Table Lateral Radiographs with Computed Tomography Scans.

Authors:  Tarun Goyal; Souvik Paul; Arghya Kundu Choudhury; Tushar Gupta
Journal:  Clin Orthop Surg       Date:  2021-08-17

5.  An accelerometer-based navigation system provides acetabular cup orientation accuracy comparable to that of computed tomography-based navigation during total hip arthroplasty in the supine position.

Authors:  Tomonori Tetsunaga; Kazuki Yamada; Tomoko Tetsunaga; Tomoaki Sanki; Yoshi Kawamura; Toshifumi Ozaki
Journal:  J Orthop Surg Res       Date:  2020-04-15       Impact factor: 2.359

Review 6.  Total hip arthroplasty planning.

Authors:  Alessandro Colombi; Daniele Schena; Claudio Carlo Castelli
Journal:  EFORT Open Rev       Date:  2019-11-01

7.  Preoperative factors improving the prediction of the postoperative sagittal orientation of the pelvis in standing position after total hip arthroplasty.

Authors:  Maximilian C M Fischer; Kunihiko Tokunaga; Masashi Okamoto; Juliana Habor; Klaus Radermacher
Journal:  Sci Rep       Date:  2020-09-29       Impact factor: 4.379

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.