| Literature DB >> 28911481 |
Kraingkrai Ponhong1, Sam-Ang Supharoek2, Watsaka Siriangkhawut1, Kate Grudpan3.
Abstract
A simple, rapid, and sensitive spectrophotometric method for the determination of benzoyl peroxide (BPO) in wheat flour samples was developed. The detection principle is based on BPO reacted with 2,2'-azino-bis(3-ethylbenzothiazoline-6-sulfonic acid) (ABTS) to obtain a blue-green colored product that was detected at 415 nm by spectrophotometry. The effect of factors influencing the color reaction was investigated. Under the selected conditions, the linear range for quantification of BPO was observed between 0.2-1.0 mg L-1 with r2 = 0.998. The limit of detection (LOD) was 0.025 mg L-1. The developed method obtained superior precision (relative standard deviation < 2%) using 11 repeatability at 0.2 mg L-1, 0.6 mg L-1, and 0.8 mg L-1. The proposed methodology was successfully applied to determine BPO in wheat flour samples.Entities:
Keywords: ABTS; benzoyl peroxide; colorimetric reaction; wheat flour
Year: 2015 PMID: 28911481 PMCID: PMC9345465 DOI: 10.1016/j.jfda.2015.03.006
Source DB: PubMed Journal: J Food Drug Anal Impact factor: 6.157
Fig. 1The possibility reaction of the proposed system.
Fig. 2Absorption spectra of 1.0 mg L−1 BPO (dotted line), ABTS 10 mg L−1 (broken line) and the blue-green color solution after mixed 1.0 mg L−1 BPO with 10 mg L−1 ABTS (unbroken line). ABTS = 2,2′-azino-bis(3-ethylbenzothiazoline-6-sulfonic acid); BPO = benzoyl peroxide.
Fig. 3Effects of 2,2′-azino-bis(3-ethylbenzothiazoline-6-sulfonic acid) (ABTS) concentration.
Fig. 4Calibration curve graph of the developed method.
Analytical table of merits.
| Linear range (mg L−1) | Linear equation | Correlation coefficient ( | LOD (mg L−1) | %RSD |
|---|---|---|---|---|
| 0.2–1.0 | 0.998 | 0.025 | < 2% |
Abs = absorbance; C = concentration of BPO; LOD = limit of detection; RSD = relative standard deviation.
LOD was calculated by 3σ of blank/slope (σ is standard deviation of blank).
Studied at 0.2 mg L−1, 0.4 mg L−1, and 0.8 mg L−1 of BPO with 11 repeats.
Analytical characteristics of some methods for determination of benzoyl peroxide in wheat flour samples.
| Detection methods | Sample | Working range | LOD (mg L−1) | LOQ (mg L−1) | %RSD | Recovery (%) | Advantage | Disadvantage | Refs |
|---|---|---|---|---|---|---|---|---|---|
| AOAC official method (benzoyl peroxide bleach (benzoic acid) in flour) | Wheat flour | 0–1.2 mg L−1 | — | — | — | — | Simple | Low sensitivity and time-consuming | [ |
| TMB-based colorimetric method | Wheat flour | 0.67–16 mg L−1 | 0.45 | — | 2.68 | 97–118 | Sensitive, convenient and rapid | Batch-wise method and high amount of waste | [ |
| Fluorescence based on | Wheat flour | 0.2–3.2 mg L−1 | 0.06 | — | — | — | Simple and sensitive | Long time analysis (30 min) | [ |
| HPLC | Wheat flour | 0.07–0.15 μg g−1 | 0.03 | — | 0.4–3.2 | 96.0–99.3 | Simple and reliable | Long time analysis (up to 25 min) | [ |
| Differential pulse voltammetry | Wheat flour | 0.61–24.2 mg L−1 | 0.061 | — | — | 94.8–106 | Selective | Using organic solvent | [ |
| Ion chromatography with pre-column derivative | Wheat flour | 0.12–20 mg L−1 | 0.019 | — | — | 81–92 | Sensitive and selective | Long time analysis | [ |
| Chemiluminescence microfluidic chip | Flour | 0.8–100 mg L−1 | 0.4 | — | 2.41–3.06 | — | Less reagent consumption | Complication device and tedious step of operation | [ |
| ABTS-based colorimetric method | Wheat flour | 0.2–1.0 mg L−1 | 0.025 | 0.087 | 0.2–1.7 | 87–104 | Rapid, sensitive and selective | Batch-wise method and high waste generation | This study |
ABTS = 2,2′-azino-bis(3-ethylbenzothiazoline-6-sulfonic acid); AOAC = Association of Official Analytical Chemists; HPLC = high performance liquid chromatography; LOD = limit of detection; LOQ = limit of quantitation; RSD = relative standard deviation; TMB = tetramethylbenzidine.
Recovery and the results for the determination of benzoyl peroxide in wheat flour samples.
| Sample no. | Original amount | Spiked (μg/g) | ABTS method | HPLC |
|---|---|---|---|---|
|
|
| |||
| Found value | Found value | |||
|
|
| |||
| Recovery | Recovery | |||
| 1 | — | 50 | 46.1 ± 0.1 | 44.1 ± 4.2 |
| 92 ± 0.2 | 88 ± 8.4 | |||
| 75 | 72.3 ± 0.2 | 74.9 ± 4.7 | ||
| 96 ± 0.2 | 100 ± 6.3 | |||
| 90 | 80 ± 0.2 | 81.6 ± 4.6 | ||
| 89 ± 0.2 | 91 ± 5.2 | |||
| 2 | — | 50 | 50.9 ± 0.3 | 46.7 ± 4.8 |
| 102 ± 0.5 | 93 ± 9.7 | |||
| 75 | 74.8 ± 0.2 | 71.0 ± 3.9 | ||
| 100 ± 0.3 | 95 ± 5.2 | |||
| 90 | 89.9 ± 0.2 | 88.5 ± 4.0 | ||
| 100 ± 0.3 | 98 ± 4.4 | |||
| 3 | — | 50 | 46.5 ± 0.1 | 49.9 ± 3.8 |
| 93 ± 0.3 | 99 ± 7.6 | |||
| 75 | 69.5 ± 0.2 | 67.3 ± 5.1 | ||
| 93 ± 0.3 | 90 ± 6.8 | |||
| 90 | 83.7 ± 0.3 | 85.5 ± 4.2 | ||
| 93 ± 0.3 | 95 ± 4.7 | |||
| 4 | — | 50 | 45.4 ± 0.1 | 46.6 ± 4.2 |
| 91 ± 0.2 | 93 ± 8.5 | |||
| 75 | 64.9 ± 0.2 | 76.0 ± 4.5 | ||
| 87 ± 0.3 | 101 ± 6.0 | |||
| 90 | 80.9 ± 0.0 | 82.0 ± 4.4 | ||
| 90 ± 0.0 | 91 ± 4.8 | |||
| 5 | — | 50 | 51.8 ± 0.2 | 50.6 ± 4.2 |
| 104 ± 0.3 | 101 ± 8.0 | |||
| 75 | 72.2 ± 0.2 | 74.9 ± 4.4 | ||
| 96 ± 0.3 | 99 ± 5.9 | |||
| 90 | 87.5 ± 0.1 | 87.8 ± 4.6 | ||
| 97 ± 0.1 | 98 ± 5.1 | |||
| 6 | — | 50 | 51.1 ± 0.2 | 54.8 ± 3.9 |
| 102 ± 0.3 | 109 ± 7.7 | |||
| 75 | 68.4 ± 0.0 | 75.3 ± 4.2 | ||
| 91 ± 0.1 | 100 ± 5.7 | |||
| 90 | 85.6 ± 0.4 | 94.0 ± 4.9 | ||
| 95 ± 0.4 | 104 ± 5.4 |
ABTS = 2,2′-azino-bis(3-ethylbenzothiazoline-6-sulfonic acid); HPLC = high performance liquid chromatography.
Not detectable.
Values found are the average of three determinations (n = 3) ± the corresponding standard deviation.
Values found are the average of three determinations (n = 3) ± the corresponding standard deviation.
Limit value for benzoyl peroxide residues in wheat flour stipulated in Codex Alimentarius Commission.
Fig. 5Chromatograms of (A) sample No. 6 without spiked benzoyl peroxide (BPO) standard, (B) BPO standard 50 μg g−1 and (C) sample No. 6 with spiked BPO 90 μg g−1.