| Literature DB >> 28904631 |
N Eddison1, M Mulholland1, N Chockalingam1.
Abstract
OBJECTIVES: The purpose of this article is to determine how many of the current peer-reviewed studies of ankle foot or-thoses (AFOs) on children with cerebral palsy (CP) have included adequate details of the design and material of the AFO, to enable the study to be reproduced and outcomes clearly understood.Entities:
Keywords: AFO, orthoses; cerebral palsy; clinical outcomes; materials testing; orthotic devices
Year: 2017 PMID: 28904631 PMCID: PMC5584494 DOI: 10.1302/1863-2548.11.160256
Source DB: PubMed Journal: J Child Orthop ISSN: 1863-2521 Impact factor: 1.548
Extraction data.
| AFO design | Choose response |
|---|---|
| Is the type of AFO described? | Type of AFO/Incomplete/- |
| Is the AFO bespoke or stock? | Bespoke/Stock/- |
| Is the AAAFO described? | Yes/- |
| Is the manufacturer of the AFO identified? | Complete/- |
| Are the trim lines of the ankle described? | Complete/Incomplete/- |
| Are the trim lines of the foot plate described? | Complete/Incomplete/- |
| Is the height of the AFO described? | Complete/Incomplete/- |
| Is the strapping system described? | Complete/Incomplete/- |
| Is there detail of the stiffness of the AFO in stance phase? | Complete/Incomplete/- |
| If hinges are stipulated are these described? | Complete/Incomplete/- |
| Is there a justification for choosing the AFO design? | Complete/Incomplete/- |
| Is the material described? | Material/- |
| Is the thickness of the material described? | Thickness/- |
| Is there a justification for choosing the material and thickness? | Complete/incomplete/- |
| Has stiffness testing been carried out? | Yes/- |
Complete, all information present; Incomplete, some information missing; N/A, not applicable to this paper; -, all information missing
Data extraction grouped per question.
| Data Extracted: | |
|---|---|
| Is the type of AFO described? | See Fig. |
| Is the AFO bespoke or stock/off the shelf? |
32 papers reportedly used bespoke AFOs 1 paper reportedly used bespoke and stock 1 paper used stock 21 papers did not state |
| Is the AAAFO described? |
15 papers described the AAAFO 10 papers gave incomplete details 30 Papers did not state |
| Is the manufacturer of the AFOs described? |
6 papers detailed the manufacturer 1 paper gave incomplete details 48 papers did not state |
| Are the trim lines of the ankle described? |
10 Papers detailed the trim lines at the ankle 7 papers gave incomplete details 38 papers did not state |
| Is the full design of the footplate described, including length, stiffness and medio-lateral borders? |
0 papers gave a full description of the footplate 21 papers gave an incomplete description of the footplate 34 papers did not state any description of the footplate |
| Is the height of the AFO described? |
14 papers described the height of the AFO 4 papers gave an incomplete description 37 papers did not describe the height of the AFO |
| Is the strapping system described? |
9 papers described the strapping system 5 papers gave an incomplete description 31 papers did not describe the strapping system |
| Is the stiffness of the AFO in the stance phase described? |
9 papers described the stiffness of the AFO in stance phase 6 papers gave an incomplete description 40 papers did not state |
| If hinges are stipulated, are these described? |
9 papers described the hinges on the AFO 2 papers gave an incomplete description 21 papers did not state 23 papers were N/A |
| Is the material described? |
24 papers described the material used 31 papers did not state the material used 20 papers used polypropylene 2 papers used copolymer 1 paper used homopolymer 1 paper used graphite 1 papers used carbon fibre 1 paper gave an incomplete description |
| Is the thickness of the material described? |
40 papers did not state the material thickness 1 paper stated 6 mm 1 paper stated 5 mm 4 papers stated 4.8 mm 1 paper stated 4.5 mm 4 papers stated 4 mm 1 paper stated 3.2 mm 7 papers stated 3 mm 1 paper stated 2.4 mm |
| Is there a justification for choosing the material and thickness? |
2 papers gave a justification for material and or thickness choice 1 paper gave an incomplete description 52 papers did not state |
| Has stiffness testing been carried out? |
2 paper carried out stiffness testing 53 did not state |
| What was the outcome measure of the study? |
25 papers measured lower limb kinetics and kinematics of gait and temporal spatial parameters 10 papers measured energy expenditure 5 papers measured standing balance 4 papers measured functional motor skills 2 papers measured sit to stand transfers 2 papers measured stair locomotion 2 papers measured trunk posture 2 papers measured EMG 3 papers measured knee extension 1 paper measured level walking 1 paper measured ankle function 1 paper measured skin tissue pressure mobility 1 paper measured ankle range of motion 1 paper measured hamstring length 1 paper measured functional balance 1 paper measured ambulation 1 paper measured static foot alignment 1 paper measured lower limb intersegmental co-ordination 2 paper measured gross motor function 1 paper measured pelvis, thorax and arm kinematics |
Fig. 1The type of ankle foot orthoses (AFO) design reported in each study.