Rita Schmidt1,2, Andrew Webb1. 1. Department of Radiology, Leiden University Medical Center , Leiden 2333 ZA, Netherlands. 2. Department of Neurobiology, Weizmann Institute of Science , Rehovot 7610001, Israel.
Abstract
Magnetic resonance imaging and spectroscopy (MRI and MRS) are both widely used techniques in medical diagnostics and research. One of the major thrusts in recent years has been the introduction of ultrahigh-field magnets in order to boost the sensitivity. Several MRI studies have examined further potential improvements in sensitivity using metamaterials, focusing on single frequency applications. However, metamaterials have yet to reach a level that is practical for routine MRI use. In this work, we explore a new metamaterial implementation for MRI, a dual-nuclei resonant structure, which can be used for both proton and heteronuclear magnetic resonance. Our approach combines two configurations, one based on a set of electric dipoles for the low frequency band, and the second based on a set of magnetic dipoles for the high frequency band. We focus on the implementation of a dual-nuclei metamaterial for phosphorus and proton imaging and spectroscopy at an ultrahigh-field strength of 7 T. In vivo scans using this flexible and compact structure show that it locally enhances both the phosphorus and proton transmit and receive sensitivities.
Magnetic resonance imaging and spectroscopy (MRI and MRS) are both widely used techniques in medical diagnostics and research. One of the major thrusts in recent years has been the introduction of ultrahigh-field magnets in order to boost the sensitivity. Several MRI studies have examined further potential improvements in sensitivity using metamaterials, focusing on single frequency applications. However, metamaterials have yet to reach a level that is practical for routine MRI use. In this work, we explore a new metamaterial implementation for MRI, a dual-nuclei resonant structure, which can be used for both proton and heteronuclear magnetic resonance. Our approach combines two configurations, one based on a set of electric dipoles for the low frequency band, and the second based on a set of magnetic dipoles for the high frequency band. We focus on the implementation of a dual-nuclei metamaterial for phosphorus and proton imaging and spectroscopy at an ultrahigh-field strength of 7 T. In vivo scans using this flexible and compact structure show that it locally enhances both the phosphorus and proton transmit and receive sensitivities.
Entities:
Keywords:
dual-band; in vivo; magnetic resonance imaging; metamaterial; ultrahigh field
Magnetic resonance imaging (MRI) is one
of the most important modalities for clinical disease diagnosis, and
it plays a key role in fundamental preclinical research. Its major
advantages include nonionizing radiation, lack of penetration effects,
ability to acquire fully isotropic three-dimensional (3D) data, and
ability to produce a variety of image contrasts between tissues using
different data acquisition parameters. The major disadvantage of MRI
arises from its low sensitivity, which is a result of the small energy
difference between energy levels. Alternative noninductive detection
methods have been explored at very low magnetic fields (≪1
T), including optical atomic magnetometers[1] and superconducting quantum interference devices,[2] but these are not suitable for the much higher field strengths
(1.5–7 T) used in human imaging. Magnetic resonance force microscopy,[3−5] which uses a nanoscale cantilever for mechanical detection of the
MR signal, is a method that has extremely high sensitivity, with the
ability to detect single electronic spins, but it is not applicable
to human imaging because the sample must be placed in a vacuum at
low temperatures.An alternative method to increase the sensitivity
is to incorporate novel materials into the setup: one such example
is a metamaterial. Metamaterials and artificial materials represent
a novel group of structures, creating new means to control wave propagation.[6−9] Applications commenced with optical superlensing,[6,10] continued
with transformation optics[11] and invisibility
cloaking,[12] and presently have incorporated
a wide range of device miniaturization and efficiency enhancements.[13−15] Several studies in MRI have investigated potential improvements
using metamaterials in the detection sensitivity,[16,17] RF transmit efficiency,[18] and decoupling
between the individual coils in a multicoil array.[19,20] In this study, we explore a completely new metamaterial implementation
for MRI, namely, a dual-nuclei resonant structure. We have designed
a hybrid metamaterial, comprising a two-dimensional (2D) metamaterial
surface and a high permittivity dielectric substrate, and explored
the properties controlling the two resonant modes corresponding to
the Larmor frequencies of the two different nuclei. Most of the nonproton
nuclei that are of interest for in vivo acquisition[21] have a Larmor frequency that is far away from that of protons,
due to the very different gyromagnetic ratios. The two nuclei we chose
to focus on in this study are phosphorus[22−24] (with γ,
the gyromagnetic ratio, of 17.4 MHz/T) and proton (γ of 42.58
MHz/T).While initial publications on acquiring a 31P signal date back to 1974,[25,26] phosphorus magnetic
resonance spectroscopy (MRS) is not in wide clinical use due to its
low signal-to-noise ratio (SNR). The low SNR arises from low concentration
and a low gyromagnetic ratio compared to those of 1H, present
in high concentration as water or lipid in tissue (SNR ∝ γ2B0c where B0 is the magnetic field strength and c is the concentration). In vivo 31PMRS acquisition
is always performed in combination with 1H imaging, because
the latter provides localization for the 31P volume-of-interest
and complementary diagnostic information. This requirement is the
motivation behind dual-nuclei excitation and acquisition. Therefore,
we explored a new flexible and compact metamaterial that can be added
as a pad to an existing RF coil setup and augment the SNR for both
the 31P and 1H imaging using a single structure.Several studies have demonstrated metamaterial designs for MRI,
which have included swiss-rolls,[18,27] split-rings,[16,28] wires,[17] and magnetoinductive waveguides.[29] These structures shape the RF magnetic field,
enabling local increases in the MRI transmit and receive efficiency.
Two proposed metamaterials have recently garnered interest. One is
based on a metamaterial with negative permeability,[28] implemented using split-rings, and its feasibility has
recently been demonstrated at 3 T using a thin (11 mm) metamaterial
slab. The second proposal is based on a metamaterial with negative
permittivity,[17] using a metasurface based
on a set of wires or conductive strips: this has been demonstrated
at 1.5[17] and 7 T[30] and showed a local enhancement in SNR. Both designs are based on
a quasi-static approximation applicable for relatively low RF frequencies
relevant for MRI (≲300 MHz). In this approximation, the magnetic
and electric fields can be analyzed separately, and therefore, a system
of either magnetic dipoles (split-rings[28] or short wires that function equivalently to split-rings[31−33]) or electric dipoles (long wires[17]) can
support resonant modes at the frequencies of interest. In our study,
at 7 T, the quasi-static approximation is well-suited for 31P (121 MHz) but is marginal for 1H (298 MHz). Therefore,
full electromagnetic (EM) simulations were used to determine the resonant
modes. Our main aim in this study was to design a new type of metamaterial
with the capability to enhance the signal for both nuclei (31P and 1H), using the same structure. This new feature
arises from a combination of the above two types of metamaterials,
creating a novel dual-nuclei metamaterial.
Design of the Dual-Nuclei
MRI Metamaterial
Several studies have examined metamaterials
for dual-band settings in the IR and microwave ranges using negative
refraction.[34−36] Such structures rely on different characteristics
depending on the RF frequency range, for example, using 3D structures
(with perpendicular setups for dual bands)[36] or a subunit that incorporates dual dimension settings.[34] For the frequency range in the applications
in optics and microwaves, the metamaterials were devised to generate
negative refraction, requiring both negative permittivity and negative
permeability. For the lower frequencies, which are of interest in
our work, implementation based on previous works would result in large
3D structures that are not well-suited to the very restricted space
in the MRI environment. In contrast to previous works, two frequency
bands can be realized based on the same substrate at the frequencies
relevant for MRI: one frequency band produced via coupling with a
set of long strips and the other with a set of short strips. Such
a metamaterial does not require the property of negative refraction.
A set of long strips produces resonant modes for the low frequency
band of interest, and a set of short strips produces resonant modes
for the high frequency band of interest. In this study, the two structures
are harnessed to generate a unique single metamaterial structure for
dual-nuclei purposes (as shown in Figure ). This allows us to generate frequency bands
shifted from each other by a factor of two and more. Such frequency
band shifts are required for MRI dual-nuclei implementations, due
to the very different gyromagnetic ratios of the nuclei of interest
compared to that of protons. The exception for this are 19F and 1H, which are very close to each other (γ
of 19F is 0.94 vs 1H), and therefore can be
implemented with very close resonant systems.[37] In our current study, the gyromagnetic ratio of 31P is
40% of the gyromagnetic ratio of 1H. A new metamaterial
structure and RF field distributions of the resonant modes, relevant
for 31P and 1H, are shown in Figure .
Figure 1
Schematic view of the
hybrid metamaterial pad and its dual-nuclei behavior. (a) Schematic
view of the metamaterial pad combined from subsetups: (top left) 31P nucleus band, tuned to 121 MHz; (top right) 1H nucleus band tuned for 298 MHz. On top of the structure, a central
cross section of the magnetic field |H| is shown (for the combined
structure two maps for two bands are shown). (b) Schematic view of
the setup in MRI including the RF “birdcage” coil and
a metamaterial placed behind the leg. A cross section through the
leg shows the distribution in the central slice. A scaled color map
shows the enhancement ratio distribution for the proton image, and
a contour map on top shows an enhancement ratio for the phosphorus.
In a and b, the 31P maps and the contours are displayed
with a “hot” color map, and the 1H maps are
displayed with a “jet” color map.
Schematic view of the
hybrid metamaterial pad and its dual-nuclei behavior. (a) Schematic
view of the metamaterial pad combined from subsetups: (top left) 31P nucleus band, tuned to 121 MHz; (top right) 1H nucleus band tuned for 298 MHz. On top of the structure, a central
cross section of the magnetic field |H| is shown (for the combined
structure two maps for two bands are shown). (b) Schematic view of
the setup in MRI including the RF “birdcage” coil and
a metamaterial placed behind the leg. A cross section through the
leg shows the distribution in the central slice. A scaled color map
shows the enhancement ratio distribution for the proton image, and
a contour map on top shows an enhancement ratio for the phosphorus.
In a and b, the 31P maps and the contours are displayed
with a “hot” color map, and the 1H maps are
displayed with a “jet” color map.In a previous study, we reduced the dimensions of a hybrid
metamaterial based on copper strips combined with a high dielectric
substrate by using a CaTiO3 suspension.[30] In the current work, we further reduced the metamaterial
dimensions by utilizing zigzag-shaped copper strips instead of straight-line
strips. The zigzag shape enables extension of the electric dipole
to the length required by the desired frequency while maintaining
the physical dimensions of the pad (see Figure S1 that compares a metamaterial based on a set of zigzag-shaped
strips versus straight-line strips).To generate the relevant
set of modes, we used a set of long copper zigzag-shaped strips immersed
in water acting as a dielectric substrate (εr = 80)
as the structure for the 31P frequency (121.7 MHz) at 7
T to produce a compact setup (overall dimensions 14 × 23 ×
1 cm3). The setup for the higher frequency band was designed
using a matrix of short copper strips. The magnetic dipoles are generated
in this setting by the short strips. It has been shown in several
previous works[31−33] that short wires are equivalent to split-rings. In
our case, it is especially advantageous to use the short strips configuration,
because we can integrate them with the long zigzag strips in a similar
manner in one setup. Figures S2.1 and S2.2 summarize the distributions of the H- and E-fields in the main three
planes for both frequencies.Our analysis shows that, although
some parameters such as the dimensions and electric properties of
the dielectric substrate substantially affect both frequencies, there
are also properties that have a major effect on only one frequency
band and a very minor effect on the other. This allows us to independently
control the tuning of each frequency band. Long strip lengths (llong) mainly affect the low frequency band,
with a minor effect on the high frequency. The effect on the high
frequency is due to the effective impedance change of the overall
system. Conversely, short strip lengths (lshort) and the spacing between them affect the high frequency band and
have a minor effect on the low frequency. Figure shows the effect of the parameters that
allow separate mode tuning in the combined structure. Variation of llong in the range of 30–50 cm had a marked
effect (∼35%) on the low frequency band, and a much smaller
effect (∼6%) on the high frequency band. On the other hand,
variation of lshort in the range 4–9
cm had only a very small effect (∼4%) on the low frequency
band, while having a significant effect (∼38%) on the high
frequency band. Another feature worth noting is the similar characteristics
of these resonant modes to transverse electric (TE) mode behavior.
This is especially relevant for MRI since the RF H-field must be perpendicular
to the main static magnetic field (B0)
in order to create the required phase coherence between the spins.
Figure 2
Parametric
dependence of the resonant mode for the low and high bands. (a) Schematic
view of the structure including three long strips in a dielectric
substrate and frequency dependence versus strips length (llong) and thickness of the dielectric (thk). (b) Schematic view of the structure including a matrix of 3 ×
3 short strips in a dielectric substrate and the frequency dependence
versus strips length (lshort) and spacing
(dshort) between the strips. (a+b) Dependence
of the dual-band frequencies in the combined structure as a function
of llong and lshort: blue plot for the low frequency band, and red plot for the high
frequency band. The parameters that were kept constant in all simulations
were dielectric layer width and height (180 mm and 500 mm, respectively)
and the distance between the strips of the same type (40 mm in the x axis). In b and a+b, the structure thickness was set to
1 cm. In the combined setup, for the plot showing the dependence of
the long strips, lshort = 60 mm, and dshort = 20 mm, and for the plot for the short
strips, the llong was set to 500 mm.
Parametric
dependence of the resonant mode for the low and high bands. (a) Schematic
view of the structure including three long strips in a dielectric
substrate and frequency dependence versus strips length (llong) and thickness of the dielectric (thk). (b) Schematic view of the structure including a matrix of 3 ×
3 short strips in a dielectric substrate and the frequency dependence
versus strips length (lshort) and spacing
(dshort) between the strips. (a+b) Dependence
of the dual-band frequencies in the combined structure as a function
of llong and lshort: blue plot for the low frequency band, and red plot for the high
frequency band. The parameters that were kept constant in all simulations
were dielectric layer width and height (180 mm and 500 mm, respectively)
and the distance between the strips of the same type (40 mm in the x axis). In b and a+b, the structure thickness was set to
1 cm. In the combined setup, for the plot showing the dependence of
the long strips, lshort = 60 mm, and dshort = 20 mm, and for the plot for the short
strips, the llong was set to 500 mm.
Results
To optimize the required
penetration depth and the overall distribution, the dual-nuclei configurations
can include different numbers of short and long strips. Figure demonstrates two dual-nuclei
configurations with different penetrations of the 31P band
into tissue. Figure also shows the frequency response of the implemented setups, measured
with a pick-up loop, demonstrating the dual-nuclei bands of interest
as expected in simulations.
Figure 3
Comparison of configurations. (a) Dual-nuclei
with two long strips. (b) Dual-nuclei with three long strips. From
left to right: (left) For each setup, the schematic structure and
the dimensions are shown; (center) Central cross section in the xz plane are displayed for the 31P and 1H frequency bands; (right) plots of the frequency response (measured
with a pick-up loop adjacent to the structure). The 31P
maps are displayed with a “hot” color map, and the 1H maps are displayed with a “jet” color map.
The border of the structure is at z = 5.5 mm.
Comparison of configurations. (a) Dual-nuclei
with two long strips. (b) Dual-nuclei with three long strips. From
left to right: (left) For each setup, the schematic structure and
the dimensions are shown; (center) Central cross section in the xz plane are displayed for the 31P and 1H frequency bands; (right) plots of the frequency response (measured
with a pick-up loop adjacent to the structure). The 31P
maps are displayed with a “hot” color map, and the 1H maps are displayed with a “jet” color map.
The border of the structure is at z = 5.5 mm.Another optimization criterion
is the electric field distribution. The power deposition in a sample
or patient is most commonly quantified via the specific absorption
rate (SAR), which is proportional to the sample conductivity and the
square of the electric field. The maximum intensity of the electric
field is reduced by replacing the straight-line strips by the zigzag-shaped
strips (see Figure S1). This indicates
that further optimization of the electric field distribution is possible,
perhaps by using a snake-shaped dipole antenna as implemented for
ultrahigh-field MRI.[38]One of the
major applications of 31P spectroscopy is the study of
metabolism in muscle physiology of the lower extremities in diseases
such as muscular dystrophy. Therefore, we examined the local enhancement
that can be achieved for imaging the lower extremities. A full EM
simulation is shown in Figure , demonstrating strong local enhancement at both frequencies.
The simulation includes a “birdcage” transmit and receive
RF coil for each of the frequencies, the metamaterial structure shown
in Figure a, and a
human model of the lower extremities. The maximum enhancement of the
RF field, normalized to the square root of the maximum SAR (B1+ / √SAR), is a factor of
2 for 31P and 2.3 for 1H.
Figure 4
EM simulations using
a human body model of the lower extremities. (a) EM simulations at
the 31P frequency with and without metamaterial, placed
under the calf muscle. (b) EM simulations for 1H with and
without metamaterial. Sagittal and axial cross sections through the
calf are shown. The B1+ maps are normalized
to the square root of the maximum local SAR averaged over 10 g. The 31P maps are displayed with a “hot” color map,
and the 1H maps are displayed with a “jet”
color map. A schematic cross-sectional overlay of the metamaterial
(white) is shown in the images.
EM simulations using
a human body model of the lower extremities. (a) EM simulations at
the 31P frequency with and without metamaterial, placed
under the calf muscle. (b) EM simulations for 1H with and
without metamaterial. Sagittal and axial cross sections through the
calf are shown. The B1+ maps are normalized
to the square root of the maximum local SAR averaged over 10 g. The 31P maps are displayed with a “hot” color map,
and the 1H maps are displayed with a “jet”
color map. A schematic cross-sectional overlay of the metamaterial
(white) is shown in the images.The metamaterial was experimentally realized by placing the
copper strips on a very thin plastic substrate, which was placed in
a water pad sealed in a flexible plastic container. The flexibility
of this structure allows close contact with the calf. The experiment
was performed using a transmit/receive double-tuned birdcage coil. 1H scanning comprised a gradient echo sequence with a low-tip
angle excitation. This type of sequence produces images in which the
SNR is proportional to the product of the RF transmit field (B1+) and the complex conjugate of
the receive sensitivity (B1–*) (see the Experimental Section for more
details). A chemical shift imaging (CSI) pure phase encoded method
was used for the 31P spectroscopic imaging. Figure shows experimental imaging
and spectroscopy results from a humancalf muscle, with measured local
enhancement similar to the simulation results (see also Figure S3 for the corresponding phantom results).
The maximum enhancement ratios of the SNR, for the same excitation
tip angle, were calculated from the images as 1.8 for 31P and 2.1 for 1H (the noise was calculated as a standard
deviation in a region placed outside the objects in the image). The
measured signal enhancement was slightly lower than in the simulation
(shown in Figure ),
which might be partially explained by a small local coupling between
the coil and the metamaterial that was not included in the simulations.
Figure 5
In vivo
results. (a) 1H images with and without the metamaterial.
(b) 31P spectra of two size voxels (single voxel with maximal
enhancement and four-combined voxels) with and without the metamaterial.
(c) Phosphocreatine (PCr) image with the metamaterial in place and
a map of the enhancement ratio with the metamaterial compared to without
the metamaterial. The maximum enhancements (for the same excitation
tip angle) were 1.8 and 2.1 for 31P and 1H,
respectively. A schematic cross-sectional overlay of the metamaterial
(white) is shown in the images.
In vivo
results. (a) 1H images with and without the metamaterial.
(b) 31P spectra of two size voxels (single voxel with maximal
enhancement and four-combined voxels) with and without the metamaterial.
(c) Phosphocreatine (PCr) image with the metamaterial in place and
a map of the enhancement ratio with the metamaterial compared to without
the metamaterial. The maximum enhancements (for the same excitation
tip angle) were 1.8 and 2.1 for 31P and 1H,
respectively. A schematic cross-sectional overlay of the metamaterial
(white) is shown in the images.
Discussion and Conclusion
In this study, we have demonstrated
for the first time a metamaterial that can be used to enhance the
sensitivity of two different nuclei in MRI scans. The prototype demonstrates
a local enhancement by combining two substructures, one based on electric
dipoles and the other on magnetic dipoles, and optimally arranging
them for the two frequency bands. Specifically in the demonstrated 31P/1H example, we have shown an enhancement ratio
of approximately a factor of two in the SNR for both nuclei. This
enhancement can be converted into a higher imaging resolution or alternatively
reduced scanning times: the latter consideration is especially important
for 31P spectroscopic imaging that can typically take tens
of minutes. Furthermore, although our focus here was a 31P/1H dual-nuclei metamaterial, this method can be extended
to other important nuclei. Figure S4 shows
a potential design for sodium and proton imaging, which have Larmor
frequencies even further apart (the gyromagnetic ratio of sodium is
0.26 times that of the proton).An additional aspect requiring
analysis is the metamaterial’s finite structure implementation
with a relatively low number of units. Considering the dielectric
environment of the modes of interest in this study, the distances
between the strips are in the range λ/6 to λ/8 (where
λ is the effective wavelength). Thus, although the subunits
are not orders of magnitude smaller than the wavelength, they are
small enough in order to produce metamaterial properties. To examine
the resonant mode with the low number of units as shown in Figure , we performed a
series of simulations varying the number of strips. Figures S2.1–S2.3 show that the resonant mode field
distributions for both frequency bands are maintained for the low
number of units actually used compared to the higher number of units,
with a slight shift of the modal frequency. Local peaks in the fields
close to the metamaterial surface can be seen in Figure . These arise from two different
sources, the relatively low number of subunits and local coupling
between the long and short strips. This local inhomogeneity can be
reduced by a higher density of the subunits (as shown in Figures S2.1–S2.3). However, it is important
to note that, because the effect takes place near the structure and
the relevant imaging area is 1–2 cm from the structure, the
presence of these small peaks has negligible impact on the imaging
performance in the region of interest.In addition to conventional
methods of improving the sensitivity of the RF coils, there is growing
interest in exploring complementary methods that utilize the EM properties
of materials for signal enhancement.[39,40] For example,
passive pads based on a suspension of Perovskite materials (CaTiO3 and BaTiO3) with relative permittivities (εr) of 100–300 have been shown to locally increase the
SNR.[41−43] The convenience of passive dielectric pads is that
they are simply added to the existing RF coil setup. However, significant
improvements for 31PMRS require very high relative permittivity
values, in the range 500–1000 or very bulky pads,[44−46] both of which make it difficult to integrate into a clinical setup.
In this work, our goal was a new flexible and compact structure that
can provide patient comfort and ease of use in a realistic environment;
that is, it can be integrated into closely fitting receive arrays.
The flexibility of the metamaterial also provides an additional advantage,
which is the close contact to the region-of-interest, maximizing the
available SNR. The approach introduced in this work is a further step
toward new design principles of metamaterials for manipulating RF
magnetic and electric fields. Such manipulation can be optimized for
different goals in MRI, such as enhancing the local signal (as shown
in this work), reducing SAR (due to a reduced electric field), or
improving signal homogeneity, especially for ultrahigh-field MRI.
In the current implementation, we used the metamaterial as a passive
pad; however, it could also be implemented as an active resonator
and serve as a building block in future RF coil design.
Experimental Section
Analysis and Simulations
3D EM simulations
were performed using finite integration technique (FIT) software (CST
Microwave Studio, Darmstadt, Germany). The first step used the eigenmode
solver to analyze the spatial distribution and the resonant frequencies
of the relevant mode. The EM mode used for MRI must have an H-field
component perpendicular to the main static magnetic field and should
propagate into the object being imaged. We choose here a “TE01-like” mode that satisfies the first criterion and
is also the mode with the deepest penetration. Simulations included
analysis of the frequency dependence of the modes as a function of
the distance between the short and long strips and the thickness of
the dielectric substrate. Water was chosen as the dielectric substrate. Figures S1 and S2 show full details of the mode
characterization.The RF transmit field is defined as a left
circularly polarized field transverse to the main B0 magnetic field: the transmit field B1+ = (B1 + B1) / 2 and
the receive field B1– = (B1 – B1) / 2. The magnitude of the
RF transmit field defines the excitation tip angle θ applied
to the spins in the excited volume, θ = γB1τ where γ is
the gyromagnetic ratio and τ is the pulse duration. The SNR
of the image is proportional to sin(θ) • B1–* / √P, where P is the accepted power of the coil. All RF transmit (B1+) maps were normalized to an accepted
power of 1 W. The simulations were performed separately for 31P and 1H with a setup that included 8-rung high-pass quadrature
birdcage coils for 31P and 1H (i.d. = 19 cm
and rung length = 14 cm for 31P; i.d. = 19.5 cm and rung
length = 14 cm for 1H), corresponding to the transmit coil
used for the experimental measurements.[30] The coil was loaded with the human model data set “Gustav”
provided by the CST software. The mesh resolution was 2.0 × 2.0
× 2.0 mm3. The simulations compared the RF field distribution
with and without the metamaterial positioned below the calf region.
Characterization of Materials
The metamaterial prototype
included two long strips of 26 cm in total length and a 3 × 3
matrix of short strips (6 cm length) (as shown in Figure a). The distance between the
strips was 2.5 cm. The implemented structure was constructed from
25 μm thick copper strips, 5 mm wide each. The full structure
size was 14 × 23 × 1.1 cm3 including a 1.1 cm
thick water layer. The dielectric layer and the copper strips were
sealed in a plastic container. The pad was placed under the calf muscle
of the leg to assess the local enhancement of the prototype.
Phantom
and in Vivo Experiments
Phantom and in vivo images of a volunteer
were acquired on a Philips Achieva 7 T MRI system. The phantom setup
consisted of a plastic cylinder containing polyvinylpyrrolidone (∼1.8
M) in water and salt added to achieve dielectric properties of εr = 48 and σ = 0.3 S/m (in order to mimic the electric
properties of calf muscle), in which 500 mM phosphoric acid was dissolved.All experimental protocols were approved by the Leiden University
Medical Centre Medical Ethics Committee, and all methods were carried
out in accordance with Leiden University Medical Centre guidelines
and regulations. Phantom and in vivo scans were acquired using custom-built
double-resonant birdcage coils reported in ref (30). The 1H images
were produced by a low-tip-angle gradient-echo sequence with the following
scan parameters: field-of-view (FOV) = 24 × 24 cm2, spatial resolution = 1.5 × 1.5 × 5.0 mm3,
TR/TE = 10/3.4 ms, and flip angle = 5°. The 31P spectroscopic
imaging consisted of a nonselective 2D CSI sequence. The CSI scan
parameters for the phantom were as follows: FOV = 13 × 13 cm2, phase encoding matrix = 13 × 13, flip angle = 45°,
echo time = 1.2 ms, repetition time = 2000 ms, and number of averages
= 2. For the in vivo scans, the parameters were as follows: FOV =
20 × 20 cm2, phase encoding matrix = 10 × 10,
flip angle = 45°, echo time = 1.3 ms, repetition time = 2000
ms, and number of averages = 12.
Authors: Shoujun Xu; Valeriy V Yashchuk; Marcus H Donaldson; Simon M Rochester; Dmitry Budker; Alexander Pines Journal: Proc Natl Acad Sci U S A Date: 2006-08-02 Impact factor: 11.205
Authors: Anna A Hurshkainen; Tatyana A Derzhavskaya; Stanislav B Glybovski; Ingmar J Voogt; Irina V Melchakova; Cornelis A T van den Berg; Alexander J E Raaijmakers Journal: J Magn Reson Date: 2016-05-26 Impact factor: 2.229