| Literature DB >> 28890606 |
Alberto Di Blasio1, Chiara Di Blasio2, Giuseppe Pedrazzi3, Diana Cassi4, Marisabel Magnifico1, Edoardo Manfredi1, Mauro Gandolfini1.
Abstract
OBJECTIVE: This study was performed to evaluate the feasibility of noninvasive measurement of the ANB angle using photographic and ultrasonographic methods.Entities:
Keywords: Cephalometry; Dimensional measurement accuracy; Orthodontics; Radiation protection; Ultrasonography
Year: 2017 PMID: 28890606 PMCID: PMC5570769 DOI: 10.1007/s11282-017-0275-y
Source DB: PubMed Journal: Oral Radiol ISSN: 0911-6028 Impact factor: 1.852
Fig. 1Ultrasonographic soft tissue thickness over N cephalometric point
Fig. 2Ultrasonographic soft tissue thickness over A cephalometric point
Fig. 3Ultrasonographic soft tissue thickness over B cephalometric point
Fig. 4General electric LOGIQ P6 ultrasound system
Fig. 5Ultrasonographic soft tissue thickness measurement
Fig. 6Ultrasonographic soft tissue thickness over N, A, and B cephalometric points marked on a 1:1 lateral photograph. ANB angle construction and measurement
Comparison of ultrasonographic measurements between the two ultrasonographers
| Measurement point | Ultrasonographer 1, mean (SD) (mm) | Ultrasonographer 2, mean (SD) (mm) | Mean diff. (mm) |
|
|---|---|---|---|---|
| Point N | 5.31 (1.14) | 5.36 (1.01) | 0.05 | 0.53 |
| Point A | 10.87 (0.98) | 10.69 (0.98) | 0.18 | 0.37 |
| Point B | 10.02 (1.40) | 10.17 (1.25) | 0.15 | 0.59 |
SD standard deviation, diff. difference
*Paired t test. Wilcoxon’s test (data not reported) provided similar results
Correlation coefficients between the two ultrasonographers
| Pearson’s |
| Spearman’s |
| ICC (95% CI) | CCC (95% CI) | |
|---|---|---|---|---|---|---|
| Ultra 1, N |
| 0.000 |
| 0.000 |
|
|
| Ultra 1, A |
| 0.005 |
| 0.003 |
|
|
| Ultra 1, B |
| 0.005 |
| 0.047 |
|
|
Ultra 1 Ultrasonographer 1, Ultra 2 Ultrasonographer 2, ICC intraclass correlation coefficient, CCC Lin’s concordance correlation coefficient, CI confidence interval
Bold values indicate the Pearson's and Spearman's correlation coefficients between the measurement of the two operators overlying the N, A and B points. There is also segnalated the intraclass correlation coefficient and the Lin's concordance correlation coefficients using 95% as the confidence interval
Comparison of ultrasonographic versus radiographic measurements of the soft tissue thickness covering the N, A, and B bony points
| Measurement point | Ultrasonography, mean (SD) (mm) | X-ray, mean (SD) (mm) | Mean diff. (mm) |
|
|---|---|---|---|---|
| Point N | 5.33 (1.07) | 5.58 (1.00) | 0.24 | 0.55 |
| Point A | 10.78 (0.94) | 11.22 (0.94) | 0.44 | 0.24 |
| Point B | 10.09 (1.27) | 9.39 (1.24) | 0.70 | 0.20 |
SD standard deviation, diff. difference
*t test. Wilcoxon’s test (data not reported) provided similar results
Fig. 7Most of the single measurements showed a difference of ≤20%
Correlation coefficients of ultrasonographic versus radiographic measurements of the soft tissue thickness covering the N, A, and B bony points
| X-ray versus ultrasonography | Pearson’s |
| Spearman’s |
| ICC (95% CI) | CCC (95% CI) |
|---|---|---|---|---|---|---|
| Point N |
| 0.336 |
| 0.343 |
|
|
| Point A |
| 0.316 |
| 0.195 |
|
|
| Point B |
| 0.472 |
| 0.861 |
|
|
ICC intraclass correlation coefficient, CCC Lin’s concordance correlation coefficient, CI confidence interval
Bold values indicate the Pearson's and Spearman's correlation coefficients between the measurement of the two operators overlying the N, A and B points. There is also segnalated the intraclass correlation coefficient and the Lin's concordance correlation coefficients using 95% as the confidence interval
Comparison of ultrasonographic versus radiographic measurements of the ANB angle
| Measurement point | Ultrasonography, mean (SD) (mm) | X-ray, mean (SD) (mm) | Mean diff. (mm) |
|
|---|---|---|---|---|
| ANB angle | 6.00 (2.83) | 6.06 (2.74) | 0.06 | 0.76 |
Wilcoxon’s test (data not reported) provided similar results
SD standard deviation, diff. difference
Correlation coefficients of ultrasonographic versus radiographic measurements of the ANB angle
| X-ray versus ultrasonography | Pearson’s |
| Spearman |
| ICC (95% CI) | CCC (95% CI) |
|---|---|---|---|---|---|---|
| ANB |
| 0.000 |
| 0.000 |
|
|
ICC intraclass correlation coefficient, CCC Lin’s concordance correlation coefficient, CI confidence interval
Bold values indicate the Pearson's and Spearman's correlation coefficients between the measurement of the two operators overlying the N, A and B points. There is also segnalated the intraclass correlation coefficient and the Lin's concordance correlation coefficients using 95% as the confidence interval