Literature DB >> 28870716

Endovascular baroreflex amplification for resistant hypertension: a safety and proof-of-principle clinical study.

Wilko Spiering1, Bryan Williams2, Jan Van der Heyden3, Monique van Kleef4, Rob Lo5, Jorie Versmissen6, Adriaan Moelker7, Abraham Kroon8, Hannes Reuter9, Gary Ansel10, Gregg W Stone11, Mark Bates12.   

Abstract

BACKGROUND: Carotid baroreflex activation lowers blood pressure and might have potential application for the treatment of resistant hypertension. We did a proof-of-principle trial with a novel endovascular baroreceptor amplification device, MobiusHD (Vascular Dynamics, Mountain View, CA, USA), in patients with resistant hypertension.
METHODS: CALM-FIM_EUR was a prospective, first-in-human, open-label study done at six European centres. Eligible patients were adults with resistant hypertension (office systolic blood pressure ≥160 mm Hg despite taking at least three antihypertensive agents, including a diuretic). MobiusHD devices were implanted unilaterally in the internal carotid artery. The primary endpoint was the incidence of serious adverse events at 6 months. Secondary endpoints included changes in office and 24 h ambulatory blood pressure. This trial is registered with ClinicalTrials.gov, number NCT01911897.
FINDINGS: Between December, 2013, and February, 2016, 30 patients were enrolled and underwent successful implantation. Mean age was 52 years (SD 12), 15 patients (50%) were men, and mean antihypertensive use was 4·4 drugs (1·4). Mean office blood pressure was 184/109 mm Hg (18/14) at baseline and was reduced by 24/12 mm Hg (13-34/6-18) at 6 months (p=0·0003 for systolic and p=0·0001 diastolic blood pressure). Mean baseline 24 h ambulatory blood pressure was 166/100 mm Hg (17/14) at baseline and was reduced by 21/12 mm Hg (14-29/7-16) at 6 months (p<0·0001 for systolic and diastolic blood pressure). Five serious adverse events had occurred in four patients (13%) at 6 months: hypotension (n=2), worsening hypertension (n=1), intermittent claudication (n=1) and wound infection (n=1).
INTERPRETATION: In patients with resistant hypertension, endovascular baroreceptor amplification with the MobiusHD device substantially lowered blood pressure with an acceptable safety profile. Randomised, double-blind, sham-controlled trials are warranted to investigate the use of this treatment further. FUNDING: Vascular Dynamics.
Copyright © 2017 Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2017        PMID: 28870716     DOI: 10.1016/S0140-6736(17)32337-1

Source DB:  PubMed          Journal:  Lancet        ISSN: 0140-6736            Impact factor:   79.321


  25 in total

Review 1.  Hypertension: history and development of established and novel treatments.

Authors:  Milan Wolf; Sebastian Ewen; Felix Mahfoud; Michael Böhm
Journal:  Clin Res Cardiol       Date:  2018-06-27       Impact factor: 5.460

2.  Renal Denervation to Treat Heart Failure.

Authors:  Thomas E Sharp; David J Lefer
Journal:  Annu Rev Physiol       Date:  2020-10-19       Impact factor: 19.318

3.  [Therapy-resistant and therapy-refractory arterial hypertension].

Authors:  M Wallbach; M J Koziolek
Journal:  Internist (Berl)       Date:  2018-06       Impact factor: 0.743

Review 4.  Does treatment-resistant hypertension exist in children? A review of the evidence.

Authors:  Ian Macumber; Joseph T Flynn
Journal:  Pediatr Nephrol       Date:  2019-05-30       Impact factor: 3.714

5.  Hypertension: Device therapy for uncontrolled hypertension: new approaches to an old problem.

Authors:  Peter J Blankestijn; Michiel L Bots
Journal:  Nat Rev Nephrol       Date:  2017-11-15       Impact factor: 28.314

Review 6.  Device-based therapies for arterial hypertension.

Authors:  Lucas Lauder; Michel Azizi; Ajay J Kirtane; Michael Böhm; Felix Mahfoud
Journal:  Nat Rev Cardiol       Date:  2020-04-14       Impact factor: 32.419

7.  Resistant Hypertension: Detection, Evaluation, and Management: A Scientific Statement From the American Heart Association.

Authors:  Robert M Carey; David A Calhoun; George L Bakris; Robert D Brook; Stacie L Daugherty; Cheryl R Dennison-Himmelfarb; Brent M Egan; John M Flack; Samuel S Gidding; Eric Judd; Daniel T Lackland; Cheryl L Laffer; Christopher Newton-Cheh; Steven M Smith; Sandra J Taler; Stephen C Textor; Tanya N Turan; William B White
Journal:  Hypertension       Date:  2018-11       Impact factor: 10.190

Review 8.  Resistant Hypertension Updated Guidelines.

Authors:  Irene Chernova; Namrata Krishnan
Journal:  Curr Cardiol Rep       Date:  2019-08-30       Impact factor: 2.931

Review 9.  Device profile of the MobiusHD EVBA system for the treatment of resistant hypertension: overview of its mechanism of action, safety and efficacy.

Authors:  Mark C Bates; Gregg W Stone; Chao-Yin Chen; Wilko Spiering
Journal:  Expert Rev Med Devices       Date:  2020-06-16       Impact factor: 3.166

Review 10.  Contemporary Strategies to Manage High Blood Pressure in Patients with Coexistent Resistant Hypertension and Heart Failure With Reduced Ejection Fraction.

Authors:  Katherine Lang; Erik H Van Iterson; Luke J Laffin
Journal:  Cardiol Ther       Date:  2020-11-17
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.