Tomoyuki Kabutoya1, Yasushi Imai1, Satoshi Hoshide1, Kazuomi Kario1.
Abstract
The authors evaluated a new algorithm for detecting atrial fibrillation (AF) using a home blood pressure monitor. Three serial blood pressure values were measured by the monitor in 16 patients with AF and 20 patients with sinus rhythm. The authors defined "monitor AF in irregular pulse peak (IPP) 25" as follows: (1) IPP: |interval of pulse peak - the average of the interval of the pulse peak| ≥ the average of the interval of the pulse peak ×25%; (2) irregular heart beat: beats of IPP ≥ total pulse ×20%; and (3) monitor AF: two or more irregular heart beats of the three blood pressure measurements. Cutoff IPP values were set at 20% (IPP20) and 15% (IPP15). The monitor's AF specificity was 1.0 in IPP25, IPP20, and IPP15, and its sensitivity was 0.88 in IPP25, 0.94 in IPP20, and 1.0 in IPP15. The new algorithm had high diagnostic accuracy for detecting AF and a low false-positive rate. ©2017 Wiley Periodicals, Inc.Entities:
Keywords: atrial fibrillation; blood pressure monitor; hypertension
Mesh:
Year: 2017 PMID: 28861938 PMCID: PMC8030935 DOI: 10.1111/jch.13076