| Literature DB >> 28861029 |
Edwin Hong-Teck Loh1, Yi-Ting Ong1,2, Narayanaswamy Venketasubramanian3,4, Saima Hilal3, Naing Thet5, Tien Yin Wong1,2,5, Christopher P L Chen3,6, Carol Yim-Lui Cheung1,7.
Abstract
BACKGROUND: With increasing interest in determining if measurement of retinal neuronal structure with spectral-domain optical coherence tomography (SD-OCT) is useful in accessing neurodegenerative process in cognitive decline and development of dementia, it is important to evaluate whether the SD-OCT measurements are repeatable and reproducible in these patients.Entities:
Keywords: dementia; mild cognitive impairment; optical coherence tomography; repeatability; reproducibility; retina
Year: 2017 PMID: 28861029 PMCID: PMC5559462 DOI: 10.3389/fneur.2017.00359
Source DB: PubMed Journal: Front Neurol ISSN: 1664-2295 Impact factor: 4.003
Figure 1Flowchart of Participants for the overall assessment.
Baseline characteristics of study subjects.
| Baseline characteristics | All subjects ( | Alzheimer’s disease ( | Mild cognitive impairment ( |
|---|---|---|---|
| Age, mean (SD) | 73.1 (8.3) | 76.3 (5.5) | 69.7 (9.5) |
| Male, | 28 (45.9) | 13 (40.6) | 15 (51.7) |
| Chinese ethnicity, | 45 (73.8) | 25 (78.1) | 20 (69.0) |
| Hypertension, | 42 (68.9) | 27 (84.4) | 15 (51.7) |
| Diabetes, | 23 (37.7) | 12 (37.5) | 11 (37.9) |
| Hypercholesterolemia, | 45 (73.8) | 22 (68.8) | 23 (79.3) |
| Clinical dementia rating (CDR) global = 0.5, | 29 (47.5) | 0 (0.0) | 29 (100) |
| CDR global = 1, | 25 (41.0) | 25 (78.1) | 0 (0.0) |
| CDR global = 2, | 7 (11.5) | 7 (21.9) | 0 (0.0) |
Distributions of macular ganglion cell-inner plexiform layer (GC-IPL) and peripapillary retinal nerve fiber layer (RNFL) thicknesses.
| Mean (SD) | Baseline | 6-Month follow-up | 1-Year follow-up | |
|---|---|---|---|---|
| GC-IPL average thickness | 77.44 (9.56) | 77.37 (9.93) | 77.13 (9.19) | 0.985 |
| GC-IPL minimum thickness | 72.54 (12.46) | 72.38 (12.67) | 72.13 (12.42) | 0.986 |
| GC-IPL temporal superior | 77.04 (10.78) | 76.79 (11.13) | 76.58 (9.71) | 0.976 |
| GC-IPL superior | 78.42 (8.89) | 78.17 (9.71) | 77.75 (8.94) | 0.932 |
| GC-IPL nasal superior | 79.37 (10.12) | 79.15 (10.14) | 79.10 (9.80) | 0.990 |
| GC-IPL nasal inferior | 76.85 (10.23) | 76.77 (10.17) | 77.10 (10.20) | 0.985 |
| GC-IPL inferior | 75.42 (10.75) | 75.52 (10.87) | 75.31 (10.11) | 0.995 |
| GC-IPL temporal inferior | 77.38 (10.55) | 77.48 (11.26) | 76.67 (10.28) | 0.915 |
| RNFL average thickness | 90.77 (12.48) | 90.75 (12.74) | 89.54 (12.94) | 0.810 |
| RNFL temporal quadrant | 65.87 (12.57) | 66.19 (13.49) | 65.88 (13.16) | 0.987 |
| RNFL superior quadrant | 110.77 (18.65) | 110.20 (17.62) | 108.45 (18.17) | 0.737 |
| RNFL nasal quadrant | 69.06 (10.75) | 69.36 (10.13) | 68.17 (10.76) | 0.789 |
| RNFL inferior quadrant | 117.54 (24.38) | 117.23 (24.80) | 115.54 (24.78) | 0.877 |
Intra-visit repeatability and inter-visit reproducibility of macular ganglion cell-inner plexiform layer (GC-IPL) measurement.
| Parameters | Coefficient of variation (%) | ICC |
|---|---|---|
| GC-IPL average thickness | 1.14 | 0.985 (0.976–0.991) |
| GC-IPL minimum thickness | 2.12 | 0.962 (0.939–0.976) |
| GC-IPL superotemporal | 1.49 | 0.983 (0.973–0.989) |
| GC-IPL superior | 1.96 | 0.962 (0.940–0.976) |
| GC-IPL superonasal | 1.51 | 0.974 (0.959–0.984) |
| GC-IPL inferonasal | 2.33 | 0.948 (0.918–0.967) |
| GC-IPL inferior | 2.40 | 0.954 (0.928–0.971) |
| GC-IPL inferotemporal | 1.56 | 0.983 (0.972–0.989) |
| GC-IPL average thickness | 1.24 | 0.985 (0.977–0.991) |
| GC-IPL minimum thickness | 1.79 | 0.983 (0.974–0.990) |
| GC-IPL superotemporal | 1.96 | 0.960 (0.938–0.976) |
| GC-IPL superior | 2.32 | 0.941 (0.909–0.964) |
| GC-IPL superonasal | 1.74 | 0.977 (0.965–0.986) |
| GC-IPL inferonasal | 2.12 | 0.960 (0.937–0.975) |
| GC-IPL inferior | 2.25 | 0.966 (0.946–0.979) |
| GC-IPL inferotemporal | 1.86 | 0.975 (0.960–0.985) |
Intra-visit repeatability and inter-visit reproducibility of peripapillary retinal nerve fiber layer (RNFL) measurement.
| Parameters | Coefficient of variation (%) | ICC |
|---|---|---|
| RNFL average thickness | 1.34 | 0.986 (0.978–0.991) |
| RNFL temporal quadrant | 2.51 | 0.951 (0.924–0.969) |
| RNFL superior quadrant | 2.40 | 0.972 (0.956–0.982) |
| RNFL nasal quadrant | 2.90 | 0.937 (0.904–0.960) |
| RNFL inferior quadrant | 2.48 | 0.977 (0.964–0.985) |
| RNFL average thickness | 2.71 | 0.952 (0.928–0.968) |
| RNFL temporal quadrant | 3.38 | 0.961 (0.942–0.974) |
| RNFL superior quadrant | 3.84 | 0.922 (0.885–0.948) |
| RNFL nasal quadrant | 5.25 | 0.845 (0.781–0.896) |
| RNFL inferior quadrant | 3.97 | 0.955 (0.934–0.971) |
Figure 2Bland–Altman plot of intra-visit peripapillary retinal nerve fiber layer (RNFL) thickness.
Figure 5Bland–Altman plot of inter-visit macular ganglion cell-inner plexiform layer (GC-IPL) average thickness.