PURPOSE: To study the accuracy and reliability of optical pachymetry using the Alcon WaveLight EX500 during laser-assisted in situ keratomileusis (LASIK).Entities:
Keywords: femtosecond laser; flap thickness; optical coherence tomography
Year: 2017 PMID: 28860701 PMCID: PMC5566874 DOI: 10.2147/OPTH.S138459
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Figure 1High-definition optical coherence tomography at center of cornea.
Note: Top caliper measures flap thickness at 132 μm and bottom caliper measures residual stromal bed thickness at 346 μm.
Abbreviations: T, temporal; N, nasal.
Figure 2High-definition optical coherence tomography measurements.
Notes: Example of measurements across the cornea. Five measurements are spaced approximately 1.5 mm apart from each another, with the center measurement at the center of the cornea. Top calipers measure flap thickness. Bottom calipers measure residual stromal bed thickness.
Abbreviations: T, temporal; N, nasal.
Timing of OCT measurements
| Months after surgery | Eyes, n | Percentage |
|---|---|---|
| 1 | 24 | 53% |
| 2–3 | 7 | 16% |
| 4–6 | 6 | 13% |
| 14–17 | 8 | 18% |
| Total | 45 | 100% |
Abbreviation: OCT, optical coherence tomography.
Flap- and RSB-thickness comparison
| Measure | Mean (mm) | 95% CI | |
|---|---|---|---|
| FT difference: FS200 – EX500 | 9 | 6–12 | <0.001 |
| FT difference: FS200 – OCT | −0.2 | −2 to 2 | 0.87 |
| RSB difference: EX500 (programmed) – EX500 | 15 | 8–22 | <0.001 |
| RSB difference: EX500 (programmed) – OCT | −0.3 | −5 to 4 | 0.89 |
Note: Comparison of difference in flap and RSB measurements between FS200 and EX500 using OCT.
Abbreviations: FT, flap thickness; OCT, optical coherence tomography; RSB, residual stromal bed.
Summary of overall clinical outcome of LASIK patients
| Measure | n | Mean | SD | Minimum | Maximum |
|---|---|---|---|---|---|
| UCVA: logMAR (Snellen) | 88 | −0.024 (20/19) | 0.17 | −0.165 (20/15 +2) | 0.66 (20/80 −1) |
| BCVA: logMAR (Snellen) | 71 | −0.08 (20/17) | 0.07 | −0.125 (20/15) | 0.14 (20/25 −2) |
| Spherical equivalent | 79 | −0.18 | 0.44 | −2 | 0.5 |
| Sphere | 77 | −0.26 | 0.24 | 0 | 1 |
| Cylinder | 77 | 0.15 | 0.45 | −2 | 0.25 |
Note:
Number of individual eyes measured.
Abbreviations: LASIK, laser-assisted in situ keratomileusis; UCVA, uncorrected visual acuity; BCVA, best-corrected visual acuity; logMAR, logarithm of minimum angle of resolution.