Misaligned pCONus Device: Case Report of a Unique Complication.

Krishna Prasad Bellam Premnath1, Gigy Kuruttukulam1, Anil Kumar Shivan1.   

Abstract

pCONus is a stent-like endovascular device which aids in retention of coils within wide-necked bifurcation aneurysms. It is retrievable even after complete deployment and is detached electrolytically. The pCONus aided coiling of wide-necked bifurcation aneurysms has a high technical success rate and a good safety profile. Different complications have been described in literature with the usage of pCONus. This case report describes a unreported complication of inappropriately deployed pCONus device.

Entities:  

Keywords:  Aneurysm; endovascular coiling; pCONus device; waffle-cone technique

Year:  2017        PMID: 28828203      PMCID: PMC5545776          DOI: 10.4103/jcis.JCIS_36_17

Source DB:  PubMed          Journal:  J Clin Imaging Sci        ISSN: 2156-5597


INTRODUCTION

Treatment of wide-necked bifurcation aneurysms incorporating part of parent vessel has always been difficult.[12345] pCONus device is an improvised stent made for coiling wide-necked bifurcation aneurysms using the “waffle-cone” technique.[2] Studies on pCONus-assisted coiling have demonstrated a high technical success and a desirable safety profile.[14] Various technical and long-term complications have been observed.[145] Technical complication of a inappropriately deployed pCONus device has never been described.

CASE REPORT

A 64-year-old female came with complaints of sudden onset severe headache which had started 3 days earlier. There was no neurological deficit. Plain computed tomography (CT) scan of the brain showed Fisher Grade 1 subarachnoid hemorrhage in the right sylvian fissure. Digital subtraction angiography through a right femoral access showed a wide-necked middle cerebral artery bifurcation aneurysm measuring 7.3 mm × 5.8 mm, a neck measuring 4 mm [Figure 1]. A decision to treat the aneurysm endovascularly with pCONus device assistance was made after an interdisciplinary discussion. Under adequate heparinization, the right femoral angiographic sheath was exchanged over a wire for a 7 French long sheath, and a 0.070” 6F Neuron guiding catheter (Penumbra Inc., Alameda, California, USA) was placed coaxially with its tip in the upper cervical segment of internal carotid artery. A Vasco 0.021” microcatheter (Balt, Montmorency, France) was placed through the guiding catheter with its tip in the center of an aneurysm. A 20 mm long pCONus device with 6 mm crown was delivered through the microcatheter, the crown was deployed within the aneurysm, and the microcatheter was pulled back to place the crown at the base of the aneurysm covering the neck. Four tablets of 75 mg prasugrel were administered through a nasogastric tube at this point. The rest of the device was deployed in the middle cerebral artery by unsheathing it. After unsheathing the device, there was prolapse of one of the crown petals into the inferior division of middle cerebral artery [Figure 2]. Repeated attempts at re-sheathing the device with the aim to reposition it by two operators failed. It was decided to proceed with coiling of the aneurysm with this misaligned position of the pCONus device rather than retrieve the device in its deployed state. A Vasco 0.010 inch microcatheter (Balt, Montmorency, France) was used to intubate the aneurysm through the pCONus device, and the aneurysm was coiled progressively using multiple coils (a three-dimensional (3D) coil followed by soft coils) with frequent repositioning of prolapsing coil loops. After the aneurysm was near totally filled and there was minimal residual filling, further coil loops prolapsed into the parent vessel relentlessly on repeated trying and dislodged a loop of an already placed coil into the parent vessel. Further attempts were hence abandoned. After the procedure, there was minimal residual filling of an aneurysm, and good flow in the middle cerebral artery and its branches [Figure 3]. Nonflow-limiting dissection was seen in the distal internal carotid artery. There was no clinical complication, and the patient was discharged 5 days after the procedure. The patient was kept on dual antiplatelets (prasugrel and aspirin) for 3 months and single antiplatelet (aspirin) thereafter.
Figure 1

A 64-year-old woman with a ruptured wide-necked bifurcation aneurysm of the right middle cerebral artery who presented with sudden onset severe headache. Digital subtraction angiography image taken during right internal carotid artery contrast injection showing a wide-necked right middle cerebral artery bifurcation aneurysm (arrow). Note mild irregularity of the distal cervical segment of internal carotid artery due to vasospasm (arrowhead).

Figure 2

A 64-year-old woman with a ruptured wide-necked bifurcation aneurysm of the right middle cerebral artery who presented with sudden onset severe headache. Intraprocedural digital subtraction angiography during right internal carotid artery injection after deploying the pCONus device showing one petal of the device (arrow) prolapsed out of the aneurysm neck into the inferior division of middle cerebral artery. Inset (a) Magnified image of the aneurysm with the deployed pCONus device, the prolapsed petal shown by arrowhead. Inset (b) Schematic diagram showing approximate outlines of the aneurysm and parent vessels (black outline), the four radiopaque markers of the petals (red lines) and the position of the distal petals and stent which are not radiopaque, hence not visible in angiogram. (blue-colored dashed shapes); the prolapsed petal is indicated by a black arrowhead.

Figure 3

A 64-year-old woman with a ruptured wide-necked bifurcation aneurysm of the right middle cerebral artery who presented with sudden onset severe headache. Postprocedural digital subtraction angiography during right carotid artery injection showing minimal filling of the aneurysm, patent middle cerebral artery branches, and prolapse of a coil loop into the middle cerebral artery (arrow).

A 64-year-old woman with a ruptured wide-necked bifurcation aneurysm of the right middle cerebral artery who presented with sudden onset severe headache. Digital subtraction angiography image taken during right internal carotid artery contrast injection showing a wide-necked right middle cerebral artery bifurcation aneurysm (arrow). Note mild irregularity of the distal cervical segment of internal carotid artery due to vasospasm (arrowhead). A 64-year-old woman with a ruptured wide-necked bifurcation aneurysm of the right middle cerebral artery who presented with sudden onset severe headache. Intraprocedural digital subtraction angiography during right internal carotid artery injection after deploying the pCONus device showing one petal of the device (arrow) prolapsed out of the aneurysm neck into the inferior division of middle cerebral artery. Inset (a) Magnified image of the aneurysm with the deployed pCONus device, the prolapsed petal shown by arrowhead. Inset (b) Schematic diagram showing approximate outlines of the aneurysm and parent vessels (black outline), the four radiopaque markers of the petals (red lines) and the position of the distal petals and stent which are not radiopaque, hence not visible in angiogram. (blue-colored dashed shapes); the prolapsed petal is indicated by a black arrowhead. A 64-year-old woman with a ruptured wide-necked bifurcation aneurysm of the right middle cerebral artery who presented with sudden onset severe headache. Postprocedural digital subtraction angiography during right carotid artery injection showing minimal filling of the aneurysm, patent middle cerebral artery branches, and prolapse of a coil loop into the middle cerebral artery (arrow). Follow-up angiography done at 3 months showed minimal residual filling and stable aneurysm size. There was a complete occlusion on follow-up angiography performed 6 months later, and the dissection in the internal carotid artery had healed without luminal compromise [Figure 4].
Figure 4

A 64-year-old woman with a ruptured wide-necked bifurcation aneurysm of the right middle cerebral artery who presented with sudden onset severe headache. Digital subtraction angiography during right internal carotid artery contrast injection, performed during the 6th month follow-up visit showing occlusion of aneurysm, persistence of prolapsed coil loop (arrow), and normal flow in the middle cerebral artery and its branches.

A 64-year-old woman with a ruptured wide-necked bifurcation aneurysm of the right middle cerebral artery who presented with sudden onset severe headache. Digital subtraction angiography during right internal carotid artery contrast injection, performed during the 6th month follow-up visit showing occlusion of aneurysm, persistence of prolapsed coil loop (arrow), and normal flow in the middle cerebral artery and its branches.

DISCUSSION

Wide-necked bifurcation aneurysms have always posed a challenge for endovascular treatment. Various strategies available for endovascular treatment of bifurcation aneurysms include coiling using 3D coils, balloon-assisted coiling using a single complaint or two balloons, stent-assisted coiling with two stents in Y or T configuration or single stent using “shelf” technique or waffle-cone technique, barrel stent-assisted coiling, intraaneurysmal flow diversion using WEB device, and pCONus-assisted coiling.[13567] TriSpan and PulseRider devices which are designed specifically for wide-necked bifurcation aneurysms are not widely used.[4] 3D coils have their limitations if the aneurysm is shallow, where even their shape alone is not enough to prevent prolapse into parent vessel.[5] Balloon or stent-assisted techniques require cannulation of the efferent vessels from the neck region of the aneurysm which is many times technically very challenging.[15] The technique of cannulating an efferent vessel by creating of loop in the aneurysm and straightening the loop either with the “wire anchor loop traction” method or using a balloon, stent, or a coil for anchoring is hazardous due to traction forces exerted on the aneurysm wall, especially if ruptured.[18910] Simultaneous placement of multiple devices and catheters as required in balloon or stent-assisted coiling may cause hemodynamic compromise resulting in cerebrovascular events.[5] In addition, balloons have a propensity to cause dissection if overinflated and there is a possibility of coil prolapse on balloon deflation.[511] Placement of double balloon or double stents for coiling of wide-necked bifurcation aneurysms is laden with significant complication rates.[1112] Deployment of a stent with its distal end within an aneurysm instead of bridging the neck is called the “waffle-cone technique” and has been successfully applied using Neuroform, Enterprise, and Solitaire AB stents.[3131415] The pCONus device is a stent exclusively made for this purpose.[16] The pCONus device is a nitinol stent with distal radially flared four loops (petals) which are to be positioned at the base of the aneurysm and an additional six polyamide fibers at the distal diameter of the stent forming a barrier preventing coil prolapse into the parent vessels. The device is compatible with a 0.021 inch microcatheter, is retrievable even after complete deployment, and is detached electrolytically. pCONus-assisted coiling is technically easier and superior to double stent-assisted coiling since there is no necessity to cannulate the efferent vessels and there is less intravascular metal placement.[5] There is, however, an argument that there may be a detrimental effect on the hemodynamics due to flow diversion toward the aneurysm.[1718] A study on intraaneurysmal flow kinetics using DSA optical flow imaging after deployment of the pCONus device has shown no adverse increase in aneurysmal flow; on the contrary, it has shown a minimal reduction.[16] An added advantage of the device is that, in the case of aneurysm re-growth, the petals still remain at the neck of the aneurysm and facilitate coil retention within the aneurysm.[5] Usage of pCONus device for ruptured aneurysms requires dual antiplatelets which might confer additional risk to the procedure, however, has been used successfully in multiple studies.[145] Thromboembolic complications, both transient (5%–20%) and permanent (2.5%–10.5%), have been documented in various studies.[1451920] Failed detachment of a pCONus device has been reported, where the device was retrieved without re-sheathing into the microcatheter resulting in a nonflow-limiting dissection.[1] Prolapse of a petal of the pCONus device can occur when the device sizing is improper (choosing smaller crown size than appropriate), if the aneurysm is eccentric to the parent vessel, or if excessive traction is applied to the pCONus device after it is deployed at the aneurysm neck. The pCONus device is compatible and has been used with a 0.021” microcatheter. Un-re-sheath-ability of a deployed pCONus device is of serious concern in case the device is misaligned with one or more of its distal petals prolapsing into the parent vessels. In such cases, aneurysm can be coiled without an additional device if the first deployed 3D coil is supported within the aneurysm optimally by the device; if not, an additional balloon or a stent placed within the branch harboring the prolapsed petal can be used to support coil retention within the aneurysm. Metal coverage of pCONus of <5%[1] would not cause any difficulty in placement of an additional stent or a balloon, and there is a description of placement of two stents through a pCONus device for a successful treatment of a basilar artery aneurysm.[5] Removal of the misaligned pCONus device in its deployed state using gentle traction is another option, as has been described once in a case of failed detachment of the device by Fischer et al.,[1] but was not performed in this case fearing the risk of endothelial damage to the aneurysm neck and parent vessel which might result in a dissection, which in unfortunate circumstances could be flow limiting or occluding. Probably, usage of a 0.027” inner diameter microcatheter is likely to avoid this complication. To the best of our knowledge, un-re-sheath-ability of a misaligned pCONus device has not been described yet in literature; and such a complication is of serious concern if the misaligned device does not support coil retention within the aneurysm.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.
  19 in total

1.  Coil occlusion of wide-neck bifurcation aneurysms assisted by a novel intra- to extra-aneurysmatic neck-bridging device (pCONus): initial experience.

Authors:  M Aguilar-Pérez; W Kurre; S Fischer; H Bäzner; H Henkes
Journal:  AJNR Am J Neuroradiol       Date:  2013-12-19       Impact factor: 3.825

2.  The stent anchor technique for distal access through a large or giant aneurysm.

Authors:  Kyle M Fargen; Gregory J Velat; Matthew F Lawson; Brian L Hoh; J Mocco
Journal:  J Neurointerv Surg       Date:  2012-04-27       Impact factor: 5.836

3.  "Waffle cone"technique for intra/extra-aneurysmal stent placement for the treatment of complex and wide-necked bifurcation aneurysm.

Authors:  Tzu Hsien Yang; Ho Fai Wong; Ming Shiang Yang; Chang Hsien Ou; Tzu Lung Ho
Journal:  Interv Neuroradiol       Date:  2009-01-02       Impact factor: 1.610

4.  Waffle-cone technique with Solitaire™ AB remodeling device: endovascular treatment of highly selected complex cerebral aneurysms.

Authors:  Vojtech Sychra; Joachim Klisch; Maren Werner; Christian Dettenborn; Alexander Petrovitch; Christoph Strasilla; Rüdiger Gerlach; Steffen Rosahl; Markus Holtmannspötter
Journal:  Neuroradiology       Date:  2010-09-07       Impact factor: 2.804

5.  The balloon anchor technique: a novel technique for distal access through a giant aneurysm.

Authors:  Kenneth V Snyder; Sabareesh K Natarajan; Erik F Hauck; J Mocco; Adnan H Siddiqui; L Nelson Hopkins; Elad I Levy
Journal:  J Neurointerv Surg       Date:  2010-06-25       Impact factor: 5.836

6.  Waffle Y technique: pCONus for tandem bifurcation aneurysms of the middle cerebral artery.

Authors:  Anastasios Mpotsaris; Hans Henkes; Werner Weber
Journal:  J Neurointerv Surg       Date:  2013-12-20       Impact factor: 5.836

7.  LVIS Jr 'shelf' technique: an alternative to Y stent-assisted aneurysm coiling.

Authors:  Elizabeth Hai Yen Du; Jai Jai Shiva Shankar
Journal:  J Neurointerv Surg       Date:  2016-02-04       Impact factor: 5.836

8.  Waffle-Cone Technique Using Solitaire AB Stent.

Authors:  Hye-Ran Park; Seok-Mann Yoon; Jai-Joon Shim; Sung-Ho Kim
Journal:  J Korean Neurosurg Soc       Date:  2012-04-30

9.  Intra-aneurysmal hemodynamics: evaluation of pCONus and pCANvas bifurcation aneurysm devices using DSA optical flow imaging.

Authors:  Marta Aguilar Pérez; Hans Henkes; Pierre Bouillot; Olivier Brina; Lee-Anne Slater; Vitor Mendes Pereira
Journal:  J Neurointerv Surg       Date:  2015-12-23       Impact factor: 5.836

10.  Single-center experience in the endovascular treatment of wide-necked intracranial aneurysms with a bridging intra-/extra-aneurysm implant (pCONus).

Authors:  S Fischer; A Weber; A Titschert; C Brenke; A Kowoll; W Weber
Journal:  J Neurointerv Surg       Date:  2015-12-09       Impact factor: 5.836

View more
  1 in total

1.  Erratum: Un-re-sheath-able Misaligned pCONus Device: Case Report of a Unique Complication.

Authors: 
Journal:  J Clin Imaging Sci       Date:  2017-12-11
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.