Literature DB >> 28817961

Evaluating the Ergonomics of Flexible Ureteroscopy.

Wesley W Ludwig1, Gyusung Lee2, Justin B Ziemba1, Joan S Ko1, Brian R Matlaga1.   

Abstract

INTRODUCTION: To date, the ergonomics of flexible ureteroscopy (URS) have not been well described. We performed a study to assess the biomechanical stresses on urologists performing URS and to investigate the effect of ureteroscope type on these parameters.
METHODS: Electromyography (EMG) was used to quantify the activation level of muscle groups involved in URS. Surface EMG electrodes (Delsys, Boston, MA) were placed on the right and left thenar, flexor carpi ulnaris (FCU), extensor carpi ulnaris (ECU), biceps, triceps, and deltoid. Three endoscopes were studied: single-use digital (Boston Scientific LithoVue), reusable digital (Karl Storz Flex-Xc), and reusable fiber-optic (Karl Storz Flex-X2). Each ureteroscope was used to perform a set sequence of navigation and procedural tasks in a training model. EMG data were processed and normalized to compare the maximum voluntary contractions between muscle groups. Cumulative muscular workload (CMW) and average muscular work per second (AWS) were used for comparative analysis.
RESULTS: For navigational tasks, CMW and AWS were greatest for the ECU, followed in descending order by right and left thenar, FCU, biceps, deltoid, and triceps. For procedural tasks, CMW and AWS were greatest for the right thenar, followed in descending order by the left thenar, ECU, FCU, triceps, biceps, and deltoid. During navigational tasks, both LithoVue and Flex-Xc had lower CMWs for every muscle group than Flex-X2 (p < 0.05). LithoVue and Flex-Xc had similar AWS and both were lower than Flex-X2 for the right thenar, ECU, biceps, and deltoid activation (p < 0.05). During procedural tasks, both LithoVue and Flex-Xc had lower CMWs and AWS for right and left thenar, ECU, and biceps than Flex-X2 (p < 0.05).
CONCLUSIONS: This study provides the first description of EMG-measured ergonomics of URS. Both the single-use and reusable digital ureteroscopes have similar profiles, and both have significantly better ergonomic metrics than the reusable fiber-optic ureteroscope.

Entities:  

Keywords:  electromyography; ergonomics; ureteroscopy; urolithiasis

Mesh:

Year:  2017        PMID: 28817961     DOI: 10.1089/end.2017.0378

Source DB:  PubMed          Journal:  J Endourol        ISSN: 0892-7790            Impact factor:   2.942


  3 in total

Review 1.  Which flexible ureteroscope is the best for upper tract urothelial carcinoma treatment?

Authors:  Etienne Xavier Keller; Steeve Doizi; Luca Villa; Olivier Traxer
Journal:  World J Urol       Date:  2019-02-15       Impact factor: 4.226

Review 2.  Surgical ergonomics for urologists: a practical guide.

Authors:  Andrew T Gabrielson; Marisa M Clifton; Christian P Pavlovich; Michael J Biles; Mitchell Huang; Jacqueline Agnew; Phillip M Pierorazio; Brian R Matlaga; Petar Bajic; Zeyad R Schwen
Journal:  Nat Rev Urol       Date:  2021-01-11       Impact factor: 14.432

3.  Ergonomic advantage of pistol-grip endoscope in the ENT practice.

Authors:  Itaru Watanabe; Makoto Miyamoto; Hideki Nakagawa; Koichiro Saito
Journal:  Laryngoscope Investig Otolaryngol       Date:  2021-02-25
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.