Literature DB >> 28804597

Size and concentration analyses of extracellular vesicles by nanoparticle tracking analysis: a variation study.

Beate Vestad1,2, Alicia Llorente2,3,4, Axl Neurauter2,5, Santosh Phuyal2,3,4, Bente Kierulf2,5, Peter Kierulf1,2, Tore Skotland2,3,4, Kirsten Sandvig2,3,4, Kari Bente F Haug1,2, Reidun Øvstebø1,2.   

Abstract

Current methods for characterisation of extracellular vesicles (EVs) need further standardisation in order to obtain an acceptable level of data comparability. Size and concentration of EVs can be determined by nanoparticle tracking analysis (NTA). However, both the heterogeneity of EVs and the choice of instrument settings may cause an appreciable analytical variation. Intra-assay (within-day, n = 6) and inter-assay (day-to-day, n = 6) variations (coefficient of variation, % CV) of different preparations of EVs and artificial vesicles or beads were determined using two NanoSight NS500 instruments, located at different laboratories. All analyses were performed by the same operator. The effect of applying identical software settings or instrument-optimised settings for each sample type and instrument was also evaluated. Finally, the impact of different operators and the use of two different software versions were investigated. The intra-assay CVs were 1-12% for both EVs and artificial samples, measured on the same instrument. The overall day-to-day variation was similar for both instruments, ranging from 2% to 25%. However, significantly different results were observed between the two instruments using identical software settings. The effect of applying instrument-optimised settings reduced the mismatch between the instruments, resulting in little to no significant divergences. The impact of using different operators and software versions when analysing silica microspheres and microvesicles from monocytes using instrument-optimised settings on the same instrument did not contribute to significant variation compared to the overall day-to-day variation of one operator. Performance differences between two similar NTA instruments may display significant divergences in size and concentration measurements when analysing EVs, depending on applied instrument settings and technical conditions. The importance of developing a streamlined and standardised execution of analysis, as well as monitoring longitudinal variation parameters on both biological and synthetic samples, should be highlighted.

Entities:  

Keywords:  Exosomes; microvesicles; nanoparticle tracking analysis; repeatability; reproducibility

Year:  2017        PMID: 28804597      PMCID: PMC5533132          DOI: 10.1080/20013078.2017.1344087

Source DB:  PubMed          Journal:  J Extracell Vesicles        ISSN: 2001-3078


Introduction

Cells release different types of membrane-embedded extracellular vesicles (EVs) that, based on their mechanisms of release and size, are grouped into different classes. Exosomes are secreted after fusion of multivesicular bodies with the plasma membrane, whereas microvesicles are thought to bud directly from the plasma membrane. The majority of EVs occur within the sub-micron range (30–1000 nm) [1], where exosomes are believed to be most abundantly present in the lowest size range (30–150 nm). The potential of EVs for diagnosis and treatment of different diseases is being intensely investigated, as EV size, composition and concentration may give important clinical information. However, their small size and heterogeneity present challenges for their isolation, analysis and biomedical use. Methods for isolation, identification and quantification of EVs are far from standardised, and there is an on-going call for consensus in order to ensure an acceptable level of data comparability [2]. Nanoparticle tracking analysis (NTA) is a powerful characterisation technique that combines the properties of both laser light scattering microscopy and Brownian motion in order to obtain size distributions of particles in liquid suspension. NanoSight instruments (Malvern, UK), which currently are the most widely used instruments for NTA in the EV field, are equipped with one or more lasers and an optical microscope connected to a digital camera. According to the manufacturer, NanoSight enables characterisation of particles from 10–2000 nm in solution. Particles are visualised by the light they scatter upon laser illumination, and their Brownian motion is monitored. The NTA software enables sizing of single particles by tracking their mean squared displacement and thereby calculating their theoretical hydrodynamic diameter using the Stokes Einstein equation. On the basis of knowing the analysed sample volume, NTA also allows for an estimation of particle concentration [3]. More details on the principle of NTA have been described elsewhere [4,5]. Analysis of size distribution of homogeneous particle preparations of either same or mixed sizes by NanoSight has previously been shown to be accurate [6], whereas analysis of heterogeneous biological vesicles has proven to be more challenging. In 2013, a study showed that silica microspheres, with a refractive index close to EVs, could be used for calibration of NanoSight measurements [7]. Moreover, operator adjustments of capture and analysis settings, such as camera level and detection threshold, have been shown to strongly affect the NanoSight-based quantification of EVs [7,8]. Consequently, a standardised protocol for the best capture and analysis settings, both with respect to the type of sample being analysed and individual instrument could minimise the variability and imprecision associated with the technique. In addition, results from an inter-laboratory comparison of size measurements on synthetic, monodisperse nanoparticles [9], revealed the importance of a well-established procedure for highly reproducible and accurate NTA measurements. The study showed improved reproducibility of an average coefficient of variation (CV) from 38.5% down to 4.4% throughout four rounds of measurements, by introducing protocols with analysis guidelines to the laboratories involved. The importance of assessing measurement variation regarding size and concentration of EVs has been strongly emphasised in the literature, in order to increase sensitivity and reproducibility of particle measurement techniques [10-12]. To our knowledge, assessing measurement variation of EVs using NTA has otherwise been sparsely described in the literature, thus there still is a need to elucidate this issue. In this study, we aimed to evaluate intra- and inter-assay variation of NanoSight measurements of both size and concentration of different preparations of EVs including microvesicles from human monocytes, exosomes from cell lines and outer membrane vesicles (OMVs) from Neisseria meningitidis. In addition, we performed variation analyses of different commercially available artificial vesicles/beads. The effect of applying identical vs. instrument-optimised measurement and analysis settings on two identical NanoSight instruments located separately, was also evaluated.

Methods

Sample preparation and isolation of EVs

Exosomes from cell cultures

PC-3 cells (ATCC) were grown in Ham’s F-12/DMEM (1:1) medium supplemented with 7% w/v foetal bovine serum (FBS), 100 units/mL penicillin and 100 units/mL streptomycin, in a humidified 5% CO2 atmosphere at 37°C. When cells were approximately 70% confluent, they were washed twice with FBS-free media, and incubated in this medium for approximately 24 h. Exosomes were then isolated from the culture medium by sequential ultracentrifugation [13]. Briefly, the supernatant was centrifuged at different speeds, 300g for 10 min, 1000g for 10 min and 10,000g for 30 min to remove cells, cell debris and microvesicles. Finally, the supernatant was ultracentrifuged at 100,000g for 70 min to pellet exosomes. The exosome pellet was washed with PBS and ultracentrifuged again at 100,000g for 70 min. Exosome pellets were resuspended in PBS, aliquoted and stored at −80°C until further use. Jurkat cells (Clone E6-1, ATCC) were cultivated for 7 days in a CELLine bioreactor (CELLine CL 1000, Integra) with RPMI 1640 supplemented with 10% w/v FBS. The exosome-containing medium from the cell compartment was centrifuged stepwise: 200g for 10 min, 400g for 10 min, then 2000g for 30 min at room temperature. The exosomes were then enriched from the cell culture media using the Total Exosome Isolation Reagent (Thermo Fisher Scientific) according to the manufacturer’s instructions.

OMVs from N. meningitidis

The bacteria were grown on blood agar plates for 15 h at 37°C, harvested in STA buffer (0.2 M LiCl, 0.1 M NaAc, pH 5.8), and inactivated for 30 min at 60°C. Cells and debris were removed from the suspension by sequential centrifugation (4000g, 20 min and 18,000g, 15 min). The supernatant was further centrifuged at 140,000g for 90 min and the pellet resuspended in PBS, aliquoted and stored at −80°C.

Microvesicles from primary human monocytes

Microvesicles were harvested from supernatants (30 mL) of E. coli LPS (10 ng/mL)-stimulated primary human monocytes (1 × 108) [14] after incubation for 4 h in RPMI 1640 medium containing 5% v/v FBS depleted for EVs (ultracentrifuged at 113,000g for 16 h). Cells and debris were removed by centrifugation at 4500g for 5 min, and the microvesicles pelleted by centrifugation at 17,000g for 30 min. Pelleted microvesicles were dissolved in 200 µL medium (same as above), aliquoted and stored at −80°C.

Artificial vesicles/beads

Artificial vesicles (Invivofectamine® 2.0, Invitrogen), 100 nm polystyrene latex beads (calculated concentration 1.8 × 1013 particles/mL, given at a prediluted stock solution of 1:100, Duke Scientific Corp./NanoSight) and 150 nm silica microspheres (predicted concentration 3.0 × 1013 particles/mL, Polysciences, Inc.) were stored at +4°C. Predicted concentrations of 100 nm polystyrene beads were calculated as follows: where W equals % mass of polymer (% solid), D is diameter in microns of latex particles and ρ is density of polymer in grams per millilitre (1.05 for polystyrene).

Transmission electron microscopy of EVs

To verify the presence of intact EVs, the preparations were analysed using transmission electron microscopy (TEM). Briefly, fixed (4% formaldehyde, 0.2% glutaraldehyde) EV samples were allowed to attach to Formvar/carbon-coated grids for 15–20 min. For labelling, PC-3-derived exosomes were first blocked with 0.5% BSA and then successively incubated with mouse anti-CD63 (H5C6) (Developmental Studies Hybridoma Bank, Iowa City, IA) followed by rabbit anti-mouse (Dako, Glostrup, Denmark) and then by 10 nm protein A-gold conjugates (Cell Microscopy Centre, Utrecht, Netherlands). Samples were then contrasted and embedded in a mixture of 0.4% uranyl acetate and 1.8% methyl cellulose. Exosomes from Jurkat cells were blocked with 0.5% BSA for 10 min, followed by labelling with anti-human CD81 (M38) (Invitrogen, Thermo Fisher Scientific) for 30 min, subsequent washing with PBS and addition of rabbit anti-mouse for 30 min. The EVs were then washed with PBS, incubated with 10 nm Protein A-gold particles for 15 min, washed again in PBS followed by distilled water and finally embedded in 0.3% uranyl acetate. For the other preparations, the grids were washed with PBS followed by double distilled water and stained with 0.4% uranyl acetate/1.8% methyl cellulose, and then dried. Exosomes from PC-3 cells were observed using a JEOL-JEM 1230 (JEOL Ltd., Tokyo, Japan) at 80 kV and images were acquired using a Morada digital camera and iTEM software (Olympus, Münster, Germany). The other EV preparations were observed using a FEI CM200 transmission electron microscope at 120 kV, and the images were recorded with a Quemesa CCD digital camera (Olympus Soft Imaging Solutions). The TEM images suggested that the preparations contained vesicles of various sizes (Figure 1).
Figure 1.

Representative transmission electron microscopy (TEM) images of negatively stained EVs used for NTA measurement variation assessment. The images show (a) exosomes derived from PC-3 cells labelled for CD63, (b) exosomes from Jurkat cells labelled for CD81, (c) OMVs derived from Neisseria meningitidis bacteria and (d) microvesicles from human monocytes. Scale bar sizes are 200, 100, 500 and 500 nm for (a), (b), (c) and (d), respectively. All images indicate the presence of intact vesicles of various sizes.

Representative transmission electron microscopy (TEM) images of negatively stained EVs used for NTA measurement variation assessment. The images show (a) exosomes derived from PC-3 cells labelled for CD63, (b) exosomes from Jurkat cells labelled for CD81, (c) OMVs derived from Neisseria meningitidis bacteria and (d) microvesicles from human monocytes. Scale bar sizes are 200, 100, 500 and 500 nm for (a), (b), (c) and (d), respectively. All images indicate the presence of intact vesicles of various sizes.

Intra/inter-assay variation of NanoSight measurements

Analyses were performed by the same operator on two different NanoSight NS500 instruments (Malvern Instruments, Amesbury, UK) located at different laboratories. The instruments were equipped with a 488 nm laser, a high sensitivity sCMOS camera and a syringe pump. The EVs and artificial vesicles/beads were mixed by vortexing, and subsequently diluted in particle-free PBS (0.02 µm filtered) to obtain a concentration within the recommended measurement range (1–10 × 108 particles/mL), corresponding to dilutions from 1:100 to 1:100,000 depending on the initial sample concentration (Tables 1 and 2). Experiment videos were analysed using NTA 2.3 build 17 or NTA 3.1 build 54 software (Malvern) (specified in Tables 1 and 2) after capture in script control mode (3 videos of 60 s per measurement) using syringe pump speed 20. A total of 1500 frames were examined per sample. Samples were captured and analysed by applying either identical or instrument-optimised (see “Inter-assay variation, instrument-optimised settings”) settings (Tables 1 and 2, respectively). Further settings, such as blur, minimum track length and minimum expected size were set to “automatic” and viscosity to “water” (0.909–0.90 cP).
Table 1.

Instrument settings for “identical settings” setup.

 “Identical settings” setup
Instruments 1 and 2
 Sample/vesicle typeSample dilutionCamera levelDetection threshold
EVsExosomes from PC-3 cells200123
Exosomes from Jurkat cells1000134
OMV from Neisseria meningitidis10,000123–4
Microvesicles from monocytes400123–4
Artificial vesicles/beadsArtificial vesicles (Invivofectamine® 2.0)100,00013–143–4
Polystyrene latex beads 100 nm50084
Silica microspheres 150 nm100,000104

Samples were analysed using NTA 2.3 build 17 software (Malvern).

Table 2.

Instrument settings for “instrument-optimised settings” setup.

 “Optimised settings” setup Instrument 1
Instrument 2
 Sample/vesicle typeSample dilutionCamera levelDetection ThresholdCamera levelDetection Threshold
EVsExosomes from PC-3 cells10014–15314–163
Microvesicles from monocytes50013–14314–153
BeadsPolystyrene latex beads 100 nm1,00011–12312–133

Samples were analysed using NTA 3.1 build 54 software (Malvern).

Instrument settings for “identical settings” setup. Samples were analysed using NTA 2.3 build 17 software (Malvern). Instrument settings for “instrument-optimised settings” setup. Samples were analysed using NTA 3.1 build 54 software (Malvern).

Intra-assay variation, identical settings

Intra-assay variation of vesicle size and concentration measurements was determined by analysing the same samples sequentially within the same day (n = 6) on instrument 1. Consequently, % CVs of mean vesicle size (nm), mode size (nm) and concentration (particles/mL) were calculated (Table 3). Camera levels were determined for each sample type by scatter properties of the first preparation. All samples were measured at 23°C, as selected in the software script. Video analysis settings were determined by increasing the screen gain and adjusting the detection threshold for single particle tracking.
Table 3.

Intra-assay variation (within day, n = 6) of size and concentration measurements on instrument 1, using identical settings in NTA 2.3 build 17 software.

 Intra-assay (within day)Mean size
Mode size
Mean concentration
 Sample/vesicle type(nm)% CV(nm)% CVE8 particles/mL% CV
EVsExosomes from PC-3 cells12829455.85
OMV from Neisseria meningitidis138210446.45
Microvesicles from monocytes141110346.87
Artificial vesicles/beadsArtificial vesicles (Invivofectamine® 2.0)8728145.412
Polystyrene latex beads 100 nm10229434.98
Silica microspheres 150 nm140213227.88
 Overall mean % CV 2 3 8
 Range % CV 1–2 2–5 5–12
Intra-assay variation (within day, n = 6) of size and concentration measurements on instrument 1, using identical settings in NTA 2.3 build 17 software.

Inter-assay variation, identical settings

Inter-assay variation was assessed by analysing the same sample aliquots with the same settings on two identical instruments located in two different laboratories on different days (n = 6) (Table 4). See also “Intra-assay variation, identical settings” section. Measurement curves from instrument 1 were generated (Supplementary Figure 1).
Table 4.

Inter-assay variation (day-to-day, n = 6) of size and concentration measurements on two NS500 instruments, using identical analysis settings in NTA 2.3 build 17 software.

  Mean sizeMean size Mode sizeMode size Mean concentrationMean concentration 
  Instrument 1
Instrument 2
 Instrument 1
Instrument 2
 Instrument 1
Instrument 2
 
 Inter-assay (day-to-day)Sample/vesicle type(nm)% CV(nm)% CVPer cent mismatch instrument 2 from 1(nm)% CV(nm)% CVPer cent mismatch instrument 2 from 1E8 particles/mL% CVE8 particles/mL% CVPer cent mismatch instrument 2 from 1
EVsExosomes from PC-3 cells1283142211*942113320*6.31810.81072*
Exosomes from Jurkat cells1304144611*91699695.6125.178
OMV from Neisseria meningitidis1392156412*1025122620*5.8168.9854*
Microvesicles from monocytes1474166613*1044127522*7.01010.01244*
Artificial vesicles/beadsArtificial vesicles (Invivofectamine® 2.0)8628913*7668248*5.81511.717102*
Polystyrene latex beads 100 nm99110324*9429612*6.0137.6527*
Silica microspheres 150 nm139314645130213846*7.61110.31135*
 Overall mean % CV 3 4  4 4  14 10 
 Range % CV 1–4 1–6  2–6 1–6  10–18 5–17 

*Significant difference (paired t-test, p < 0.05).

Inter-assay variation (day-to-day, n = 6) of size and concentration measurements on two NS500 instruments, using identical analysis settings in NTA 2.3 build 17 software. *Significant difference (paired t-test, p < 0.05).

Inter-assay variation, instrument-optimised settings

Inter-assay variation was assessed by analysing the same sample aliquots with instrument-optimised settings on different days (n = 3) (Table 5). The term “instrument-optimised” settings were in these experiments referred to as best visualisation of particles by applying software adjustments (camera level, focus and detection threshold) in order to optimise analysis results with respect to both sample type and individual instrument, without changing the settings within comparable sample series (Appendix 1). Samples were measured at 24°C. Automated image setup (camera level and focus) was chosen whenever applicable. See also the “Intra-assay variation, identical settings” section.
Table 5.

Inter-assay variation (day-to-day, n = 3) of two NS500 instruments, using instrument-optimised analysis settings in NTA 3.1 build 54 software.

  Mean sizeMean size Mode sizeMode size Mean concentrationMean concentration 
 Inter-assay (day-to-day)
Instrument 1
Instrument 2
 Instrument 1
Instrument 2
 Instrument 1
Instrument 2
 
 Sample/vesicle type(nm)% CV(nm)% CVPer cent mismatch instrument 2 from 1(nm)% CV(nm)% CVPer cent mismatch instrument 2 from 1E8 particles/mL% CVE8 particles/mL% CVPer cent mismatch instrument 2 from 1
EVsExosomes from PC-3 cells1252135128898921633.1152.9258
Microvesicles from monocytes17141672212731191064.343.9138
BeadsPolystyrene latex beads 100 nm97210528*9619914*4.714.11313
 Overall mean % CV 3 5  4 9  7 17 
 Range % CV 2–4 2–12  1–8 1–16  1–15 13–25 

*Significant difference (paired t-test, p < 0.05).

Inter-assay variation (day-to-day, n = 3) of two NS500 instruments, using instrument-optimised analysis settings in NTA 3.1 build 54 software. *Significant difference (paired t-test, p < 0.05).

Inter-operator variation

Analyses were independently performed by four different operators on Instrument 1 by measurement of 150 nm silica microspheres and microvesicles from monocytes, using instrument-optimised settings. Additionally, all sample videos were acquired and analysed in software versions NTA 2.3 and NTA 3.1 (Table 6).
Table 6.

Inter-operator variation by measurement of 150 nm silica microspheres (1:100,000) and microvesicles from monocytes (1:400) in software versions NTA 2.3 and NTA 3.1, using instrument-optimised settings.

 Inter-operator
Operator 1
Operator 2
Operator 3
Operator 4
 Instrument-optimised settingsNTA v2.3NTA v3.1NTA v2.3NTA v3.1NTA v2.3NTA v3.1NTA v2.3NTA v3.1
Silica microspheres 150 nmMean size (nm)142137136135129136130136
Mode size (nm)135135131129123133121132
Concentration (E8 particles/mL)6.57.47.29.98.18.59.77.1
Camera level1012101312121211
Detection threshold53445553
Microvesicles from monocytesMean size (nm)182182182184179166168179
Mode size (nm)138148128134134123125138
Concentration (E8 particles/mL)4.63.95.15.54.92.93.93.1
Camera level1414141514151514
Detection threshold33444353
Inter-operator variation by measurement of 150 nm silica microspheres (1:100,000) and microvesicles from monocytes (1:400) in software versions NTA 2.3 and NTA 3.1, using instrument-optimised settings.

Statistical analysis

Calculation of % CVs of size and concentration measurements was performed using GraphPad PRISM® Version 6.03 (San Diego, CA, USA). Additionally, statistical significance in measurements between instruments (Tables 4 and 5) and between software versions (Supplementary Table 1) were determined using paired t-tests (p < 0.05). Observational errors for polystyrene beads and silica microspheres were calculated as % BIAS. Longitudinal measurements of 100 nm polystyrene beads were monitored in Levey Jennings diagrams for up to 2 SD of mean values.

Results

Intra-assay variation

In order to investigate the performance of one of the NanoSight instruments (Instrument 1), three different EV preparations (exosomes from PC-3 cells, microvesicles from monocytes and OMVs from N. meningitidis (Figure 1)) and three different preparations of artificial vesicles/beads (Invivofectamine® 2.0, 100 nm polystyrene latex beads, 150 nm silica microspheres) were measured. The intra-assay CVs were determined within the same day (n = 6) using identical settings within sample series, and ranged from 1–5% for size measurements and from 5–12% for concentration measurements (Table 3).

Inter-assay variation

Inter-assay CVs of four different preparations of EVs (same as in “Intra-assay variation” in addition to exosomes from Jurkat cells) and three different artificial vesicles or beads on two differently located NS500 instruments using identical settings were evaluated from day to day (n = 6). The inter-assay CVs ranged from 1–6% for size measurements, and from 5–18% for concentration measurements. Significantly different results on both size and concentrations were observed when comparing the two instruments (p < 0.05) (Table 4). Measurement curves illustrating inter-assay variation (day-to-day) from Instrument 1 are shown in Supplementary Figure 1. Longitudinal monitoring of 100 nm polystyrene latex beads measured on Instrument 1 over a period of 18 months is shown in Levey Jennings plots (Supplementary Figure 2), corresponding to inter-assay CVs of 3%, 4% and 9% on mean size, mode size and concentration, respectively. To investigate the effect of applying instrument-optimised measurement settings (best visualisation of particles), three different sample preparations were measured on three different days on both instruments. The resulting inter-assay CVs ranged from 1–16% for size measurements, and from 1–25% for concentration measurements. Use of instrument-optimised settings on the two instruments resulted in no significant divergences considering EVs (Table 5). For polystyrene beads, observational errors (% BIAS) were calculated to be 158% in Instrument 1 and 128% in Instrument 2 (data not shown), reflecting predicted versus measured concentrations of 1.8 vs 4.7 × 108 particles/mL and 1.8 vs 4.1 × 108 particles/mL, respectively (Table 5). For silica microspheres, % BIAS was calculated to be 168% in Instrument 1 (data not shown), reflecting predicted versus mean measured concentration of 3.0 vs 8.0 × 108 particles/mL (Table 6). In order to investigate the influence of the operator on instrument-optimised settings, associated operator-dependent variations were also determined (Table 6 and Supplementary Table 1). For 150 nm silica microspheres and microvesicles from monocytes, the choice of camera levels ranged from 10 to 13 and 14 to 15, respectively. Detection threshold levels ranged from 3 to 5 for both sample types. There were no significant differences between the % CVs for either size or concentration measurements when comparing inter-operator variations in the two software versions NTA 2.3 and 3.1 (Supplementary Table 1).

Discussion

Standardisation of methodological platforms used for characterisation of EVs, both within and between laboratories, in order to compare research practice in the EV field, is of utmost importance [2,10]. The current study has evaluated single particle sizing and enumeration using NTA, for the primary purpose of obtaining standardised measurement practices across two laboratories. This was performed by analysing intra- and inter-assay variations on different biological and artificial vesicles or beads on two NanoSight NS500 instruments at different locations. The analyses were carried out by the same operator and by applying identical or instrument-optimised measurement settings. To elucidate the influence of the operator using instrument-optimised settings, four different operators performed measurements of silica microspheres and microvesicles from monocytes on the same instrument independently. Moreover, the sample videos were acquired and analysed in software versions NTA 2.3 and 3.1 to investigate potential software differences.

Repeatability and reproducibility of EV sizing and enumeration by NTA

Repeatability in particle sizing techniques may reflect the stability of a system when assessing precision of repeated consecutive measurements on a group of particles under identical conditions. The term reproducibility may be used when evaluating results obtained from measuring multiple aliquots of a bulk sample, but may also reflect variation occurring from inter-laboratory testing, such as instrument-to-instrument or day-to-day [15]. Evaluating repeatability and reproducibility in NTA measurements may be to consider intra- (within-day repeatability) and inter-assay (day-to-day reproducibility) variation of the measurements performed, as well as instrument-to-instrument reproducibility. A large number of variables may influence this assessment, either dependent on the operator or on the technical conditions of the instrument. Adjusting software settings, such as camera level (shutter and gain) and detection threshold may alone entail a noticeable variation when analysing samples. Various instrumental challenges such as air bubbles, temperature changes or laser misalignment may as well arise during use [7]. In addition, NTA detection limits may differ between various instrument versions depending on the hardware, such as illumination intensity, efficiency of the collection optics and quantum efficiency of the camera. Refractive index of the particles and the medium also influences NTA detection limits [16,17]. In our study, the two NTA instruments possessed identical hardware, and the sample aliquots were prepared uniformly in order to reduce pre-analytical variability. Variation originating from pre-analytical conditions, such as sample preparation, storage, freezing, thawing or precision of pipetting, may further contribute to the total variation of the samples to be compared. Pre-analytical issues are, however, beyond the scope of this paper and therefore have not been elucidated. Different methodological platforms can be used for single particle sizing and enumeration of EVs. Recent studies using tuneable resistive pulse sensing (TRPS) have evaluated the reproducibility of size and concentration measurements on EVs using optimised operating procedures. These studies showed inter-assay CVs ranging from 5–15% on size measurement of urine vesicles [12], while CVs on the concentration measurements of vesicles isolated from citrated plasma and analysed in six different laboratories was up to 52.5% [18]. Most studies that have evaluated NTA variation measurements to date have utilised synthetic beads or artificial vesicles, such as liposomes [7,9]. Only a few studies have reported measurement variation in biological samples analysed by NTA, however, verification on the presence of EVs has largely lacked. Analyses of particles with a diameter less than 300 nm in human unprocessed urine showed a within-sample variation of 47% when measuring the area under the curve of particle size vs. concentration [19]. Recently, a study assessing analytical, pre-analytical and biological variation on particle levels by NTA and TRPS in blood plasma revealed a generally low (below 10%) analytical variation on samples [20]. However, a considerable intra- and inter-individual variation was demonstrated. In our investigation, the focus has been to include biological vesicles isolated from different sources, such as cell lines (PC-3 and Jurkat), primary human monocytes and N. meningitidis colonies. As suggested by TEM, these samples are all heterogeneous in terms of size, concentration and morphology of vesicles. Moreover, we included a selection of artificial vesicles or beads, representing homogenous particle sources of known sizes and predicted concentrations.

Reduced variation between NanoSight instruments using instrument-optimised software settings

In general, the obtained % CVs observed in our study were reasonable, ranging from 3% to 14% when measuring with identical settings (Table 4) and 3–17% when measuring with instrument-optimised settings (Table 5) in both biological and artificial vesicles or beads on both instruments. Compared to inter-assay, the intra-assay variation on one of the instruments (Table 3) resulted in overall lower mean % CVs (2–8%), as could be expected when making repeated measurements on the same aliquot rather than from different aliquots on different days. When comparing results from the preparations analysed on two NanoSight instruments for size and concentration measurements using identical measurement settings (Table 4), a larger heterogeneous size distribution was observed for biological vesicles compared to artificial for both instruments. Furthermore, the measured mean and mode sizes showed significant divergences between the instruments. Also for the concentration measurements, the two instruments gave significantly different results. By applying instrument-optimised settings (Table 5), comparable inter-assay CVs were observed for both size and concentration measurements as for identical settings (Table 4). However, the divergences between the two instruments were greatly reduced for the observed sizes and concentrations of EVs to below the level of significance (Table 5). Overall, higher camera levels were applied using instrument-optimised settings (Table 2) than for identical settings (Table 1). This may partly be explained by new aliquot series for EVs, explaining the changed sample dilution. Additionally, both instruments had undergone several service operations (laser realignments, optical configurations) between the measurement periods, which might explain the need for changed camera levels. The overall mean % CVs were higher for concentration measurements than for size measurements, independently of whether identical or instrument-optimised settings were applied. Thus, we observed CVs of 10–14% compared to 3–4% for identical settings, and 7–17% compared to 3–9% for instrument-optimised settings for concentration and size measurements, respectively. This is likely due to the fact that concentration determination is more affected by camera level and detection threshold than size calculations, as explained by the Brownian motion and scatter properties of visible particles. An updated feature in newer software versions enables automated image setup (camera level and focus). However, we have found this to be most applicable for homogenous sample preparations, such as uniformly sized beads and less for heterogeneous samples, such as EVs. The influence of different operators could lead to additional variation beyond the existing inter-assay and inter-instrument variations. In our study, the inter-operator CVs for both size and concentration measurements were comparable to the inter-assay (day-to-day) CVs for one operator (Tables 4 and 5 and Supplementary Table 1). These observations suggest that the influence of the operator when applying instrument-optimised settings is low. As a bullet point in Appendix 2, we recommend to keep the software updated. However, one should take caution when comparing results measured by different software versions. Improved sizing algorithms and vibration correction in versions 3.X may result in improved measurement variation. Hence, this could have explained that our findings for instrument-optimised settings showed a lower per cent mismatch between the instruments (Tables 4 and 5). However, the overall % CVs per instrument were not markedly improved from software version 2.3 to version 3.1, also supported by the inter-operator variation results (Supplementary Table 1). It is therefore reasonable to infer that the instrument-to-instrument variation in our case was mainly reduced by applying instrument-optimised settings. In Table 4, the mean and mode sizes measured by instrument 1 are generally slightly lower than for instrument 2 using identical settings. This could suggest that instrument 1 is more sensitive for the settings used, expecting to detect higher particle concentrations than instrument 2; however, this is not the case. We may speculate that, throughout the period of the measurements, there was a slight difference in microscope magnification between the instruments, possibly due to inconsistent microscope calibration. Microscope objectives may deviate a few per cent from their specified magnifications. If the magnification of instrument 1 is underestimated, the distances of particle tracks could be overestimated, leading to overestimation of diffusion coefficients and underestimation of particle sizes. Likewise, an underestimation of the magnification of instrument 1 could result in a smaller field-of-view than assumed, and thus an underestimation of the particle concentration.

Improved reproducibility of NTA measurements by combining the use of traceable standards with close monitoring of instrument plasticity

Applying traceable standards in the field of sizing and enumeration of EVs is of utmost importance. However, challenges may arise when transferring the calibration properties of homogeneous, synthetic standards onto heterogeneous, biological samples, such as EVs. Traceable standards, such as silica beads, obviously give substantial information of instrument performance. The use of a correction factor from measuring silica microspheres in order to calibrate results obtained from measuring EVs has been suggested [7]. Additionally, the settings used for measuring silica particle size could be more transmissible than for polystyrene beads, reasoned that they have a refractive index closer to EVs. An important challenge for a calibrator is to mimic the analyte as much as possible; however, there is still a gap in order to reach calibration properties that can be directly transferable to EVs. The use of nanoerythrosomes, possessing suitable EV properties, that are stable, safe and can be manufactured relatively cheaply, is currently being investigated [21]. However, the use of such reference materials still needs to be further validated and implemented. In line with others [8,20], we yet believe that the importance of relative comparisons of particle concentrations within experiments should therefore be underlined, rather than using mathematical corrections. The fact that we, in our experiments, observed little differences in variation when comparing heterogeneous biological EVs to homogeneous synthetic samples for both size and concentration measurements, could support the future applicability of biological reference materials for EV measurements by NTA. When comparing our obtained numbers on synthetic beads from inter-assay (day-to-day) instrument-optimised settings, with the predicted concentrations (108 particles/mL), we found large observational errors (see Results section), suggesting quite unsatisfactory trueness of measurements. However, the predicted concentrations of these standard solutions are mathematically calculated from density, volume (% solids) and diameter, and could be largely inaccurate. We believe that, rather than reporting “true” concentrations of EV samples, the focus should be on relative comparisons of samples as well as monitoring measurement variability over time. Taken together, our findings suggest that instrument-optimised settings provide improved reproducibility on both size and concentrations on biological vesicles when comparing NTA instruments. Consequently, the importance of monitoring “instrument plasticity” during measurement should be highlighted. Up until present, there has been much focus on the importance of reporting NTA settings in EV publications. This could “mislead” new operators to directly copy software settings from other groups, given that their sample type and instrument hardware are similar. We believe that reproducible NTA measurements can be obtained by both the use of traceable standards and the close monitoring of instrument plasticity. This might result in different software settings on two identical instruments when measuring the same sample. Capture optimisation might be done by providing the best visualisation of particles, meaning that they are adequately exposed and focused. Furthermore, keeping the balance between generating background exposure and reduced particle visibility, potentially leading to falsely elevated or low counts, should be emphasised. Nevertheless, comparable settings when handling samples of similar refractive index [16,17] and within measurement series on one instrument, are recommended. Technical conditions should, however, remain consistent during the measurement period. Implementation of the NTA analysis in our lab has been time consuming. We have found it extremely important to have a dedicated person in the lab to develop and maintain a streamlined and standardised execution of analysis, thereby obtaining acquaintance with appropriate monitoring of the instrument and to assess any technical challenges related to the application used prior to evaluating results. Moreover, laboratories using NTA for EV research should be encouraged to monitor longitudinal variation parameters using both synthetic beads and biological samples, in order to validate the accuracy of their NTA measurements. Based on our own experience, we have listed several technical suggestions on how to reduce analytical variation when analysing EVs using NTA in Appendix 2. In summary, the NTA measurements in the present study showed good repeatability, as evaluated by the good precision of repeated consecutive measurements on a NanoSight NS500 instrument. The results were further verified by the use of two NS500 instruments at different locations (Table 4), implying fair reproducibility from day to day and between the instruments. However, the overall variation assessment showed significant divergences on size and concentrations between the two instruments compared, suggesting there could be performance differences between same type of NTA instruments depending on applied instrument settings and technical conditions. Whereas the influence of the operator and software version may become a contributing factor of variation when measuring EVs by NTA, the close monitoring of instrument plasticity and longitudinal measurements could greatly improve the variation between instruments. Click here for additional data file.
  19 in total

1.  Isolation of monocytes from whole blood by density gradient centrifugation and counter-current elutriation followed by cryopreservation: six years' experience.

Authors:  P K Lund; G B Joø; A B Westvik; R Ovstebø; P Kierulf
Journal:  Scand J Clin Lab Invest       Date:  2000-08       Impact factor: 1.713

2.  Refractive index determination of nanoparticles in suspension using nanoparticle tracking analysis.

Authors:  Edwin van der Pol; Frank A W Coumans; Auguste Sturk; Rienk Nieuwland; Ton G van Leeuwen
Journal:  Nano Lett       Date:  2014-10-02       Impact factor: 11.189

3.  Biological reference materials for extracellular vesicle studies.

Authors:  S Valkonen; E van der Pol; A Böing; Y Yuana; M Yliperttula; R Nieuwland; S Laitinen; P R M Siljander
Journal:  Eur J Pharm Sci       Date:  2016-09-10       Impact factor: 4.384

4.  Sizing and phenotyping of cellular vesicles using Nanoparticle Tracking Analysis.

Authors:  Rebecca A Dragovic; Christopher Gardiner; Alexandra S Brooks; Dionne S Tannetta; David J P Ferguson; Patrick Hole; Bob Carr; Christopher W G Redman; Adrian L Harris; Peter J Dobson; Paul Harrison; Ian L Sargent
Journal:  Nanomedicine       Date:  2011-05-04       Impact factor: 5.307

5.  Particle size distribution of exosomes and microvesicles determined by transmission electron microscopy, flow cytometry, nanoparticle tracking analysis, and resistive pulse sensing.

Authors:  E van der Pol; F A W Coumans; A E Grootemaat; C Gardiner; I L Sargent; P Harrison; A Sturk; T G van Leeuwen; R Nieuwland
Journal:  J Thromb Haemost       Date:  2014-06-19       Impact factor: 5.824

6.  Extracellular vesicle sizing and enumeration by nanoparticle tracking analysis.

Authors:  Chris Gardiner; Yannick J Ferreira; Rebecca A Dragovic; Christopher W G Redman; Ian L Sargent
Journal:  J Extracell Vesicles       Date:  2013-02-15

7.  Possibilities and limitations of current technologies for quantification of biological extracellular vesicles and synthetic mimics.

Authors:  Sybren L N Maas; Jeroen de Vrij; Els J van der Vlist; Biaina Geragousian; Louis van Bloois; Enrico Mastrobattista; Raymond M Schiffelers; Marca H M Wauben; Marike L D Broekman; Esther N M Nolte-'t Hoen
Journal:  J Control Release       Date:  2014-12-30       Impact factor: 9.776

8.  Standardization of sample collection, isolation and analysis methods in extracellular vesicle research.

Authors:  Kenneth W Witwer; Edit I Buzás; Lynne T Bemis; Adriana Bora; Cecilia Lässer; Jan Lötvall; Esther N Nolte-'t Hoen; Melissa G Piper; Sarada Sivaraman; Johan Skog; Clotilde Théry; Marca H Wauben; Fred Hochberg
Journal:  J Extracell Vesicles       Date:  2013-05-27

9.  Measurement of refractive index by nanoparticle tracking analysis reveals heterogeneity in extracellular vesicles.

Authors:  Chris Gardiner; Michael Shaw; Patrick Hole; Jonathan Smith; Dionne Tannetta; Christopher W Redman; Ian L Sargent
Journal:  J Extracell Vesicles       Date:  2014-11-24

10.  A standardized method to determine the concentration of extracellular vesicles using tunable resistive pulse sensing.

Authors:  Robert Vogel; Frank A W Coumans; Raluca G Maltesen; Anita N Böing; Katherine E Bonnington; Marike L Broekman; Murray F Broom; Edit I Buzás; Gunna Christiansen; Najat Hajji; Søren R Kristensen; Meta J Kuehn; Sigrid M Lund; Sybren L N Maas; Rienk Nieuwland; Xabier Osteikoetxea; Rosalie Schnoor; Benjamin J Scicluna; Mitch Shambrook; Jeroen de Vrij; Stephen I Mann; Andrew F Hill; Shona Pedersen
Journal:  J Extracell Vesicles       Date:  2016-09-27
View more
  71 in total

1.  Characterization of Extracellular Vesicles from Entamoeba histolytica Identifies Roles in Intercellular Communication That Regulates Parasite Growth and Development.

Authors:  Manu Sharma; Pedro Morgado; Hanbang Zhang; Gretchen Ehrenkaufer; Dipak Manna; Upinder Singh
Journal:  Infect Immun       Date:  2020-09-18       Impact factor: 3.441

Review 2.  NeuroEVs: Characterizing Extracellular Vesicles Generated in the Neural Domain.

Authors:  Christie D Fowler
Journal:  J Neurosci       Date:  2019-11-20       Impact factor: 6.167

3.  Exosomes in disease and regeneration: biological functions, diagnostics, and beneficial effects.

Authors:  Yun Lin; Johnathon D Anderson; Lily M A Rahnama; Shenwen V Gu; Anne A Knowlton
Journal:  Am J Physiol Heart Circ Physiol       Date:  2020-09-28       Impact factor: 4.733

4.  CNS-derived extracellular vesicles from superoxide dismutase 1 (SOD1)G93A ALS mice originate from astrocytes and neurons and carry misfolded SOD1.

Authors:  Judith M Silverman; Darren Christy; Chih Cheih Shyu; Kyung-Mee Moon; Sarah Fernando; Zoe Gidden; Catherine M Cowan; Yuxin Ban; R Greg Stacey; Leslie I Grad; Luke McAlary; Ian R Mackenzie; Leonard J Foster; Neil R Cashman
Journal:  J Biol Chem       Date:  2019-01-11       Impact factor: 5.157

5.  High Glucose Induces Mesangial Cell Apoptosis through miR-15b-5p and Promotes Diabetic Nephropathy by Extracellular Vesicle Delivery.

Authors:  Yi-Chun Tsai; Mei-Chuan Kuo; Wei-Wen Hung; Ling-Yu Wu; Ping-Hsun Wu; Wei-An Chang; Po-Lin Kuo; Ya-Ling Hsu
Journal:  Mol Ther       Date:  2020-01-15       Impact factor: 11.454

6.  Comparison of the capillary-channeled polymer (C-CP) fiber spin-down tip approach to traditional methods for the isolation of extracellular vesicles from human urine.

Authors:  Kaylan K Jackson; Rhonda R Powell; R Kenneth Marcus; Terri F Bruce
Journal:  Anal Bioanal Chem       Date:  2022-04-12       Impact factor: 4.142

7.  Novosphingobium sp. PP1Y as a novel source of outer membrane vesicles.

Authors:  Federica De Lise; Francesca Mensitieri; Giulia Rusciano; Fabrizio Dal Piaz; Giovanni Forte; Flaviana Di Lorenzo; Antonio Molinaro; Armando Zarrelli; Valeria Romanucci; Valeria Cafaro; Antonio Sasso; Amelia Filippelli; Alberto Di Donato; Viviana Izzo
Journal:  J Microbiol       Date:  2019-05-27       Impact factor: 3.422

8.  Exosomal MicroRNAs Associate With Neuropsychological Performance in Individuals With HIV Infection on Antiretroviral Therapy.

Authors:  Tess OʼMeara; Yong Kong; Jennifer Chiarella; Richard W Price; Rabib Chaudhury; Xinran Liu; Serena Spudich; Kevin Robertson; Brinda Emu; Lingeng Lu
Journal:  J Acquir Immune Defic Syndr       Date:  2019-12-15       Impact factor: 3.731

9.  Identification of extracellular nanoparticle subsets by nuclear magnetic resonance.

Authors:  Md Sharif Ullah; Vladimir V Zhivonitko; Anatoliy Samoylenko; Artem Zhyvolozhnyi; Sirja Viitala; Santeri Kankaanpää; Sanna Komulainen; Leif Schröder; Seppo J Vainio; Ville-Veikko Telkki
Journal:  Chem Sci       Date:  2021-04-29       Impact factor: 9.825

Review 10.  Exosomes as Naturally Occurring Vehicles for Delivery of Biopharmaceuticals: Insights from Drug Delivery to Clinical Perspectives.

Authors:  Arun Butreddy; Nagavendra Kommineni; Narendar Dudhipala
Journal:  Nanomaterials (Basel)       Date:  2021-06-03       Impact factor: 5.076

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.