Jae Min Shin1, Tae Hoon Kim1, Seokjin Haam2, Kyunghwa Han3, Min Kwang Byun4, Yoon Soo Chang4, Hyung Jung Kim4, Chul Hwan Park1.
Abstract
In this study, we examined the repeatability of computed tomography (CT) lung volume measurements in healthy individuals and patients with obstructive and restrictive lung diseases. To do this, we retrospectively enrolled 200 healthy individuals (group 1), 100 patients with obstructive lung disease (group 2), and 100 patients with restrictive lung disease (group 3) who underwent two consecutive chest CT scans within a 1-year period. The CT lung volume was measured using a threshold-based, three-dimensional auto-segmentation technique at a default range from -200 to -1024 HU. The within-subject standard deviation, repeatability coefficient, within-subject coefficient variability, and intraclass correlation coefficient were evaluated. No significant differences were identified between the two consecutive CT lung volume measurements in any of the groups (p> 0.05). The within-subject standard deviations for groups 1, 2, and 3 were 441.1, 387.0, and 288.6, respectively, while the repeatability coefficients were 1222.6, 1072.6, and 800.1, respectively. The within-subject coefficient variabilities for groups 1, 2, and 3 were 0.097, 0.083, and 0.090, respectively, while the intraclass correlation coefficients were 0.818, 0.881, and 0.910, respectively. The two CT lung volume measurements showed excellent agreement in healthy individuals and patients with obstructive or restrictive lung disease. However, the repeatability was lower in healthy individuals than it was in patients with lung diseases.Entities:
Mesh:
Year: 2017 PMID: 28796825 PMCID: PMC5552110 DOI: 10.1371/journal.pone.0182849
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Demographic data from the healthy participants (group1), patients with obstructive lung disease (group 2), and patients with restrictive lung disease (group 3).
| Group 1 | Group 2 | Group 3 | ||
|---|---|---|---|---|
| 129:71 | 79:21 | 52:48 | <0.001 | |
| 53.6 ± 9.1 | 68.0 ± 10.7 | 64.2 ± 12.8 | <0.001 | |
| 166.7 ± 8.2 | 165.5 ± 7.8 | 161.5 ± 8.7 | <0.001 | |
| 67.1 ± 12.5 | 62.9 ± 12.5 | 61.9 ± 11.8 | 0.001 | |
| 97.2 ± 9.9 | 85.3 ± 17.9 | 67.9 ± 14.5 | <0.001 | |
| 104.6 ± 11.8 | 67.1 ± 21.2 | 76.3 ± 16.3 | <0.001 | |
| 79.4 ± 5.7 | 53.3 ± 13.5 | 80.1 ± 9.2 | <0.001 |
Data are presented as the mean±the standard deviation, unless otherwise stated
FVC forced vital capacity, FEV forced expiratory volume in 1 s
*Group 1 vs. Group 2 <0.05, Group 2 vs. Group 3 < 0.05, and Group 3 vs Group 1 < 0.05
† Group 2 vs. Group 3<0.05
‡ Group 1 vs. Group 2<0.05
Two consecutive computed tomography lung volume measurements in healthy participants (group1), patients with obstructive lung disease (group 2), and patients with restrictive lung disease (group 3).
| Volume 1 (mL) | Volume 2 (mL) | Scan interval (days) | ||
|---|---|---|---|---|
| 4525.8 ± 1056.4 | 4539.9 ± 1009.6 | 0.751 | 361 (293, 365) | |
| 4657.6 ± 1138.4 | 4639.6 ± 1102.8 | 0.744 | 279 (30, 365) | |
| 3234.7 ± 947.1 | 3198.0 ± 978.6 | 0.371 | 182 (24, 365) |
Data are presented as the mean±the standard deviation, unless otherwise stated.
The median interval between the two CT scans is presented with the minimum and maximum values.
Fig 1Comparisons between the two consecutive computed tomography (CT) lung volumes in healthy participants (group 1), patients with obstructive lung disease (group 2), and patients with restrictive lung disease (group 3).
Plots show the lung volumes from the first and follow-up CT scans. From left to right are the correlations from the two CT scans in (a) group 1 (n = 200, correlation coefficient (r) = 0.818, p<0.01), (b) group 2 (n = 100, r = 0.880, p<0.01), and (c) group 3 (n = 100, r = 0.910, p<0.01).
Fig 2Bland-Altman plots of the correlations between the two consecutive computed tomography (CT) lung volumes in healthy participants (group1), patients with obstructive lung disease (group 2), and patients with restrictive lung disease (group 3).
(a) in group 1, (b) group 2,(c) group 3. The dashed line indicates the mean bias, while the solid line indicates ±1.96 standard deviation.
Repeatability of the CTvol measurements in healthy participants (group1), patients with obstructive lung disease (group 2), and patients with restrictive lung disease (group 3).
| Group | Mean | wSD | RC | CV | ICC |
|---|---|---|---|---|---|
| 4532.9 | 441.1 | 1222.6 | 0.097 | 0.818 | |
| 4648.6 | 387.0 | 1072.6 | 0.083 | 0.881 | |
| 3216.4 | 288.6 | 800.1 | 0.090 | 0.910 |
wSD within-subject standard deviation, CI confidence interval, RC repeatability coefficient, CV within-subject coefficient variability, ICC intraclass correlation coefficient