L M de Vries1, M J G Leening1,2,3, W A Dijk4, C A M Hooijschuur4, B H C Stricker5, N M van Hemel6.
Abstract
AIMS: After decades of experience and strongly improved technology, service time of pacemaker generators is expected to increase. To test this hypothesis, we conducted a retrospective review of a large cohort of patients with a pacemaker.Entities:
Keywords: Generator replacement; Pacemaker follow-up; Pacemaker infection; Pacemaker longevity; Pacemaker recall
Year: 2017 PMID: 28770398 PMCID: PMC5612868 DOI: 10.1007/s12471-017-1024-x
Source DB: PubMed Journal: Neth Heart J ISSN: 1568-5888 Impact factor: 2.380
Baseline characteristics of patients in the nationwide CPPR-SPRN database having a first pacemaker implanted (n = 96,900), the Netherlands 1984–2007
| 1984–1990 | 1991–1995 | 1996–2000 | 2001–2005 | 2006–2007 | Total study period | |
|---|---|---|---|---|---|---|
|
| ||||||
| Mean age at first implantation, years (SD) | 72.7 (13.0) | 73.0 (13.0) | 72.9 (13.2) | 73.1 (12.9) | NA | 72.9 (13.0) |
| Female gender, | 11,628 (48.7) | 9419 (47.9) | 10,569 (46.6) | 14,045 (45.8) | NA | 45,661 (47.1) |
| Deaths, | 7245 (30.4) | 6119 (31.1) | 5849 (25.8)c | 3721 (12.1)c | NA | 22,934 (23.7) |
| Follow-up duration (range), years | 17–23 | 12–16 | 7–11 | 2–6 | NA | 2–23 |
|
| ||||||
|
| 24,952 (20.0) | 23,451 (18.8) | 29,789 (23.9) | 41,569 (33.4) | 4798 (3.9) | 124,559 (100.0) |
| – First pacemaker | 23,870 (95.7) | 19,659 (83.8)c | 22,694 (76.2)c | 30,677 (73.8)c | NA | 96,900 (77.8) |
| – First replacement | 1004 (4.0) | 3352 (14.3)c | 5821 (19.5)c | 8386 (20.2) | 3571 (74.4) | 22,134 (17.8) |
| – Second replacement | 73 (0.3) | 389 (1.7)c | 1050 (3.5)c | 1930 (4.6)c | 908 (18.9) | 4350 (3.5) |
| – Third or more replacement | 5 (<0.1) | 51 (0.2)c | 224 (0.8)c | 576 (1.4)c | 319 (6.6) | 1175 (0.9) |
|
| 7238 (29.0) | 8402 (35.8)c | 9250 (31.1)c | 3000 (7.2)c | 47 (1.0) | 27,937 (22.4) |
| – Explantation <7 years, | 3233 (44.7) | 4489 (53.4)c | 5745 (62.1)c | NA | NA | 16,514 (59.1) |
| – Explantation <5 years, | 1692 (23.4) | 1972 (23.5) | 2455 (26.5)c | NA | NA | 8586 (30.6) |
| – Explantation <3 years, | 950 (13.1) | 943 (11.2)c | 1196 (12.9)c | NA | NA | 4683 (16.8) |
| Without replacement/without immediate replacement, | 41 (0.6) | 33 (0.4) | 66 (0.7) | 181 (6.0)c | 11 (23.4) | 332 (1.2) |
|
| 7.7 (4.0) | 6.8 (3.0)c | 6.0 (2.4)c | 2.9 (2.0)c | 0.5 (0.5) | 6.3 (3.3) |
|
| 7.5 | 6.8 | 6.4 | 2.8 | 0.3 | 6.4 |
For 2006–2007 we used data on replacements only
SD standard deviation, NA not available
aFor 75 patients, data on gender was missing. For 38 males and 22 females, data on age was missing
bData on number of deaths is incomplete
cSignificantly different compared to previous time interval (p ≤ 0.001), 2006–2007 not tested
Reasons for pacemaker generator replacements within 7 years of follow-up after implantation, stratified for implantation period, and for the entire cohort irrespective of duration of follow-up in the nationwide CPPR-SPRN database, the Netherlands 1984–2007
| Reason for explantation of pacemaker generators within 7 years of follow-up, stratified for implantation perioda
| All pacemakers explanted during study periodb | |||
|---|---|---|---|---|
| 1984–1990 | 1991–1995 | 1996–2000 | 1984–2007 | |
| Device failure | 126 (3.9) | 151 (3.4) | 105 (1.8)c | 509 (1.8) |
| Recall | 32 (1.0) | 174 (3.9)c | 270 (4.7) | 583 (2.1) |
| Infection | 149 (4.6) | 176 (3.9) | 248 (4.3) | 884 (3.2) |
| Other complication | 31 (1.0) | 31 (0.7) | 39 (0.7) | 153 (0.5) |
| Elective for system change | 319 (9.9) | 450 (10.0) | 554 (9.6) | 2775 (9.9) |
| System change – Haemodynamic reasons | 90 (2.8) | 261 (5.8)c | 398 (6.9) | 1513 (5.4) |
| System change – Electrode problem | 182 (5.6) | 325 (7.2) | 178 (3.1)c | 1003 (3.6) |
| System change – other reasons | 37 (1.1) | 53 (1.2) | 61 (1.1) | 214 (0.8) |
| Normal ‘end of life’, of which: | 908 (28.1) | 1800 (40.1)c | 2678 (46.6)c | 14,077 (50.4) |
| – <7 years, | 908 (28.1) | 1800 (40.1)c | 2678 (46.6)c | 6349 (38.4) |
| – <5 years, | 202 (11.9) | 393 (19.9)c | 587 (23.9) | 1759 (20.5) |
| – <3 years, | 29 (3.1) | 48 (5.1) | 87 (7.3) | 307 (6.6) |
| Premature ‘end of life’ | 77 (2.4) | 140 (3.1) | 126 (2.2) | 520 (1.9) |
| Reason uncoded, unknown or unspecified | 1282 (39.6) | 928 (20.7)c | 1088 (18.9) | 5706 (20.4) |
| Total | 3233 (100.0) | 4489 (100.0) | 5745 (100.0) | 27,937 (100.0) |
aFor each stratum, the duration of follow-up was maximised at 7 years. Hence, the years 2001–2007 were used for follow-up only. Consequently, numbers of the first three columns do not add up to the numbers in the fourth column
bRegardless duration of follow-up
cSignificantly different compared to previous time interval (p ≤ 0.001)
dSee Tab. 1
Comparison of specific replacement or explantation reasons between first pacemakers with NASPE codes VVI/VVIR and DDD/DDDR in the nationwide CPPR-SPRN database, the Netherlands 1984–2007
| VVI/VVIR | DDD/DDDR |
| |||
|---|---|---|---|---|---|
|
| % |
| % | ||
| Number of first implantations of this pacemaker typea | 47,945 | 49.5 | 38,060 | 39.3 | – |
| Number of explantations of first pacemaker of this type | 4391 | 9.2 | 5861 | 15.4 | <0.001 |
|
| |||||
| Recall | 141 | 3.2 | 209 | 3.6 | 0.328 |
| Complication | 226 | 5.1 | 350 | 6.0 | 0.073 |
| Failure | 161 | 3.7 | 154 | 2.6 | 0.003 |
| Premature ‘end of life’ | 123 | 2.8 | 164 | 2.8 | 0.993 |
|
| |||||
| <7 years | 1302 | 29.7 | 2716 | 46.3 | <0.001 |
| <5 years | 273 | 6.2 | 690 | 11.8 | <0.001 |
| <3 years | 53 | 1.2 | 46 | 0.8 | 0.408 |
NASPE North American Society of Pacing and Electrophysiology
a% compared to all first pacemaker implantations between 01.01.1984 and 01.01.2006, n = 96,900