Kimberley Legge1, Peter B Greer1,2, Paul J Keall3, Jeremy T Booth4, Sankar Arumugam5, Trevor Moodie6, Doan T Nguyen3, Jarad Martin2, Daryl John O'Connor1, Joerg Lehmann2.
Abstract
PURPOSE: The Trans-Tasman Radiation Oncology Group (TROG) 15.01 Stereotactic Prostate Adaptive Radiotherapy utilizing Kilovoltage intrafraction monitoring (SPARK) trial is a multicenter trial using Kilovoltage Intrafraction Monitoring (KIM) to monitor prostate position during the delivery of prostate radiation therapy. KIM increases the accuracy of prostate radiation therapy treatments and allows for hypofractionation. However, an additional imaging dose is delivered to the patient. A standardized procedure to determine the imaging dose per frame delivered using KIM was developed and applied at four radiation therapy centers on three different types of linear accelerator.Entities:
Keywords: dose intercomparison; dose measurement; intrafraction motion monitoring; kilovoltage imaging dose
Mesh:
Year: 2017 PMID: 28766904 PMCID: PMC5875840 DOI: 10.1002/acm2.12151
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Experimental set up for ionization chamber measurements showing position of ionization chamber, kV source, and kV detector.
HVL and per‐frame and total treatment course (five fractions) dose values for each Linac obtained using ion chamber measurements. Calculated total dose assumes a 314 s treatment time for each of five fractions delivered. TB refers to Varian Truebeam machines
| Vendor | Linac | kV imaging settings | HVL (mm Al) | Dw,z=0, iso (mGy/frame) | Dw,z=15, iso (mGy/frame) | Dtotal (1 Hz) (mGy) | Dtotal (5 Hz) (mGy) | Dtotal (11 Hz) (mGy) |
|---|---|---|---|---|---|---|---|---|
| Varian | Clinac iX | 125 kV, 1.04 mAs per frame | 4.95 | 0.16 | 0.023 | 35.9 | 179.4 | 394.7 |
| Trilogy | 4.7 | 0.17 | 0.023 | 36.7 | 183.7 | 404.2 | ||
| TB 1 | 4.4 | 0.19 | 0.027 | 41.9 | 209.4 | 460.6 | ||
| TB 2 | 4.65 | 0.20 | 0.028 | 43.7 | 218.6 | 480.9 | ||
| TB 3 | 4.68 | 0.19 | 0.026 | 41.0 | 205.0 | 451.0 | ||
| TB 4 | 4.7 | 0.19 | 0.026 | 41.4 | 206.9 | 455.2 | ||
| Elekta | Synergy | 120 kV, 0.96 mAs per frame | 7.5 | 0.09 | 0.025 | 38.9 | 194.7 | 428.3 |
Estimated dose delivered to a point on the skin
| Vendor | Linac | DSkin (1 Hz) (mGy) | DSkin (5 Hz) (mGy) | DSkin (11 Hz) (mGy) |
|---|---|---|---|---|
| Varian | Clinac iX | 24.5 | 122.3 | 269.0 |
| Trilogy | 25.0 | 125.2 | 275.5 | |
| TB 1 | 28.5 | 142.7 | 313.9 | |
| TB 2 | 29.8 | 149.0 | 327.8 | |
| TB 3 | 27.9 | 139.7 | 307.4 | |
| TB 4 | 28.2 | 141.0 | 310.3 | |
| Elekta | Synergy | 27.6 | 130.2 | 304.1 |