OBJECTIVE: The after reconstitution stability of L1 & L2 is 7 days when stored at 2°-8°c and tightly capped. The total content of the vial is 500 µL and per test requirement is 5 µL. Hence, in 7 days laboratories would consume only 35 µL wasting 365 µL which is 73% of the expensive control samples and such wastage should be ideally prevented. The study of after reconstitution stability proved the stability of the control samples up to 90 days resulting in proper utilization of L1 & L2.Entities:
Keywords: HPLC; HbA1c; decision charts; six sigma; total allowable error
Year: 2017 PMID: 28757822 PMCID: PMC5460012
Source DB: PubMed Journal: EJIFCC ISSN: 1650-3414
HbA1c Peer results of lots 33870, 33890, 33920
| Control | Lot number | Target (%) | Peer SD | Range (±3SDX2) | TAE (%) = Range × 100/ Peer mean | ||||
|---|---|---|---|---|---|---|---|---|---|
| Level | Peer | Insert | Peer | Insert | Peer | Insert | Peer | Insert | |
| L1 | 33870 | 5.38 | 5.4 | 0.184 | 0.2 | 1.104 | 1.2 | 20.0 | 22.22 |
| L2 | 33870 | 9.79 | 9.9 | 0.257 | 0.4 | 1.542 | 2.4 | 15.6 | 24.24 |
| L1 | 33890 | 5.48 | 5.3 | 0.166 | 0.2 | 0.996 | 1.2 | 18.2 | 22.64 |
| L2 | 33890 | 9.87 | 9.8 | 0.245 | 0.4 | 1.47 | 2.4 | 14.9 | 24.5 |
| L1 | 33920 | 5.42 | 5.4 | 0.184 | 0.2 | 1.104 | 1.2 | 20.0 | 22.22 |
| L2 | 33920 | 9.84 | 9.8 | 0.328 | 0.4 | 1.968 | 2.4 | 20.0 | 24.5 |
Statistical evaluation of HbA1c values of Lot 33870
| N | Mean HbA1c (%) | SD | CV% | Deviation from peer mean | % Bias | Tests per vial | Sigma | z-score | Allowable bias | Allowable imprecision | Critical error |
|---|---|---|---|---|---|---|---|---|---|---|---|
| 182 | 5.47 | 0.127 | 2.32 | 0.09 | 1.67 | 45+46 | 8.0 | 0.49 | 8.3 | 11.6 | -0.641 |
| 182 | 9.52 | 0.191 | 2.01 | -0.27 | 2.96 | 45+46 | 6.3 | -1.13 | 19 | 13 | -1.911 |
*Allowable Bias- Normal Operational Specifications along Y-axis. Calculated as: %Bias*100/TAE (%)
*Allowable Imprecision- Normal Operational Specifications along X-axis. Calculated as: CV%*100/TAE (%)
N- Number of tests performed
Critical error – Calculations obtained from “Quality Control Grid Calculator”. Critical error
< ±2.0 shows excellent performance
Statistical evaluation of Lot 33920 (Laboratory 1)
| N | Mean HbA1c (%) | SD | CV% | Deviation from peer mean | % Bias | Tests per vial | Sigma | z-score | Allowable bias* | Allowable imprecision* | Critical error* |
|---|---|---|---|---|---|---|---|---|---|---|---|
| 135 | 5.49 | 0.17 | 2.5 | 0.07 | 1.3 | 45+45 | 7.5 | 0.27 | 6.5 | 12.5 | -0.569 |
| 135 | 9.97 | 0.203 | 2.04 | 0.13 | 1.32 | 45+45 | 9.2 | 0.39 | 6.6 | 10.2 | -0.336 |
Statistical evaluation of Lot 33890
| N | Mean HbA1c (%) | SD | CV% | Deviation from peer mean | % Bias | Tests per vial | Sigma | z-score | Allowable bias* | Allowable imprecision* | Critical error* |
|---|---|---|---|---|---|---|---|---|---|---|---|
| 180 | 5.56 | 0.127 | 2.285 | 0.08 | 1.46 | 45+45 | 7.3 | 0.48 | 8 | 12.5 | -0.834 |
| 183 | 9.67 | 0.262 | 2.2 | -0.20 | 2.02 | 46+46 | 5.9 | -0.82 | 13.6 | 14.7 | -1.337 |
Statistical evaluation of Lot 33920 (Laboratory 2)
| N | Mean HbA1c (%) | SD | CV% | Deviation from peer mean | % Bias | Tests per vial | Sigma | z-score | Allowable bias* | Allowable imprecision* | Critical error* |
|---|---|---|---|---|---|---|---|---|---|---|---|
| 135 | 5.55 | 0.23 | 2.2 | 0.13 | 2.4 | 45+45 | 8.0 | 0.5 | 12 | 11 | -0.923 |
| 135 | 10.18 | 0.206 | 2.02 | 0.34 | 3.45 | 45+45 | 8.2 | 1.03 | 17.2 | 10.1 | -1.37 |