Literature DB >> 28740145

Validation of asynchronous quantitative bone densitometry of the spine: Accuracy, short-term reproducibility, and a comparison with conventional quantitative computed tomography.

Ling Wang1, Yongbin Su1, Qianqian Wang2, Yangyang Duanmu3, Minghui Yang4, Chen Yi4, Xiaoguang Cheng5.   

Abstract

Asynchronous calibration quantitative computed tomography (QCT) is a new tool that allows the quantification of bone mineral density (BMD) without the use of a calibration phantom during scanning; however, this tool is not fully validated for clinical use. We used the European spine phantom (ESP) with repositioning during scanning and assessed the accuracy and short-term reproducibility of asynchronous QCT. Intra-scanner and intra-observer precision were each calculated as the root mean square of the standard deviation (RMSSD) and the coefficient of variation (CV-RMSSD). We also compared asynchronous and conventional QCT results in 50 clinical subjects. The accuracy of asynchronous QCT for three ESP vertebrae ranged from 1.4-6.7%, whereas intra-scanner precision for these vertebrae ranged from 0.53-0.91 mg/cc. Asynchronous QCT was most precise for a trabecular BMD of 100 mg/cc (CV-RMSSD = 0.2%). For intra-observer variability, overall precision error was smaller than 3%. In clinical subjects there was excellent agreement between the two calibration methods with correlation coefficients ranging from 0.96-0.99. A Bland-Altman analysis demonstrated that methodological differences depended on the magnitude of the BMD variable. Our findings indicate that the asynchronous QCT has good accuracy and precision for assessing trabecular BMD in the spine.

Entities:  

Mesh:

Year:  2017        PMID: 28740145      PMCID: PMC5524691          DOI: 10.1038/s41598-017-06608-y

Source DB:  PubMed          Journal:  Sci Rep        ISSN: 2045-2322            Impact factor:   4.379


Introduction

Bone mineral density is a surrogate indicator of bone strength that plays an important role in the management of osteoporosis and related fractures[1, 2]. Different from areal bone mineral density computed by dual-energy X-ray absorptiometry (DXA), bone mineral density (BMD) derived from quantitative computed tomography (QCT) is a volumetric measure of the vertebral trabecular bone. Given the high turnover rate of trabecular bone compared to cortical bone, BMD calculated from QCT offers substantially higher sensitivity. Yet, radiation doses associated with CT limit the application of QCT in osteoporosis screening. Recently, asynchronous calibration QCT was described as a new tool for quantifying BMD[3]. Asynchronous calibration means that a calibration phantom is not necessary during QCT scanning. Rather, asynchronous QCT utilizes phantom data obtained separately from CT scans to calibrate data in Hounsfield units for the measurement of BMD. This approach is convenient for the assessment of BMD during routine abdominal and/or lung CT scans, and for the identification of patients who have an increased risk of fracture with diagnostic CT scanning. The 2015 International Society for Clinical Densitometry (ISCD) official position states that the in-scan calibration phantom for density-based QCT measurement can be replaced with asynchronous calibration if scanner stability is maintained[4, 5]. Opportunistic screening enabled by the use of asynchronous calibration may therefore improve the current understanding of bone health status and decrease the number of undiagnosed or overlooked cases of osteoporosis. Few studies to date have compared conventional QCT, asynchronous QCT, and DXA[3, 6–8]. Moreover, asynchronous QCT has not yet been fully validated for the clinical measurement of spinal BMD. Although an early study by Brown et al. reported the precision of asynchronous QCT using the Mindways QA phantom, this study did not perform an accuracy assessment using the European spine phantom (ESP), which is a standard evaluation tool for bone densitometry. Thus, we performed 10 ESP scans with repositioning in order to assess the accuracy and short-term reproducibility of asynchronous quantitative bone densitometry, and to compare spine BMD results of asynchronous and conventional QCT in 50 clinical subjects.

Methods

Evaluation of accuracy and short-term reproducibility with ESP

The ESP (QRM, Erlangen, Germany) was recommended by the International DXA Standardization Committee as a possible standard for use in DXA as well as QCT[9]. The ESP consists of three simulated vertebrae that are designed to give trabecular density values of 50, 100, and 200 mg/cm3 of hydroxyapatite, respectively. To simulate slight differences seen on daily quality control (QC) charts due to repositioning of the phantom, we performed each phantom scan with repositioning of the ESP. This method was recommended in the 2005 ISCD Official Position and reported to improve the estimate of the mean measure by a factor of nearly three[10]. Briefly, the ESP was placed on the scanner table on top of the table pad and aligned along the long axis of the table. CT scanning parameters were summarized in Table 1. Images were transferred to a QCT workstation and analysed using the 3D spine function version 5.10 of Mindways QCT pro software (Mindways Software Inc., Austin, TX, USA). This novel version includes a conventional QCT module and an asynchronous calibration module. For asynchronous QCT measurements, a new Model 4 asynchronous calibration phantom (Mindways Software Inc.) (Fig. 1) was scanned for quality assurance (QA) calibration. A Model 3 conventional calibration phantom (Mindways Software Inc.) was used to validate QA calibration for conventional QCT analysis. To evaluate short-term intra-scanner precision, a ten-scan series was performed in duplicate over two sessions, with one month between each scanning session. For the evaluation and comparison of asynchronous and conventional QCT, a Toshiba Aquilion 80-slices CT scanner (Toshiba Medical Systems Corp., Tokyo, Japan) was used to scan the ESP (no. 145) in the presence and absence of a Mindways calibration phantom, ten times each. Each group of ten scans was analysed individually and the average of each parameter was used to compute accuracy. To further compare of the effect of inter-scanner differences, The ESP was also scanned in 16-slice CT scanner for ten times with a calibration phantom. Based on the BMD measured using conventional QCT on each scanner in the different ESP regions, the interchangeability of the two CT systems was showed. Scan parameters of the two CT scanners were summarized on Table 1.
Table 1

Computed tomography (CT) scan parameters for phantom and patient image acquisition.

European spine phantomPatient spine
Scanner type80-slice Toshiba Aquilion16-slice Toshiba Aquilion
Voltage (kV)120120
Exposure (mAs)100Auto exposure
Reconstruction kernelStandardStandard
DFOV (mm)400400
Slice thickness (mm)1.01.0
Table height (cm)12090

DFOV, display field of view.

Figure 1

Quality assurance (QA) for asynchronous quantitative computed tomography using the Model 4 asynchronous QA phantom.

Computed tomography (CT) scan parameters for phantom and patient image acquisition. DFOV, display field of view. Quality assurance (QA) for asynchronous quantitative computed tomography using the Model 4 asynchronous QA phantom.

Subjects and clinical data

Fifty subjects (25 female subjects and 25 male subjects) were retrospectively enrolled in August 2016, one week before the date of Model4 Quality Assurance Scanning, from the emergency room radiology department of Beijing Jishuitan Hospital. In our emergency room radiology department, there are 10 to 20 clinical spine QCT scans on each day. For inclusion, eligible patients had to have conventional spine QCT BMD examinations within one week to avoid uncertainty regarding calibration stability in retrospective measurements. The exclusion criteria included fixation of the spine affecting image quality and more than two vertebrae compressive fractures. The 3D spine function version 5.10 of Mindways QCT pro software (Mindways Software Inc., Austin, TX, USA) was used to perform L1–L3 BMD analysis of all 50 cases, once with the conventional QCT module and once with the asynchronous calibration module. To further assess intra-observer precision, all 50 cases were analysed with the asynchronous QCT module two times by one radiologist, with one week between each analysis. This study was approved by the Ethics Committee of Beijing Jishuitan Hospital, Xicheng District, Beijing, China, and all methods were performed in accordance with the approved guidelines, and all the subjects signed informed consents.

Image acquisition

For ESP, QCT scans were acquired using a Toshiba Aquilion 80-slice CT scanner (Toshiba Medical Systems Corp., Tokyo, Japan) with and without a calibration phantom (Mindways Software Inc., Austin, TX, USA). For clinical assessments, spine QCT scans were acquired using a Toshiba Aquilion 16-slice CT scanner (Toshiba Medical Systems Corp., Tokyo, Japan) with a calibration phantom (Mindways Software Inc., Austin, TX, USA). As the clinical scans were retrospectively from patients who aimed to have conventional QCT examinations, in our department the clinical QCT scans were all performed by 16-slice CT scanner in the emergency room radiology department. The reason for performing ESP scans on 80-slice CT scanner is that we want to investigate the intra-scanner precision on another scanner to show the equivalence. Scan parameters for both CT scanners were as follows: 120 kVp, 125 mAs, 1-mm slice thickness, 50-cm scan field of view, and a matrix size of 512 × 512 in spiral and standard reconstructions.

Statistics analysis

The accuracies of conventional and asynchronous QCT for ESP scans were assessed by calculating the average of percentages of difference between measured values and expected values given by the manufacturer. Mann-Whitney U tests were used to compare values acquired using the two methods and to compare the interchangeability of the two CT scanners systems. A linear regression analysis was used to compare the density values acquired using the two methods, and a Bland–Altman analysis was conducted to assess differences between the conventional and asynchronous QCT measurements. As per the ISCD recommendation, short-term precision (intra-observer and intra-scanner precision) was calculated in terms of RMSSD and CV-RMSSD values.

Data Availability

The datasets generated during and/or analyzed during the current study are available from the corresponding author on reasonable request.

Results

Accuracy

The results of the accuracy analysis based on 10 ESP scans with repositioning are summarized in Table 2. The accuracies of conventional and asynchronous QCT were different, ranging from 3.7–5.9% for conventional QCT and from 1.4–6.7% for asynchronous QCT. Asynchronous QCT had higher accuracy than conventional QCT for measuring trabecular BMD in 50 and 100 mg/cc ESP (L1 and L2, respectively) vertebrae, whereas conventional QCT showed better accuracy for measuring the 200 mg/cc ESP L3 vertebra. Conventional QCT tended to overestimate ESP trabecular bone, whereas asynchronous QCT underestimated the BMD of the ESP L3 vertebra. The mean trabecular volumetric BMD (vBMD) for all three ESP vertebrae using conventional QCT was significantly higher than that obtained using asynchronous QCT. Figure 2 shows the results of ten scans performed with asynchronous QCT and conventional QCT; consistency was excellent for both methods.
Table 2

The accuracies of asynchronous and conventional quantitative computed tomography (QCT) and comparisons of inter-scanner differences.

Accuracies of asynchronous and conventional QCT using 80-slice CT scannerInter-scanner differences
SiteESP BMD (mg/cm3)Conv. values (mg/cm3)REAsyc. values (mg/cm3)REConv. − Asyc. (mg/cm3)P-valueConv. values on 16-slice scanner (mg/cm3)Diff. (mg/cm3)P-value
L15052.94 ± 0.995.9%51.89 ± 0.533.8%1.05<0.000152.14 ± 1.450.800.66
L2100104.18 ± 0.854.2%101.40 ± 0.611.4%2.78<0.0001105.00 ± 0.37−0.820.37
L3200207.40 ± 0.973.7%186.70 ± 0.536.7%20.710.0029204.85 ± 1.352.550.023

Asyc., asynchronous QCT method; BMD, bone mineral density; Conv., conventional QCT; ESP, European spine phantom; RE, Relative error; Diff. difference defined as 80-slice scanner – 16-slice scanner. Data represent the mean ± standard deviation.

Figure 2

Ten-scan results of asynchronous and conventional quantitative computed tomography (QCT) for ESP study. Asyc., asynchronous QCT method; Conv., conventional QCT method.

The accuracies of asynchronous and conventional quantitative computed tomography (QCT) and comparisons of inter-scanner differences. Asyc., asynchronous QCT method; BMD, bone mineral density; Conv., conventional QCT; ESP, European spine phantom; RE, Relative error; Diff. difference defined as 80-slice scanner – 16-slice scanner. Data represent the mean ± standard deviation. Ten-scan results of asynchronous and conventional quantitative computed tomography (QCT) for ESP study. Asyc., asynchronous QCT method; Conv., conventional QCT method.

Intra-scanner short-term reproducibility

The intra-scanner precision of asynchronous QCT for vBMD across all three vertebrae ranged from 0.53–0.91 mg/cc. Asynchronous QCT was most precise (in terms of the root mean square of the standard deviation [RMSSD]) for the 100 mg/cc ESP L2 vertebra with a coefficient of variance (CV)-RMSSD value of 0.2%.

Clinical study

Table 3 shows that among 25 male subjects and 25 female subjects, the mean age was 62 ± 9 years (range, 33–85 years), the mean height was 165.9 ± 7.9 cm (range, 150.0–181.0 cm), the mean weight was 70.7 ± 9.8 kg (range, 50.0–89.5 kg), and the mean BMI was 25.6 ± 2.7 kg/m2 (range, 20.4–32.8 kg/m2). Of the 50 subjects, twelve female subjects and 10 male subjects had osteoporosis, and two patients had other vertebrae compressive fractures.
Table 3

Characteristics of clinical subjects with QCT scans acquired using a 16-slice CT scanner.

Male subjects (n = 25)Female subjects (n = 25)Total subjects (n = 50)
Age (years)62 ± 1162 ± 862 ± 9
Weight (kg)76.1 ± 7.065.1 ± 9.170.7 ± 9.8
Height (cm)170.6 ± 5.7161.0 ± 6.0165.9 ± 7.9
BMI (kg/m2)26.2 ± 1.925.1 ± 3.325.6 ± 2.7
Osteoporosis10/2512/2522/50

BMI, body mass index. Data represent the mean ± standard deviation.

Characteristics of clinical subjects with QCT scans acquired using a 16-slice CT scanner. BMI, body mass index. Data represent the mean ± standard deviation.

Relationships between conventional and asynchronous QCT scanning results

Figure 3 shows correlations between the results of conventional and asynchronous QCT for each vertebra (L1–L3) and the average of all 3 vertebrae of the clinical subjects. The analysis demonstrated excellent agreement between the two methods, with correlation coefficients (R2) ranging from 0.96–0.99. A Bland–Altman analysis (small upper panel inserts in Fig. 3) showed differences between conventional and asynchronous QCT for different vertebrae with respect to a linear correlation. Methodological differences (expressed as absolute differences) depended on the magnitude of the BMD variable.
Figure 3

Correlation scatter plots for conventional and asynchronous quantitative computed tomography (QCT) outcomes of clinical subjects and Bland–Altman plots (small upper panel inserts) for conventional and asynchronous QCT BMD measurements of clinical subjects. Small upper inserts: Y-axis (Diff) is defined as (Asyc. −Conv.); X-axis is defined as the mean value of (Asyc. +Conv.).

Correlation scatter plots for conventional and asynchronous quantitative computed tomography (QCT) outcomes of clinical subjects and Bland–Altman plots (small upper panel inserts) for conventional and asynchronous QCT BMD measurements of clinical subjects. Small upper inserts: Y-axis (Diff) is defined as (Asyc. −Conv.); X-axis is defined as the mean value of (Asyc. +Conv.).

Intra-observer variability

Table 4 summarizes the RMMS and CV-RMSS values for repeated measures with the asynchronous calibration method. The intra-observer precision for asynchronous QCT measured across all three ESP vertebrae ranged from 2.68–3.80 mg/cc. Overall precision error for BMD was smaller than 3%, consistent with the known precision of QCT technology.
Table 4

Intra-observer and intra-scanner reproducibility of asynchronous quantitative computed tomography (QCT).

Intra-observer variabilityIntra-scanner variability
Subjects (n = 50)Obs. 1Obs. 2RMSSDCV-RMSSDESP scans (n = 10)Scan 1Scan 2RMSSDCV-RMSSD
L1 vBMD93.04 ± 34.5894.02 ± 35.143.122.5L1 vBMD51.89 ± 0.5351.79 ± 0.200.560.7
L2 vBMD86.76 ± 34.0187.64 ± 34.433.732.6L2 vBMD101.40 ± 0.61101.67 ± 0.400.910.6
L3 vBMD81.44 ± 33.4281.39 ± 33.512.682.2L3 vBMD186.70 ± 0.53186.55 ± 0.460.530.2

CV, coefficient of variation; Obs., observation; RRMSD, root mean square of the standard deviation; vBMD, volumetric bone mineral density. Data represent the mean ± standard deviation.

Intra-observer and intra-scanner reproducibility of asynchronous quantitative computed tomography (QCT). CV, coefficient of variation; Obs., observation; RRMSD, root mean square of the standard deviation; vBMD, volumetric bone mineral density. Data represent the mean ± standard deviation.

Discussion

In this study, we determined the accuracy and intra-operator precision of asynchronous QCT using an ESP vertebrae dataset and data retrospectively collected from clinical subjects. In both datasets, the precision of BMD measurement was around 3%. Our results suggest that vertebral trabecular BMD measurements can be conveniently obtained by asynchronous QCT with good accuracy and precision. The results of an analysis based on ten ESP scans demonstrated that asynchronous QCT had excellent accuracy for measuring trabecular BMD and agreed well with conventional QCT. Asynchronous QCT had slightly higher accuracy than conventional QCT for measuring trabecular BMD in 50 and 100 mg/cc ESP vertebrae; however, the differences between the two methods for 50 and 100 mg/cc vertebrae were relatively small (1.05 mg/cc and 2.78 mg/cc, respectively). In contrast, conventional QCT was more accurate for measuring the 200 mg/cc ESP vertebra, suggesting that the asynchronous QCT method may underestimate the BMD of high-density bone. While traditional QCT utilizes a calibration equation obtained from each individual slice, asynchronous QCT uses the same calibration equation for all data. This may result in lower variability in high-BMD bone for asynchronously calibrated results. Inconsistent with our finding, Brown et al. showed excellent agreement between conventional and asynchronous QCT using high-BMD Mindways QA phantom scan data[3]. The reason for this discrepancy is unclear; however, one explanation may be related to the fact that peak histogram values used in the calibration procedure depend on noise and thus exposure settings, the reconstruction kernel, and slice thickness[5]. True in vivo precision measurements were obtained by measuring subjects twice at a one-month interval with repositioning. Repositioning has a lower impact on precision in three-dimensional QCT than in traditional two-dimensional slice-based imaging, where the location of the slice relative to the vertebral body is determined at the time of acquisition using the scout view taken before the actual CT scan. In an earlier study, Brown et al.[3] investigated inter-observer variability for the asynchronous method using data from 43 patients aged 63.8 ± 8.6 years and reported RMSSD and CV-RMSSD values of 4.34 mg/cc and 3.67%, respectively; bias was not considered to be clinically important in the context of osteoporosis screening. In our study, we assessed the intra-observer reproducibility for asynchronous QCT and calculated RMSSD values of 3.12, 3.73, and 2.68 mg/cc and CV-RMSSD values of 2.5%, 2.6%, and 2.2% for the ESP L1–L3 vertebrae, respectively. Inter-observer reproducibility for asynchronous QCT reported by Brown et al. is therefore similar to the calculated intra-observer variability in our study. In one phantomless QCT BMD study, inter-observer variability was 3.1 mg/cc and CV-RMSSD was 4.0%[11], which are high compared to the results of our study. Although previous reports have indicated that non-calibrated Hounsfield unit values from CT scanners may be used for the opportunistic screening of low bone mass[11-13], the use of a phantom calibration standard guarantees that the derived BMD computations will be consistent across CT scanners from different manufacturers and consistent across different scanning X-ray energy levels. Lastly, we found that asynchronous and synchronous QCT results were highly correlated; regression lines for each method were not significantly different. Accordingly, it appears that bias (expressed as the absolute difference between method results) depends on the magnitude of the BMD variable. Different from conventional QCT utilizing a calibration equation obtained from each individual slice, the asynchronous QCT uses the same calibration equation for all slices and could have such calibration before or after CT images acquisition. The different calibration methods might be one of important causes of the bias. Further, there might be a phantom-induced bias between asynchronous and conventional QCT. Brown et al. had investigated the phantom-induced bias on clinical individuals and found that the bias induced by the presence of the phantom was 2.3 mg/cm3 when asynchronous calibration was applied[3]. The present study had several limitations that most notably affected our precision assessments. First, due to concerns about radiation doses, it was not appropriate to scan patients twice in one session, so we performed ESP scans to investigate intra-scanner bias; however, this approach does not meet conservative recommendations (>27 degree of freedom, DOF) and may have underestimated precision error[14]. Second, we did not analyse the long-term precision of asynchronous QCT; this should be investigated in future studies. Lastly, the BMD of volunteers recruited in precision studies is often normal. In this study, we selected patients from a population with low-to-normal BMD. Given that BMD typically decreases with age, precision errors are typically larger in elderly individuals[15]. Additionally, precision errors are typically estimated by scanning a subject twice on the same day with repositioning in between scans and all scans performed by the same operator. Instead, we report precision using repeat scans taken at a one-month interval. We believe that these aspects of our study provide a more accurate reflection of the real precision of asynchronous QCT measurement. Further studies are necessary to confirm and extend our results. In conclusion, we demonstrate that asynchronous QCT could be used for spine BMD screening based on the results of accuracy assessment for volumetric trabecular BMD in the spine and good short-term precision; however, even when the asynchronous and synchronous methods are highly correlated, the presence of bias was observed.
  15 in total

1.  Clinical Use of Quantitative Computed Tomography-Based Advanced Techniques in the Management of Osteoporosis in Adults: the 2015 ISCD Official Positions-Part III.

Authors:  Klaus Engelke; Thomas Lang; Sundeep Khosla; Ling Qin; Philippe Zysset; William D Leslie; John A Shepherd; John T Shousboe
Journal:  J Clin Densitom       Date:  2015 Jul-Sep       Impact factor: 2.617

2.  European semi-anthropomorphic spine phantom for the calibration of bone densitometers: assessment of precision, stability and accuracy. The European Quantitation of Osteoporosis Study Group.

Authors:  J Pearson; J Dequeker; M Henley; J Bright; J Reeve; W Kalender; A M Laval-Jeantet; P Rüegsegger; D Felsenberg; J Adams
Journal:  Osteoporos Int       Date:  1995-05       Impact factor: 4.507

3.  Opportunistic Quantitative CT Bone Mineral Density Measurement at the Proximal Femur Using Routine Contrast-Enhanced Scans: Direct Comparison With DXA in 355 Adults.

Authors:  Timothy J Ziemlewicz; Alyssa Maciejewski; Neil Binkley; Alan D Brett; J Keenan Brown; Perry J Pickhardt
Journal:  J Bone Miner Res       Date:  2016-05-06       Impact factor: 6.741

4.  Asynchronously Calibrated Quantitative Bone Densitometry.

Authors:  J K Brown; W Timm; G Bodeen; A Chason; M Perry; F Vernacchia; R DeJournett
Journal:  J Clin Densitom       Date:  2016-01-15       Impact factor: 2.617

5.  Forearm bone mineral density by age in 7,620 men and women: the Tromsø study, a population-based study.

Authors:  G K Berntsen; V Fønnebø; A Tollan; A J Søgaard; J H Magnus
Journal:  Am J Epidemiol       Date:  2001-03-01       Impact factor: 4.897

6.  Direct Comparison of Unenhanced and Contrast-Enhanced CT for Opportunistic Proximal Femur Bone Mineral Density Measurement: Implications for Osteoporosis Screening.

Authors:  Timothy J Ziemlewicz; Alyssa Maciejewski; Neil Binkley; Alan D Brett; J Keenan Brown; Perry J Pickhardt
Journal:  AJR Am J Roentgenol       Date:  2016-02-11       Impact factor: 3.959

7.  Phantom-less QCT BMD system as screening tool for osteoporosis without additional radiation.

Authors:  Dirk K Mueller; Alex Kutscherenko; Hans Bartel; Alain Vlassenbroek; Petr Ourednicek; Joachim Erckenbrecht
Journal:  Eur J Radiol       Date:  2010-03-12       Impact factor: 3.528

Review 8.  Epidemiology and outcomes of osteoporotic fractures.

Authors:  Steven R Cummings; L Joseph Melton
Journal:  Lancet       Date:  2002-05-18       Impact factor: 79.321

9.  Opportunistic screening for osteoporosis using abdominal computed tomography scans obtained for other indications.

Authors:  Perry J Pickhardt; B Dustin Pooler; Travis Lauder; Alejandro Muñoz del Rio; Richard J Bruce; Neil Binkley
Journal:  Ann Intern Med       Date:  2013-04-16       Impact factor: 25.391

10.  Bone Mineral Density Estimations From Routine Multidetector Computed Tomography: A Comparative Study of Contrast and Calibration Effects.

Authors:  Johannes Kaesmacher; Hans Liebl; Thomas Baum; Jan Stefan Kirschke
Journal:  J Comput Assist Tomogr       Date:  2017 Mar/Apr       Impact factor: 1.826

View more
  14 in total

1.  Highlights of the annual scientific meeting of the 24th congress of the European Society of Musculoskeletal Radiology (ESSR) 2017.

Authors:  Carmelo Messina; Luca Maria Sconfienza; Giuseppe Guglielmi
Journal:  Skeletal Radiol       Date:  2017-09-20       Impact factor: 2.199

Review 2.  X-ray-based quantitative osteoporosis imaging at the spine.

Authors:  M T Löffler; N Sollmann; K Mei; A Valentinitsch; P B Noël; J S Kirschke; T Baum
Journal:  Osteoporos Int       Date:  2019-11-14       Impact factor: 4.507

3.  One Novel Phantom-Less Quantitative Computed Tomography System for Auto-Diagnosis of Osteoporosis Utilizes Low-Dose Chest Computed Tomography Obtained for COVID-19 Screening.

Authors:  Tang Xiongfeng; Zhang Cheng; He Meng; Ma Chi; Guo Deming; Qi Huan; Chen Bo; Yang Kedi; Shen Xianyue; Wong Tak-Man; Lu William Weijia; Qin Yanguo
Journal:  Front Bioeng Biotechnol       Date:  2022-06-28

4.  The influence of CT and dual-energy X-ray absorptiometry (DXA) bone density on quantitative [18F] sodium fluoride PET.

Authors:  Inderbir S Jassel; Musib Siddique; Michelle L Frost; Amelia E B Moore; Tanuj Puri; Glen M Blake
Journal:  Quant Imaging Med Surg       Date:  2019-02

5.  Fully automated radiomic screening pipeline for osteoporosis and abnormal bone density with a deep learning-based segmentation using a short lumbar mDixon sequence.

Authors:  Yinxia Zhao; Tianyun Zhao; Shenglan Chen; Xintao Zhang; Mario Serrano Sosa; Jin Liu; Xianfu Mo; Xiaojun Chen; Mingqian Huang; Shaolin Li; Xiaodong Zhang; Chuan Huang
Journal:  Quant Imaging Med Surg       Date:  2022-02

6.  Automatic opportunistic osteoporosis screening using low-dose chest computed tomography scans obtained for lung cancer screening.

Authors:  Yaling Pan; Dejun Shi; Hanqi Wang; Tongtong Chen; Deqi Cui; Xiaoguang Cheng; Yong Lu
Journal:  Eur Radiol       Date:  2020-02-19       Impact factor: 5.315

7.  Patient-Specific Phantomless Estimation of Bone Mineral Density and Its Effects on Finite Element Analysis Results: A Feasibility Study.

Authors:  Young Han Lee; Jung Jin Kim; In Gwun Jang
Journal:  Comput Math Methods Med       Date:  2019-01-03       Impact factor: 2.238

8.  Improved prediction of incident vertebral fractures using opportunistic QCT compared to DXA.

Authors:  Maximilian T Löffler; Alina Jacob; Alexander Valentinitsch; Anna Rienmüller; Claus Zimmer; Yu-Mi Ryang; Thomas Baum; Jan S Kirschke
Journal:  Eur Radiol       Date:  2019-02-21       Impact factor: 5.315

9.  The Association of Lumbar Disc Herniation with Lumbar Volumetric Bone Mineral Density in a Cross-Sectional Chinese Study.

Authors:  Jian Geng; Ling Wang; Qing Li; Pengju Huang; Yandong Liu; Glen M Blake; Wei Tian; Xiaoguang Cheng
Journal:  Diagnostics (Basel)       Date:  2021-05-24

Review 10.  Opportunistic diagnosis of osteoporosis, fragile bone strength and vertebral fractures from routine CT scans; a review of approved technology systems and pathways to implementation.

Authors:  Veena Aggarwal; Christina Maslen; Richard L Abel; Pinaki Bhattacharya; Paul A Bromiley; Emma M Clark; Juliet E Compston; Nicola Crabtree; Jennifer S Gregory; Eleni P Kariki; Nicholas C Harvey; Kate A Ward; Kenneth E S Poole
Journal:  Ther Adv Musculoskelet Dis       Date:  2021-07-10       Impact factor: 5.346

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.