| Literature DB >> 28725445 |
Espen Rostrup Nakstad1,2, Helge Opdahl1, Fridtjof Heyerdahl3, Fredrik Borchsenius2, Ole Henning Skjønsberg2.
Abstract
INTRODUCTION: Removal of pulmonary secretions in mechanically ventilated patients usually requires suction with closed catheter systems or flexible bronchoscopes. Manual ventilation is occasionally performed during such procedures if clinicians suspect inadequate ventilation. Suctioning can also be performed with the ventilator entirely disconnected from the endotracheal tube (ETT). The aim of this study was to investigate if these two procedures generate negative airway pressures, which may contribute to atelectasis.Entities:
Keywords: Open suction procedure; airway pressure; bronchoscopy; closed catheter system; end-expiratory pressure; endotracheal suctioning; manual ventilation; peak pressure
Year: 2017 PMID: 28725445 PMCID: PMC5501241 DOI: 10.1136/bmjresp-2016-000176
Source DB: PubMed Journal: BMJ Open Respir Res ISSN: 2052-4439
Figure 1Test design: design and connections of the mechanical lung model test system. ETT, endotracheal tube.
Figure 2Manual ventilation versus open suction procedure. Pressures distal to ETT 7 mm during bronchoscopic suctioning: changes in peak, mean and end-expiratory model airway pressures (PPEAK MA, PMEAN MA and PEEP MA) distal to an ETT with 7 mm inner diameter during bronchoscope insertion (16 Fr) and maximum suction (−765 cm H2O (−75 kPa)) are shown. For comparison, unpublished tracings from a previous study of mechanical ventilation are shown below. The total effect of suctioning compared with normal ventilation is shown as Δ PPEAK MA and Δ PEEP MA. Note that during the suction procedure, PMEAN MA is −10 cm H2O in manual ventilation, −18 cm H2O with the ETT disconnected and 7 cm H2O during mechanical volume-controlled ventilation. ETT, endotracheal tube.
Figure 3Suction pressure variation in different ETT sizes: model airway pressures are recorded as described in figure 3 using both bronchoscopic and closed catheter suction systems during manual ventilation (left panel) and with the ETT disconnected (right panel). Each manual ventilation procedure was performed with both moderate and maximum suction levels. ETT, endotracheal tube.
Effects of ETT obstruction and suction pressure
| Open suction procedure | Before suction | Moderate suction | Maximum suction | |
| PMEAN MA
| PMEAN MA
| PMEAN MA
| ||
| ETT 9 | Catheter 12 Fr | 0 | −0.5 | −1.5 |
| Catheter 14 Fr | 0 | −1 | −2 | |
| Catheter 16 Fr | 0 | −2 | −3 | |
| Br.scope 16 Fr | 0 | −1.5 | −2 | |
| ETT 8 | Catheter 12 Fr | 0 | −1 | −2 |
| Catheter 14 Fr | 0 | −2 | −4 | |
| Catheter 16 Fr | 0 | −3 | −7 | |
| Br.scope 16 Fr | 0 | −2 | −5 | |
| ETT 7 | Catheter 12 Fr | 0 | −2 | −5 |
| Catheter 14 Fr | 0 | −4 | −10 | |
| Catheter 16 Fr | 0 | −9 | −23 | |
| Br.scope 16 Fr | 0 | −8 | −18 | |
The effects of altered ETT diameter relative to suction devices and vacuum pressures during open suction procedure are shown. Pressures are measured distal to ETTs, with the ETT disconnected from the test model. Pressures did not change during 30 s of suctioning or when recording was repeated; pressure changes induced by moderate and maximum suction were significant for each permutation. 16 Fr catheters are normally not used clinically in ETT 7 mm ID, but the measurement is included for comparative reasons.
ETT, endotracheal tube.