James W Snider1, Yildirim Mutaf2, Elizabeth Nichols2, Andrea Hall1, Patrick Vadnais1, William F Regine2, Steven J Feigenberg2.
Abstract
Accelerated partial breast irradiation has caused higher than expected rates of poor cosmesis. At our institution, a novel breast stereotactic radiotherapy device has demonstrated dosimetric distributions similar to those in brachytherapy. This study analyzed comparative dose distributions achieved with the device and intensity-modulated radiation therapy accelerated partial breast irradiation. Nine patients underwent computed tomography simulation in the prone position using device-specific immobilization on an institutional review board-approved protocol. Accelerated partial breast irradiation target volumes (planning target volume_10mm) were created per the National Surgical Adjuvant Breast and Bowel Project B-39 protocol. Additional breast stereotactic radiotherapy volumes using smaller margins (planning target volume_3mm) were created based on improved immobilization. Intensity-modulated radiation therapy and breast stereotactic radiotherapy accelerated partial breast irradiation plans were separately generated for appropriate volumes. Plans were evaluated based on established dosimetric surrogates of poor cosmetic outcomes. Wilcoxon rank sum tests were utilized to contrast volumes of critical structures receiving a percentage of total dose (Vx). The breast stereotactic radiotherapy device consistently reduced dose to all normal structures with equivalent target coverage. The ipsilateral breast V20-100 was significantly reduced (P < .05) using planning target volume_10mm, with substantial further reductions when targeting planning target volume_3mm. Doses to the chest wall, ipsilateral lung, and breast skin were also significantly lessened. The breast stereotactic radiotherapy device's uniform dosimetric improvements over intensity-modulated accelerated partial breast irradiation in this series indicate a potential to improve outcomes. Clinical trials investigating this benefit have begun accrual.Entities:
Keywords: breast cancer; dosimetry; partial breast irradiation; stereotactic ablative radiotherapy; stereotactic body radiotherapy
Year: 2017 PMID: 28705082 PMCID: PMC5762064 DOI: 10.1177/1533034617718961
Source DB: PubMed Journal: Technol Cancer Res Treat ISSN: 1533-0338
Figure 1.The GammaPod device with device-specific prone patient loader (A). The internal device components including the rotating source carrier and associated collimators (B). The breast immobilization cup system incorporating 2 layered cups with a stereotactic wire fiducial system, negative pressure valve, and silicone flange (C). The breast cup system interlocks with the patient loader. The dynamic dose painting delivery technique is also pictured (D). Courtesy of Xcision Medical Systems, LLC.
Ipsilateral Breast Volume Receiving Mean Percentages (V) of Prescription Dose.a
|
| IMRT PBI PTV_10 mm | BSRTD PTV_10 mm | BSRTD PTV_3 mm |
|---|---|---|---|
|
| 72.1% (60.1%-81.7%) | 76.5% (61.6%-89.2%), | 64.1% (50.6%-75.9%), |
|
| 60.3% (43.8%-74.8%) | 42.0% (26.5%-51.3%), | 32.1% (20.6%-42.7%), |
|
| 42.5% (28.1%-55.0%) | 27.0% (16.1%-35.6%), | 17.7% (11.5%-27.1%), |
|
| 26.3% (15.4%-38.0%) | 16.8% (9.8%-22.7%), | 9.7% (6.1%-15.3%), |
|
| 11.1% (4.8%-21.2%) | 6.5% (3.1%-9.8%), | 3.3% (2.2%-5.3%), |
Abbreviations: BSRTD, breast stereotactic radiation therapy; IMRT, intensity-modulated radiation therapy; PBI, partial breast irradiation; PTV, planning target volume.
aIMRT PBI PTV_10mm (n = 7), BSRTD PTV_10mm (n = 7), BSRTD PTV_3mm (n = 9).
Figure 2.Ipsilateral breast volume receiving percentages (V) of prescription dose. Dark gray/lines: intensity-modulated radiation therapy accelerated partial breast irradiation (IMRT APBI); black/dots: breast stereotactic radiotherapy device (BSRTD) 10 mm; light gray/cross: BSRTD 3 mm.
Figure 3.Example isodose distributions for each planning technique. A, Intensity-modulated radiation therapy accelerated partial breast irradiation (IMRT APBI). B, Breast stereotactic radiotherapy device (BSRTD; planning target volume [PTV]_10mm). C, Breast stereotactic radiotherapy device (PTV_3mm). White contours, PTV_eval surrounding lumpectomy cavity; red line, 100%; dark blue line, 95%; yellow line, 80%; green line, 50%; light blue line, 20%; purple line, 5%.
Ipsilateral Organs at Risk (OAR) Receiving Reduced Mean Percentages (V x) of Prescription Dose.a
|
| IMRT PBI PTV_10mm | BSRTD PTV_10mm | BSRTD PTV_3mm |
|---|---|---|---|
|
| 29.0% (18.9%-35.3%) | 21.6% (12.6%-28.8%), | 15.9% (7.9%-25.7%), |
|
| 19.2% (1.0%-29.1%) | 13.4% (7.2%-18.4%), | 8.7% (1.7%-15.5%), |
|
| 12.5% (0%-23.0%) | 8.5% (3.9%-14.0%), | 4.6% (0.6%-11.1%), |
|
| 9.4% (0%-20.3%) | 6.1% (2.7%-11.1%), | 3.2% (0%-8.1%), |
|
| 54.2% (43.0%-68.4%) | 39.5% (25.5%-50.8%), | 26.9% (12.6%-43.1%), |
|
| 34.5% (20.8%-46.0%) | 22.0% (10.8%-36.0%), | 11.1% (1.0%-25.9%), |
|
| 11.8% (3.0%-39.1%) | 17.9% (0%-5.1%), | 0.8% (0%-1.8%), |
Abbreviations: BSRTD, breast stereotactic radiation therapy; IMRT, intensity-modulated radiation therapy; PBI, partial breast irradiation; PTV, planning target volume.
aIMRT PBI PTV_10mm (n = 7), BSRTD PTV_10mm (n = 7), BSRTD PTV_3mm (n = 9).