Literature DB >> 28687395

Multi-centre audit of VMAT planning and pre-treatment verification.

Diego Jurado-Bruggeman1, Victor Hernández2, Jordi Sáez3, David Navarro4, Francisco Pino5, Tatiana Martínez6, Maria-Elena Alayrach7, Norbert Ailleres8, Alejandro Melero9, Núria Jornet10.   

Abstract

BACKGROUND AND
PURPOSE: We performed a multi-centre intercomparison of VMAT dose planning and pre-treatment verification. The aims were to analyse the dose plans in terms of dosimetric quality and deliverability, and to validate whether in-house pre-treatment verification results agreed with those of an external audit.
MATERIALS AND METHODS: The nine participating centres encompassed different machines, equipment, and methodologies. Two mock cases (prostate and head and neck) were planned using one and two arcs. A plan quality index was defined to compare the plans and different complexity indices were calculated to check their deliverability. We compared gamma index pass rates using the centre's equipment and methodology to those of an external audit (global 3D gamma, absolute dose differences, 10% of maximum dose threshold). Log-file analysis was performed to look for delivery errors.
RESULTS: All centres fulfilled the dosimetric goals but plan quality and delivery complexity were heterogeneous and uncorrelated, depending on the manufacturer and the planner's methodology. Pre-treatment verifications results were within tolerance in all cases for gamma 3%-3mm evaluation. Nevertheless, differences between the external audit and in-house measurements arose due to different equipment or methodology, especially for 2%-2mm criteria with differences up to 20%. No correlation was found between complexity indices and verification results amongst centres.
CONCLUSIONS: All plans fulfilled dosimetric constraints, but plan quality and complexity did not correlate and were strongly dependent on the planner and the vendor. In-house measurements cannot completely replace external audits for credentialing.
Copyright © 2017 Elsevier B.V. All rights reserved.

Keywords:  Audit; Quality assurance; VMAT

Mesh:

Year:  2017        PMID: 28687395     DOI: 10.1016/j.radonc.2017.05.019

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  12 in total

Review 1.  Complexity metrics for IMRT and VMAT plans: a review of current literature and applications.

Authors:  Sophie Chiavassa; Igor Bessieres; Magali Edouard; Michel Mathot; Alexandra Moignier
Journal:  Br J Radiol       Date:  2019-07-24       Impact factor: 3.039

2.  Multi-institutional evaluation using the end-to-end test for implementation of dynamic techniques of radiation therapy in Thailand.

Authors:  Lalida Tuntipumiamorn; Puangpen Tangboonduangjit; Taweap Sanghangthum; Rattapol Rangseevijitprapa; Chirasak Khamfongkhruea; Thida Niyomthai; Boonrut Vuttiprasertpong; Supranee Supanant; Nongnuch Chatchaipaiboon; Porntip Iampongpaiboon; Pitchayut Nakkrasae; Tanwiwat Jaikuna
Journal:  Rep Pract Oncol Radiother       Date:  2018-11-28

3.  Impact of different optimization strategies on the compatibility between planned and delivered doses during radiation therapy of cervical cancer.

Authors:  Agata Jodda; Tomasz Piotrowski; Marta Kruszyna-Mochalska; Julian Malicki
Journal:  Rep Pract Oncol Radiother       Date:  2020-04-12

4.  Validation of the RapidArc Delivery System Using a Volumetric Phantom as Per Task Group Report 119 of the American Association of Physicists in Medicine.

Authors:  Lalit Kumar; Girigesh Yadav; Vimal Kishore; Manindra Bhushan; Munish Gairola; Deepak Tripathi
Journal:  J Med Phys       Date:  2019 Apr-Jun

Review 5.  Novel methodologies for dosimetry audits: Adapting to advanced radiotherapy techniques.

Authors:  Marlies Pasler; Victor Hernandez; Núria Jornet; Catharine H Clark
Journal:  Phys Imaging Radiat Oncol       Date:  2018-03-19

6.  Auditing local methods for quality assurance in radiotherapy using the same set of predefined treatment plans.

Authors:  Enrica Seravalli; Antonetta C Houweling; Leo Van Battum; Thom A Raaben; Marc Kuik; Jacco A de Pooter; Marion P R Van Gellekom; Jochem Kaas; Wilfred de Vries; Erik A Loeff; Jeroen B Van de Kamer
Journal:  Phys Imaging Radiat Oncol       Date:  2018-02-02

7.  The role of dosimetry audit in achieving high quality radiotherapy.

Authors:  Catharine H Clark; Núria Jornet; Ludvig P Muren
Journal:  Phys Imaging Radiat Oncol       Date:  2018-03-21

8.  Comparison of complexity metrics for multi-institutional evaluations of treatment plans in radiotherapy.

Authors:  Victor Hernandez; Jordi Saez; Marlies Pasler; Diego Jurado-Bruggeman; Nuria Jornet
Journal:  Phys Imaging Radiat Oncol       Date:  2018-02-22

9.  Comparison of patient-specific intensity modulated radiation therapy quality assurance for the prostate across multiple institutions.

Authors:  Kazuki Kubo; Hajime Monzen; Kohei Shimomura; Kenji Matsumoto; Tomoharu Sato; Mikoto Tamura; Kiyoshi Nakamatsu; Kentaro Ishii; Ryu Kawamorita
Journal:  Rep Pract Oncol Radiother       Date:  2019-10-17

10.  Reducing variability among treatment machines using knowledge-based planning for head and neck, pancreatic, and rectal cancer.

Authors:  Hideaki Hirashima; Mitsuhiro Nakamura; Nobutaka Mukumoto; Ryo Ashida; Kota Fujii; Kiyonao Nakamura; Aya Nakajima; Katsuyuki Sakanaka; Michio Yoshimura; Takashi Mizowaki
Journal:  J Appl Clin Med Phys       Date:  2021-06-20       Impact factor: 2.102

View more