| Literature DB >> 28686629 |
Linchuan Wang1, Kai-Hua Zhou2, He-Ping Zhao3, Ji-Han Wang3, Hai-Chao Zheng4, Yan Yu3, Wei Chen1.
Abstract
OBJECTIVES: Individuals with recent or acute HIV infection are more infectious than those with established infection. Our objective was to analyze the characteristics of detection among HIV infections in Xi'an.Entities:
Mesh:
Substances:
Year: 2017 PMID: 28686629 PMCID: PMC5501483 DOI: 10.1371/journal.pone.0180071
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Fig 1The comparison of four kits performance and the frequency of different band detection.
A, The comparison of false-positive rate and positive predictive value (PPV) between a 4th generation assay(Architect HIV Ag/Ab Combo) and three 3rdgeneration EIAs (WanTai, XinChuang and Livzon).Note: PPV was positive predictive value, FPR was false-positive rate.B, The frequency of different band was detected in the patients with initial WB positive results.
The results of Western blot and follow-up.
| screening positive | Western blot | ||
|---|---|---|---|
| Positive, n(%) | Negative, n(%) | Indeterminate, n(%) | |
| Abbott (n = 309) | 240 (77.67%) | 54 (17.48%) | 15 (4.85%)* |
| WanTai (n = 162) | 145 (89.51%) | 9 (5.56%) | 8 (4.94%)* |
| XinChuang (n = 138) | 123 (89.13%) | 13 (9.42%) | 2 (1.45%) |
| Livzon (n = 56) | 51 (91.07%) | 3 (5.36%) | 2 (3.57%) |
Frequency of 11 band patterns detected in 559 patients with initial WB positive results.
| WB band profile | Frequency (n = 559) | ||||
|---|---|---|---|---|---|
| Abbott | WanTai | XinChuang | Livzon | ||
| p24 p17 p55 p39 p31 p51 p66 gp41 gp120/gp160 | 125 | 80 | 60 | 32 | |
| p24 p17 p55 p31 p51 p66 gp41 gp120/gp160 | 67 | 50 | 39 | 12 | |
| p24 p17 p31 p51 p66 gp41 gp120/gp160 | 29 | 10 | 13 | 3 | |
| p24 p31 p51 p66 gp41 gp120/gp160 | 4 | 0 | 1 | 0 | |
| p24 p17 p51 p66 gp41 gp120/gp160 | 1 | 0 | 0 | 2 | |
| p24 p51 p66 gp41 gp120/gp160 | 2 | 2 | 2 | 0 | |
| p24 p31 p51 gp41 gp120/gp160 | 0 | 0 | 0 | 1 | |
| p24 p66 gp41 gp120/gp160 | 3 | 1 | 1 | 0 | |
| p24 p51 gp41 gp120/gp160 | 0 | 1 | 0 | 0 | |
| p24 gp41 gp120/gp160 | 1 | 1 | 0 | 0 | |
| p24 gp120/gp160 | 8 | 0 | 7 | 1 | |
The consistency and PPV of four kits at different screening values.
| Screening positive | Retesting positive | Consistency | Western blot confirmed | ||||
|---|---|---|---|---|---|---|---|
| Used kit | Value | Number | Number | FP(n) | TP(n) | PPV(%) | |
| Architect HIV Ag/Ab | |||||||
| 1.00–5.00 | 44 | 3 | 6.82 | 44 | 0 | 0 | |
| 5.00–30.00 | 39 | 25 | 64.1 | 24 | 15 | 38.46 | |
| ≥30.00 | 226 | 226 | 100 | 0 | 226 | 100 | |
| WanTai | |||||||
| 1.00–3.00 | 10 | 3 | 23.08 | 10 | 0 | 0 | |
| 3.00–10.00 | 12 | 8 | 40 | 6 | 6 | 50 | |
| ≥10.00 | 140 | 140 | 100 | 0 | 140 | 100 | |
| XinChuang | |||||||
| 1.00–3.00 | 8 | 2 | 20 | 8 | 0 | 0 | |
| 3.00–10.00 | 17 | 15 | 46.88 | 7 | 10 | 58.82 | |
| ≥10.00 | 113 | 113 | 100 | 0 | 113 | 100 | |
| Livzon | |||||||
| 1.00–3.00 | 4 | 1 | 20 | 4 | 0 | 0 | |
| 3.00–10.00 | 20 | 12 | 37.5 | 1 | 19 | 95 | |
| ≥10.00 | 32 | 32 | 100 | 0 | 32 | 100 | |
Note: FP was false positive; TP was true positive.
Fig 2Distribution of screening values using different kits in the first WB negative, indeterminate and positive result.