Abstract
BACKGROUND: Current pressurised metered dose asthma inhaler (pMDI) propellants are not inert pharmacologically as were previous chlorofluorocarbons, have smooth muscle relaxant' partial pressure effects in the lungs and inhaled hydrofluoroalkane 134a (norflurane) has anaesthetic effects. Volumes of propellant gas per actuation have never been measured.Entities:
Keywords: Metered dose inhalers; Propellants; Toxicology; Valved holding chambers (spacers); Ventilator inhaler devices
Year: 2017 PMID: 28670327 PMCID: PMC5492461 DOI: 10.1186/s13223-017-0202-0
Source DB: PubMed Journal: Allergy Asthma Clin Immunol ISSN: 1710-1484 Impact factor: 3.406
Fig. 1Full and evacuated 500 ml anaesthetic reservoir bags. The evacuated contains an inhaler and other inhalers and cartridges are shown
Propellant volumes from different pMDIs
| HFA 134a norflurane | Salamol, (IVAX) small | 490/70 puffs = 7 ml per puff |
| Ventolin (A&H) large | 480/30 puffs = 16 ml per puff | |
| Clenil Modulite™ (Chiesi) large | 400/35 puffs = 11.75 ml per puff | |
| HFA 227ea apaflurane | Intal™ (Sanofi-Aventis) large | 465/35 puffs = 13.5 ml per puff |
| Vannair™ (AstraZeneca) small | 480/53 puffs = 9 ml per puff | |
| CFCa | Serevent™ (A&H) small | 530 ml/35 puffs = 15.6 ml per puff |
| Ventolin ™ (A&H) large | 490/24 puffs = 20.4 ml per puff |
aCFCs dichlorodifluoromethane Cl2F2C; trichlorofluoromethane Cl3FC plus lecithin
Weights and volumes of salbutamol HFA134a propellant pMDIs
| Small salbutamol SalamolR | Large salbutamol VentolinTM | |
|---|---|---|
| Actuations of a first full pMDI | ||
| Cartridge weight | ||
| Full | 15.64 g | 28.75 g |
| Empty | 7.76 g | 11.43 g |
| Total volume | 1640 ml | 3885 mla |
| Actuations of a second full pMDI | ||
| 151–200 (50) | 350 ml | 800 ml |
| 201–210 (10) | 70 ml | 160 ml |
| 211–220 | 70 ml (last “sensing”) | 160 ml (last “sensing”) |
| 221–230 | 65 ml | 100 ml |
| 231–240 | 30 ml | 70 ml |
| 241–250 | 5 ml | 40 ml |
| 251–260 | 0 ml | 15 ml |
| Total volume | 1640 ml | 3745 mla |
a140 ml less total propellant gas volume of second large pMDI
Valved holding chamber type; their volumes; and drop in oxygen percentage from 21% air per actuation
| Spacer type | Volume (ml) | 1 puff (O2 %) | 2 puffs (O2 %) | 10 puffs (O2 %) |
|---|---|---|---|---|
| Ablespacer™ | 126 | 19 | 17 | 11 |
| Vortex™ | 180 | 19 | 18 | 13 |
| BabyHaler™ | 393 | 20 | 19 | 16 |
| Volumatic™ | 788 | 20 | 19 | 17 |
Falls in percentage from 100% oxygen after five actuations of pMDIs into four ventilator delivery devices
| 5× Duolin™ | 5× Respigen™ | 5× Vannair™ | 5× Intal™ | |
|---|---|---|---|---|
| L-Trace™ | 88 | 87 | 89 | 81 |
| Spirale™ | 86 | 87 | 89 | 82 |
| Intersurgical™ | 87 | 86 | 87 | 79 |
| RT 200 (ventilator tubing) | 90 | 91 | 93 | 91 |