Medical aerosol propellant interference with infrared anaesthetic gas monitors.

P D Levin1, D Levin, A Avidan.   

Abstract

BACKGROUND: 1,1,1,2 Tetrafluoroethane is a hydrofluoroalkane (HFA) that is replacing chlorofluorocarbons (CFC) as a medical aerosol propellant in an attempt to reduce damage to the ozone layer. This study compared the effects of HFA- and CFC-based inhalers on four anaesthetic gas monitoring systems.
METHODS: The HFA- and CFC-based inhalers were activated in close proximity to the sample line of two Datex Ohmeda, an Agilent and a Siemens infrared anaesthetic agent monitoring systems. The effects were recorded on each system for five common anaesthetic agents.
RESULTS: The HFA inhaler caused either maximal false positive readings (with the exception of desflurane) or transient measurement failure on all systems. The Datex Ohmeda AS/3 system misidentified the HFA inhaler as carbon dioxide at low concentration (2 +/- 0 mm Hg). The CFC-based inhaler caused a minor false-positive reading (0.4 +/- 0%) for halothane only on the Datex Ohmeda AS/3 system only and was misidentified as carbon dioxide at 33.3 (sd 2.1) mm x Hg and 22.4 (8.9) mm x Hg by the Agilent and Siemens systems.
CONCLUSIONS: The HFA inhaler adversely affected all equipment tested. The infrared spectra of HFA and the common anaesthetic gases have considerable overlap at the 8-12 microm range that is not shared by the CFCs. The differences in spectral overlap explain the different effects of the HFA and CFC propellants. Anaesthetic gas concentration data may be erroneous using the HFA-based inhalers.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15121726     DOI: 10.1093/bja/aeh154

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  7 in total

1.  Interference of anesthesia monitoring equipment by HFA propellants. Letter to the editor.

Authors:  Phillip D Levin; Alexander Avidan
Journal:  J Clin Monit Comput       Date:  2010-06-09       Impact factor: 2.502

2.  Anaesthetic in the garb of a propellant.

Authors:  Shagun Bhatia Shah; Uma Hariharan; Ajay Kumar Bhargava
Journal:  Indian J Anaesth       Date:  2015-04

3.  Halothane: I am still there.

Authors:  Pooja Bihani; Deepak Choudhary; Pradeep Kumar Bhatia; Sadik Mohammed
Journal:  Indian J Anaesth       Date:  2016-12

4.  Spectral tissue sensing to identify intra- and extravascular needle placement - A randomized single-blind controlled trial.

Authors:  Andrea J R Balthasar; Geert-Jan van Geffen; Marjolein van der Voort; Gerald W Lucassen; Stefan Roggeveen; Ivar J Bruaset; Joergen Bruhn
Journal:  PLoS One       Date:  2017-03-09       Impact factor: 3.240

Review 5.  Analysis of oxygen, anaesthesia agent and flows in anaesthesia machine.

Authors:  Rakesh Garg; Ramesh Chand Gupta
Journal:  Indian J Anaesth       Date:  2013-09

6.  Asthma pressurised metered dose inhaler performance: propellant effect studies in delivery systems.

Authors:  William F S Sellers
Journal:  Allergy Asthma Clin Immunol       Date:  2017-06-29       Impact factor: 3.406

7.  Gas analyzer aberrancy: Due to disinfectant?

Authors:  Barkha Bindu; Hemanshu Prabhakar; Siddharth Chavali
Journal:  J Anaesthesiol Clin Pharmacol       Date:  2018 Jul-Sep
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.