Three tests performed using the cobas 4800 system on a subset of cervical specimens containing large quantities of mucus
| Test | Cobas x480 | Cobas z480 | Percentage of resolved specimens | ||||
|---|---|---|---|---|---|---|---|
| N of failed results | N of passed results | N of negative results | N of positive results | N of invalid results | |||
| X3 | X4 | ||||||
| Test 1 (without treatment) | 13 | 6 | 1 | 1 | 0 | 19 | 5% |
| Test 2 (with vortexing before the test) | 5 | 1 | 14 | 5 | 9 | 6 | 70% |
| Test 3 (pretreatment with Sputasol and vortexing before the test) | 0 | 0 | 20 | 9 | 11 | 0 | 100% |
X3(Error): clot was detected and specimen was not processed; X4 (Error): pipetting error occurred and specimen was not processed.