Kristina Gorseta1, Domagoj Glavina1, Tomislav Skrinjaric1, Beata Czarnecka2, John W Nicholson3.
Abstract
This study determined the influence of coating with either petroleum jelly or light-cured varnish and storage medium on the flexural strength of glass-ionomer cements (GIC). The flexural strength of two glass-ionomer cements (Fuji Equia Fil and Ketac Molar Aplicap) was measured. Specimens (2 × 2 × 25 mm) were prepared in three groups: uncoated, coated with petroleum jelly, or coated with light-cured varnish (EquiaCoat) cured for 20 s using a cure lamp (Bluephase G2, Ivoclar Vivadent, Schaan, Liechtenstein). Specimens were stored for 1 week at 37 °C in water, artificial saliva or 20 mmol dm-3 lactic acid, then flexural strength was determined in 3-pont bend. Data were analyzed by ANOVA and Tukey HSD post hoc test (p < 0.05). In addition, the mold was filled with water and the temperature change caused by the cure lamp was measured with a thermocouple. For both materials, storage in water gave the lowest flexural strength. It was slightly higher in either saliva or lactic acid, and was improved by coating in petroleum jelly. Specimens coated with light-cured varnish, that also involved heating with a cure lamp, gave the highest flexural strength. The heating effect of the lamp was demonstrated by the temperature rise in the water in the mold after light exposure from 21.9 (± 1.0) °C to 26.8 (± 1.0) °C. hence, sealing of GIC from aqueous media improves flexural strength. The cure lamp emitted heat, which may enhance the flexural strength of specimens coated with light-cured varnish.Entities:
Keywords: Coating; flexural strength; glass-ionomer
Year: 2016 PMID: 28642913 PMCID: PMC5433215 DOI: 10.3109/23337931.2016.1160784
Source DB: PubMed Journal: Acta Biomater Odontol Scand ISSN: 2333-7931
Figure 1.Test assembly for three-point bending test.
Mean flexural strength values for Fuji Equia with various coating treatments, stored in different media (standard deviations in parentheses).
| Storage medium | Uncoated | Coated with petroleum jelly | Coated with EquiaCoat + 20 s cure |
|---|---|---|---|
| Water | 21.8 (13.9)a | 36.9 (7.8)b | 69.7 (13.4)c |
| Artificial saliva | 37.4 (22.3)b | 33.7 (6.4)b | 53.5 (27.2)c |
| Lactic acid | 36.5 (10.9)b | 37.4 (14.8)b | 66.3 (15.8)c |
a, b, c show groups which do not differ significantly from each other. Means for groups a, b and c differ significantly (p < 0.05).
Mean flexural strength values for Ketac Molar Aplicap with various coating treatments, stored in different media (standard deviations in parentheses).
| Storage medium | Uncoated | Coated with petroleum jelly | Coated with EquiaCoat + 20 s cure |
|---|---|---|---|
| Water | 10.0 (3.3)a | 17.3 (4.7)c | 59.3 (19.8)d |
| Artificial saliva | 36.6 (15.4)b | 14.8 (5.0)a,c | 62.7 (16.6)d |
| Lactic acid | 31.4 (12.1)b | 35.4 (10.2)b | 60.0 (22.4)d |
a, b, c and d show groups which do not differ significantly from each other. Means for groups a, b, c and d differ significantly (p < 0.05).
Mean temperatures of water samples in the cylinder before and after application of the cure lamp for 20 s (standard deviation in parentheses).
| Time/s | Temperature (°C) |
|---|---|
| 0 | 21.9 (0.9) |
| 20 | 26.8 (1.0) |