Literature DB >> 28626862

Acoustic field characterization of a clinical magnetic resonance-guided high-intensity focused ultrasound system inside the magnet bore.

Satya V V N Kothapalli1, Michael B Altman2, Ari Partanen3, Leighton Wan1, H Michael Gach4, William Straube2, Dennis E Hallahan2, Hong Chen5.   

Abstract

PURPOSE: With the expanding clinical application of magnetic resonance-guided high-intensity focused ultrasound (MR-HIFU), acoustic field characterization of MR-HIFU systems is needed for facilitating regulatory approval and ensuring consistent and safe power output of HIFU transducers. However, the established acoustic field measurement techniques typically use equipment that cannot be used in a magnetic resonance imaging (MRI) suite, thus posing a challenge to the development and execution of HIFU acoustic field characterization techniques. In this study, we developed and characterized a technique for HIFU acoustic field calibration within the MRI magnet bore, and validated the technique with standard hydrophone measurements outside of the MRI suite.
METHODS: A clinical Philips MR-HIFU system (Sonalleve V2, Philips, Vantaa, Finland) was used to assess the proposed technique. A fiber-optic hydrophone with a long fiber was inserted through a 24-gauge angiocatheter and fixed inside a water tank that was placed on the HIFU patient table above the acoustic window. The long fiber allowed the hydrophone control unit to be placed outside of the magnet room. The location of the fiber tip was traced on MR images, and the HIFU focal point was positioned at the fiber tip using the MR-HIFU therapy planning software. To perform acoustic field mapping inside the magnet, the HIFU focus was positioned relative to the fiber tip using an MRI-compatible 5-axis robotic transducer positioning system embedded in the HIFU patient table. To perform validation measurements of the acoustic fields, the HIFU table was moved out of the MRI suite, and a standard laboratory hydrophone measurement setup was used to perform acoustic field measurements outside the magnetic field.
RESULTS: The pressure field scans along and across the acoustic beam path obtained inside the MRI bore were in good agreement with those obtained outside of the MRI suite. At the HIFU focus with varying nominal acoustic powers of 10-500 W, the peak positive pressure and peak negative pressure measured inside the magnet bore were 3.87-68.67 MPa and 3.56-12.06 MPa, respectively, while outside the MRI suite the corresponding pressures were 3.27-67.32 MPa and 3.06-12.39 MPa, respectively. There was no statistically significant difference (P > 0.05) between measurements inside the magnet bore and outside the MRI suite for the p+ and p- at any acoustic power level. The spatial-peak pulse-average intensities (ISPPA ) for these powers were 312-17816 W/cm2 and 220-15698 W/cm2 for measurements inside and outside the magnet room, respectively. In addition, when the scanning step size of the HIFU focus was increased from 100 μm to 500 μm, the execution time for scanning a 4 × 4 mm2 area decreased from 210 min to 10 min, the peak positive pressure decreased by 14%, the peak negative pressure decreased by 5%, and the lateral full width at half maximum dimension of pressure profiles increased from 1.15 mm to 1.55 mm.
CONCLUSIONS: The proposed hydrophone measurement technique offers a convenient and reliable method for characterizing the acoustic fields of clinical MR-HIFU systems inside the magnet bore. The technique was validated for use by measurements outside the MRI suite using a standard hydrophone calibration technique. This technique can be a useful tool in MR-HIFU quality assurance and acoustic field assessment.
© 2017 American Association of Physicists in Medicine.

Entities:  

Keywords:  MR-HIFU; MR-guided HIFU; acoustic field mapping; fiber-optic hydrophone; quality assurance

Mesh:

Year:  2017        PMID: 28626862     DOI: 10.1002/mp.12412

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  6 in total

1.  Spatiotemporally-controlled transgene expression in hydroxyapatite-fibrin composite scaffolds using high intensity focused ultrasound.

Authors:  Alexander Moncion; Jonah S Harmon; Yan Li; Sam Natla; Easton C Farrell; Oliver D Kripfgans; Jan P Stegemann; Francisco M Martín-Saavedra; Nuria Vilaboa; Renny T Franceschi; Mario L Fabiilli
Journal:  Biomaterials       Date:  2018-12-13       Impact factor: 12.479

2.  Correction for Spatial Averaging Artifacts in Hydrophone Measurements of High-Intensity Therapeutic Ultrasound: An Inverse Filter Approach.

Authors:  Keith A Wear; Samuel M Howard
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2019-06-24       Impact factor: 2.725

3.  Focused Ultrasound-enabled Brain Tumor Liquid Biopsy.

Authors:  Lifei Zhu; Galen Cheng; Dezhuang Ye; Arash Nazeri; Yimei Yue; Weijun Liu; Xiaowei Wang; Gavin P Dunn; Allegra A Petti; Eric C Leuthardt; Hong Chen
Journal:  Sci Rep       Date:  2018-04-26       Impact factor: 4.379

4.  A convenient, reliable, and fast acoustic pressure field measurement method for magnetic resonance-guided high-intensity focused ultrasound systems with phased array transducers.

Authors:  Satya V V N Kothapalli; Ari Partanen; Lifei Zhu; Michael B Altman; H Michael Gach; Dennis E Hallahan; Hong Chen
Journal:  J Ther Ultrasound       Date:  2018-07-02

5.  Focused ultrasound for safe and effective release of brain tumor biomarkers into the peripheral circulation.

Authors:  Lifei Zhu; Arash Nazeri; Christopher Pham Pacia; Yimei Yue; Hong Chen
Journal:  PLoS One       Date:  2020-06-03       Impact factor: 3.240

6.  Static Magnetic Fields Dampen Focused Ultrasound-mediated Blood-Brain Barrier Opening.

Authors:  Yaoheng Yang; Christopher Pham Pacia; Dezhuang Ye; Yimei Yue; Chih-Yen Chien; Hong Chen
Journal:  Radiology       Date:  2021-07-06       Impact factor: 29.146

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.