Abstract
Aim: To evaluate the plan quality of 6MV unflattened (UFB) and flattened beam (FB) photon energy using AAA dose calculation algorithms for volumetric arc therapy. Materials andEntities:
Keywords: Flattened beam; unflattened beam; rapid arc and volumetric arc therapy
Year: 2017 PMID: 28612590 PMCID: PMC5555550 DOI: 10.22034/APJCP.2017.18.5.1377
Source DB: PubMed Journal: Asian Pac J Cancer Prev ISSN: 1513-7368
PTV Coverage and OAR’s Doses for 6MV FB and UFB. SD, Standard Deviation
| Target and OARS | Parameters | 6MV_FB RA | 6MV_UFB_RA | 6MV RA FB Vs UFB |
|---|---|---|---|---|
| Mean ± SD | Mean ± SD | (p value) | ||
| D98% (Gy) | 49.45 ± 0.2 | 49.31 ± 0.13 | NS | |
| D95% (Gy) | 50.34 ± 0.1 | 50.26 ±0. 74 | NS | |
| D50% (Gy) | 53.00 ± 0.3 | 53.20 ± 0.3 | p<0.05 | |
| D2% (Gy) | 55.31 ± 0.3 | 55.74 ± 1.5 | p<0.05 | |
| PTV | HI | 0.11 ± 0.01 | 0.12 ± 0.00 | p<0.05 |
| CI | 1.12 ± 0.04 | 1.16 ± 0.05 | p<0.05 | |
| V110% | 1.8 ± 1.4 | 3.4 ± 1.04 | NS | |
| V20Gy(%) | 26.6 ± 4.5 | 26.9 ± 3.9 | p<0.05 | |
| V45Gy(%) | 3.1 ± 1.5 | 3.7 ± 1.5 | p<0.05 | |
| Common Lung | V5Gy(%) | 84.0 ± 14.7 | 84.2 ± 14.9 | p<0.05 |
| Mean dose (Gy) | 15.47 ± 1.3 | 15.82 ± 1.4 | p<0.05 | |
| V10Gy(%) | 34.3 ± 6.7 | 37.8 ± 9.6 | p<0.05 | |
| V25Gy(%) | 9.5 ± 4.2 | 10.6 ± 4.4 | p<0.05 | |
| Heart | V35Gy(%) | 3.0± 2.1 | 3.6 ± 2.3 | p<0.05 |
| Mean dose (Gy) | 10.17 ± 2.6 | 10.84± 3.1 | p<0.05 | |
| NTID (105Gy cm3) | 142.44 ± 45 | 141.44 ± 45 | p<0.05 | |
| V1Gy (%) | 62.33 ± 11.05 | 61.50 ± 10.8 | p<0.05 | |
| V2Gy (%) | 51.16 ± 9.20 | 50.41 ± 9.03 | p<0.05 | |
| V3Gy (%) | 47.19 ± 8.33 | 46.76 ± 8.21 | p<0.05 | |
| Low Dose Volume | V4Gy (%) | 44.81 ± 7.89 | 43.45 ± 7.74 | p<0.05 |
| V5Gy (%) | 42.37 ± 7.56 | 41.89 ± 8.13 | p<0.05 | |
| V10Gy (%) | 28.55 ± 6.19 | 27.12 ± 5.89 | p<0.05 | |
| MU | Monitor Unit | 1214 ± 66 | 1638.4 ± 78 | p<0.05 |
| BOT | minutes | 3.01 ± 0.13 | 2.50 ± 0.12 | p<0.05 |