Modified ideal observer model (MIOM) for high-contrast and high-spatial resolution CT imaging tasks.

Juan P Cruz-Bastida1, Daniel Gomez-Cardona1, John Garrett1, Timothy Szczykutowicz1,2,3, Guang-Hong Chen1,2, Ke Li1,2.   

Abstract

PURPOSE: Although a variety of mathematical observer models have been developed to predict human observer performance for low contrast lesion detection tasks, their predictive power for high contrast and high spatial resolution discrimination imaging tasks, including those in CT bone imaging, could be limited. The purpose of this work was to develop a modified observer model that has improved correlation with human observer performance for these tasks.
METHODS: The proposed observer model, referred to as the modified ideal observer model (MIOM), uses a weight function to penalize components in the task function that have less contribution to the actual human observer performance for high contrast and high spatial resolution discrimination tasks. To validate MIOM, both human observer and observer model studies were performed, each using exactly the same CT imaging task [discrimination of a connected component in a high contrast (1000 HU) high spatial resolution bone fracture model (0.3 mm)] and experimental CT image data. For the human observer studies, three physicist observers rated the connectivity of the fracture model using a five-point Likert scale; for the observer model studies, a total of five observer models, including both conventional models and the proposed MIOM, were used to calculate the discrimination capability of the CT images in resolving the connected component. Images used in the studies encompassed nine different reconstruction kernels. Correlation between human and observer model performance for these kernels were quantified using the Spearman rank correlation coefficient (ρ). After the validation study, an example application of MIOM was presented, in which the observer model was used to select the optimal reconstruction kernel for a High-Resolution (Hi-Res, GE Healthcare) CT scan technique.
RESULTS: The performance of the proposed MIOM correlated well with that of the human observers with a Spearman rank correlation coefficient ρ of 0.88 (P = 0.003). In comparison, the value of ρ was 0.05 (P = 0.904) for the ideal observer, 0.05 (P = 0.904) for the non-prewhitening observer, -0.18 (P = 0.634) for the non-prewhitening observer with eye filter and internal noise, and 0.30 (P = 0.427) for the prewhitening observer with eye filter and internal noise. Using the validated MIOM, the optimal reconstruction kernel for the Hi-Res mode to perform high spatial resolution and high contrast discrimination imaging tasks was determined to be the HD Ultra kernel at the center of the scan field of view (SFOV), or the Lung kernel at the peripheral region of the SFOV. This result was consistent with visual observations of nasal CT images of an in vivo canine subject.
CONCLUSION: Compared with other observer models, the proposed modified ideal observer model provides significantly improved correlation with human observers for high contrast and high spatial resolution CT imaging tasks.
© 2017 American Association of Physicists in Medicine.

Entities:  

Keywords:  CT; Hi-Res; ROC; high-spatial resolution; human observer; image quality; observer model; reconstruction kernel

Mesh:

Year:  2017        PMID: 28600849      PMCID: PMC5603220          DOI: 10.1002/mp.12404

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  33 in total

1.  Receiver operating characteristic rating analysis. Generalization to the population of readers and patients with the jackknife method.

Authors:  D D Dorfman; K S Berbaum; C E Metz
Journal:  Invest Radiol       Date:  1992-09       Impact factor: 6.016

2.  Task-based assessment of breast tomosynthesis: effect of acquisition parameters and quantum noise.

Authors:  I Reiser; R M Nishikawa
Journal:  Med Phys       Date:  2010-04       Impact factor: 4.071

3.  A comparison of the Dorfman-Berbaum-Metz and Obuchowski-Rockette methods for receiver operating characteristic (ROC) data.

Authors:  Stephen L Hillis; Nancy A Obuchowski; Kevin M Schartz; Kevin S Berbaum
Journal:  Stat Med       Date:  2005-05-30       Impact factor: 2.373

4.  Optimization of dual-energy imaging systems using generalized NEQ and imaging task.

Authors:  S Richard; J H Siewerdsen
Journal:  Med Phys       Date:  2007-01       Impact factor: 4.071

5.  Amorphous selenium flat panel detectors for digital mammography: validation of a NPWE model observer with CDMAM observer performance experiments.

Authors:  Jennifer A Segui; Wei Zhao
Journal:  Med Phys       Date:  2006-10       Impact factor: 4.071

6.  A comparison of denominator degrees of freedom methods for multiple observer ROC analysis.

Authors:  Stephen L Hillis
Journal:  Stat Med       Date:  2007-02-10       Impact factor: 2.373

7.  Recent developments in the Dorfman-Berbaum-Metz procedure for multireader ROC study analysis.

Authors:  Stephen L Hillis; Kevin S Berbaum; Charles E Metz
Journal:  Acad Radiol       Date:  2008-05       Impact factor: 3.173

8.  Image covariance and lesion detectability in direct fan-beam x-ray computed tomography.

Authors:  Adam Wunderlich; Frédéric Noo
Journal:  Phys Med Biol       Date:  2008-04-18       Impact factor: 3.609

9.  Human observer detection experiments with mammograms and power-law noise.

Authors:  A E Burgess; F L Jacobson; P F Judy
Journal:  Med Phys       Date:  2001-04       Impact factor: 4.071

10.  Visual signal detection with two-component noise: low-pass spectrum effects.

Authors:  A E Burgess
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  1999-03       Impact factor: 2.129

View more
  2 in total

1.  Low-dose cone-beam CT via raw counts domain low-signal correction schemes: Performance assessment and task-based parameter optimization (Part II. Task-based parameter optimization).

Authors:  Daniel Gomez-Cardona; John W Hayes; Ran Zhang; Ke Li; Juan Pablo Cruz-Bastida; Guang-Hong Chen
Journal:  Med Phys       Date:  2018-04-06       Impact factor: 4.071

2.  Diagnostic Accuracy and Failure Mode Analysis of a Deep Learning Algorithm for the Detection of Intracranial Hemorrhage.

Authors:  Andrew F Voter; Ece Meram; John W Garrett; John-Paul J Yu
Journal:  J Am Coll Radiol       Date:  2021-04-03       Impact factor: 6.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.