Literature DB >> 28583022

Implementation of the new VIRTUO blood culture system: evaluation and comparison to the 3D system using simulated blood cultures.

Nathalie Miller1, Laetitia Brassinne2, Daniel Allemeersch2.   

Abstract

PURPOSE: To evaluate the performances of the newly approved BacT/ALERT VIRTUO blood culture system for the recovery of bloodstream pathogens and compare it to the BacT/ALERT 3D system.
METHODS: Simulated blood cultures of eight clinically relevant microorganisms were used: Bacteroides fragilis (ATCC 25285), Escherichia coli (ATCC 25922), Haemophilus influenzae (ATCC 49247), Pseudomonas aeruginosa (ATCC 27853), Enterococcus faecalis (ATCC 29212), Staphylococcus aureus (ATCC 29213), Streptococcus pneumoniae (ATCC 49619) and Candida krusei (ATCC 6258). Criteria for comparison were culture positivity and time to detection (TTD). The effects of delayed entry on recovery and TTD were also evaluated.
RESULTS: The VIRTUO exhibited around 3 h faster detection time compared to the 3D system. (p < 0.01) for aerobic and facultative microorganisms. The difference in TTD was greatest for the B. fragilis, with a median difference of 46.67 h. The anaerobic bottle of the VIRTUO (FN Plus) did not support the growth of obligate aerobes, whereas the 3D did so. Delayed entry (studied with an E. Coli isolate) had no effect on the recovery rate but proportionally reduced TTD.
CONCLUSIONS: The VIRTUO performed better than the 3D in terms of TTD and hands-on-time. FN Plus vial appears to be more efficient than the SN bottle in the recovery of anaerobes.

Entities:  

Keywords:  BacT/ALERT VIRTUO septicaemia; Blood cultures; anaerobes; delayed entry; time to detection

Mesh:

Year:  2017        PMID: 28583022     DOI: 10.1080/17843286.2017.1331618

Source DB:  PubMed          Journal:  Acta Clin Belg        ISSN: 1784-3286            Impact factor:   1.264


  5 in total

1.  Improved blood culture workflow in the time to detection of microorganisms placing incubators systems outside of microbiology laboratory.

Authors:  M Angeles Orellana; Fernando Chaves; Rafael Delgado
Journal:  Braz J Microbiol       Date:  2020-05-18       Impact factor: 2.476

2.  Evaluation of Optimal Blood Culture Incubation Time To Maximize Clinically Relevant Results from a Contemporary Blood Culture Instrument and Media System.

Authors:  Eric M Ransom; Zahra Alipour; Meghan A Wallace; C A Burnham
Journal:  J Clin Microbiol       Date:  2021-02-18       Impact factor: 5.948

3.  Time-to-detection of bacteria and yeast with the BACTEC FX versus BacT/Alert Virtuo blood culture systems.

Authors:  Ali Mohammed Somily; Hanan Ahmed Habib; Armen Albert Torchyan; Samina B Sayyed; Muhammed Absar; Rima Al-Aqeel; A Khalifa Binkhamis
Journal:  Ann Saudi Med       Date:  2018 May-Jun       Impact factor: 1.526

4.  Effect of delayed entry on performance of the BACT/ALERT FAN PLUS bottles in the BACT/ALERT VIRTUO blood culture system.

Authors:  Mary Adamik; Anne Hutchins; Jasmin Mangilit; Betsy Katzin; Heather Totty; Parampal Deol
Journal:  Eur J Clin Microbiol Infect Dis       Date:  2020-10-09       Impact factor: 3.267

5.  Considerations in evaluating equipment-free blood culture bottles: A short protocol for use in low-resource settings.

Authors:  Sien Ombelet; Alessandra Natale; Jean-Baptiste Ronat; Olivier Vandenberg; Jan Jacobs; Liselotte Hardy
Journal:  PLoS One       Date:  2022-04-25       Impact factor: 3.752