Optimal injection method for long-range computed tomography angiography.

Aogu Yamaguchi1, Tsukasa Sasaki2.   

Abstract

We aimed to obtain high-stable computed tomography (CT) values for CT angiography (CTA). We verified the optimal use of the new protocol for long-range scanning using a contrast material flow phantom (Ichikawa, CT imaging theory. Igaku syoin Publishing, Tokyo, 8; Awai et al., Am J Roentgenol 186(2):379-385, 1) to obtain high-stable CT values (Fig. 2). We have developed a novel contrast injection method called the stable line imaging protocol (SLIP). This method involved a gradual reduction in the rate of administration of the contrast medium, which was accompanied by simultaneous administration of saline to compensate for the resultant decrease in the rate of administration. A saline flush was further added after the contrast medium injection. The rate of administration of the saline flush was the same as that of the initial contrast medium. We injected a contrast medium from the injector into the flow phantom. The time-enhancement curve (TEC) data were obtained five times by changing the injection method, and the average CT values were determined. The maximum simulation of the aortic peak enhancement of TEC was defined as pTEC. The time required to keep the CT value more than 80% of the pTEC was defined as the time period of contrast enhancement (p80). The pTEC of a single injection was 295.7 ± 5.3 HU, while that of SLIP was 296.5 ± 2.3 HU; the two values did not differ significantly. The p80 of a single injection was 19.1 ± 0.2 s, while that of SLIP was 23.2 ± 0.4 s; p80 of SLIP was 18% longer than that of the single injection.

Entities:  

Keywords:  Computed tomography (CT); Computed tomography angiography (CTA); Contrast medium injection method; Stable line imaging protocol (SLIP); Time-enhancement curve (TEC)

Mesh:

Year:  2017        PMID: 28573550     DOI: 10.1007/s12194-017-0402-9

Source DB:  PubMed          Journal:  Radiol Phys Technol        ISSN: 1865-0333


  12 in total

1.  Cerebral CT angiography using a small volume of concentrated contrast material with a test injection method: optimal scan delay for quantitative and qualitative performance.

Authors:  N Takeyama; K Kuroki; T Hayashi; S Sai; N Okabe; Y Kinebuchi; T Hashimoto; T Gokan
Journal:  Br J Radiol       Date:  2012-03-14       Impact factor: 3.039

Review 2.  Intravenous contrast medium administration and scan timing at CT: considerations and approaches.

Authors:  Kyongtae T Bae
Journal:  Radiology       Date:  2010-07       Impact factor: 11.105

Review 3.  CT angiography: injection and acquisition technique.

Authors:  Dominik Fleischmann
Journal:  Radiol Clin North Am       Date:  2010-03       Impact factor: 2.303

4.  Optimizing scan timing of hepatic arterial phase by physiologic pharmacokinetic analysis in bolus-tracking technique by multi-detector row computed tomography.

Authors:  Isao Yamaguchi; Eiji Kidoya; Masayuki Suzuki; Hirohiko Kimura
Journal:  Radiol Phys Technol       Date:  2010-09-25

5.  Usefulness of saline pushing in reduction of contrast material dose in abdominal CT: evaluation of time-density curve for the aorta, portal vein and liver.

Authors:  F Tatsugami; M Matsuki; Y Inada; G Nakai; M Tanikake; S Yoshikawa; I Narabayashi
Journal:  Br J Radiol       Date:  2006-10-12       Impact factor: 3.039

6.  What is the best contrast injection protocol for 64-row multi-detector cardiac computed tomography?

Authors:  Jin-guo Lu; Bin Lv; Xiong-biao Chen; Xiang Tang; Shi-liang Jiang; Ru-ping Dai
Journal:  Eur J Radiol       Date:  2009-05-21       Impact factor: 3.528

7.  Effect of different saline chaser volumes and flow rates on intravascular contrast enhancement in CT using a circulation phantom.

Authors:  Florian F Behrendt; Philipp Bruners; Sebastian Keil; Cedric Plumhans; Andreas H Mahnken; Marco Das; Diana Ackermann; Rolf W Günther; Georg Mühlenbruch
Journal:  Eur J Radiol       Date:  2009-04-05       Impact factor: 3.528

8.  Effects of injection rates of contrast material on arterial phase hepatic CT.

Authors:  T Kim; T Murakami; S Takahashi; K Tsuda; K Tomoda; Y Narumi; H Oi; H Nakamura
Journal:  AJR Am J Roentgenol       Date:  1998-08       Impact factor: 3.959

9.  Simulation of aortic peak enhancement on MDCT using a contrast material flow phantom: feasibility study.

Authors:  Kazuo Awai; Atsushi Hatcho; Yoshiharu Nakayama; Shinichiro Kusunoki; Duo Liu; Masahiro Hatemura; Yoshinori Funama; Masayuki Denbo; Natsuko Sato; Yasuyuki Yamashita
Journal:  AJR Am J Roentgenol       Date:  2006-02       Impact factor: 3.959

10.  Effect of saline flush on enhancement of proximal and distal segments using 320-row coronary CT angiography.

Authors:  Nobuo Tomizawa; Fumio Suzuki; Masaaki Akahane; Rumiko Torigoe; Shigeru Kiryu; Kuni Ohtomo
Journal:  Eur J Radiol       Date:  2013-04-08       Impact factor: 3.528

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.