The accuracy of PiCCO® in measuring cardiac output in patients under therapeutic hypothermia: Comparison with transthoracic echocardiography.

T Souto Moura1, S Aguiar Rosa2, N Germano3, R Cavaco3, T Sequeira3, M Alves4, A L Papoila5, L Bento3.   

Abstract

BACKGROUND: Invasive cardiac monitoring using thermodilution methods such as PiCCO® is widely used in critically ill patients and provides a wide range of hemodynamic variables, including cardiac output (CO). However, in post-cardiac arrest patients subjected to therapeutic hypothermia, the low body temperature possibly could interfere with the technique. Transthoracic Doppler echocardiography (ECHO) has long proved its accuracy in estimating CO, and is not influenced by temperature changes.
OBJECTIVE: To assess the accuracy of PiCCO® in measuring CO in patients under therapeutic hypothermia, compared with ECHO. DESIGN AND PATIENTS: Thirty paired COECHO/COPiCCO measurements were analyzed in 15 patients subjected to hypothermia after cardiac arrest. Eighteen paired measurements were obtained at under 36°C and 12 at ≥36°C. A value of 0.5l/min was considered the maximum accepted difference between the COECHO and COPiCCO values.
RESULTS: Under conditions of normothermia (≥36°C), the mean difference between COECHO and COPiCCO was 0.030 l/min, with limits of agreement (-0.22, 0.28) - all of the measurements differing by less than 0.5 l/min. In situations of hypothermia (<36°C), the mean difference in CO measurements was -0.426 l/min, with limits of agreement (-1.60, 0.75), and only 44% (8/18) of the paired measurements fell within the interval (-0.5, 0.5). The calculated temperature cut-off point maximizing specificity was 35.95°C: above this temperature, specificity was 100%, with a false-positive rate of 0%.
CONCLUSIONS: The results clearly show clinically relevant discordance between COECHO and COPiCCO at temperatures of <36°C, demonstrating the inaccuracy of PiCCO® for cardiac output measurements in hypothermic patients.
Copyright © 2017 Elsevier España, S.L.U. y SEMNIM. All rights reserved.

Entities:  

Keywords:  Cardiac output; Doppler echocardiography; Ecocardiografía doppler; Gasto cardiaco; Hemodynamic monitoring; Hipotermia; Hypothermia; Monitor PiCCO; Monitorización hemodinámica; PiCCO monitor

Mesh:

Year:  2017        PMID: 28552462     DOI: 10.1016/j.medin.2017.03.007

Source DB:  PubMed          Journal:  Med Intensiva (Engl Ed)        ISSN: 2173-5727


  6 in total

Review 1.  Accuracy of non-invasive and minimally invasive hemodynamic monitoring: where do we stand?

Authors:  Issa Pour-Ghaz; Theodore Manolukas; Nathalie Foray; Joel Raja; Aranyak Rawal; Uzoma N Ibebuogu; Rami N Khouzam
Journal:  Ann Transl Med       Date:  2019-09

Review 2.  Protocolized Post-Cardiac Arrest Care with Targeted Temperature Management.

Authors:  Wei-Ting Chen; Min-Shan Tsai; Chien-Hua Huang; Wei-Tien Chang; Wen-Jone Chen
Journal:  Acta Cardiol Sin       Date:  2022-05       Impact factor: 1.800

Review 3.  Rationale for using the velocity-time integral and the minute distance for assessing the stroke volume and cardiac output in point-of-care settings.

Authors:  Pablo Blanco
Journal:  Ultrasound J       Date:  2020-04-21

4.  Successful treatment of acute circulatory failure of unknown cause using critical ultrasound-guided reverse fluid resuscitation: A case report.

Authors:  Haotian Zhao; Ling Long; Zekai Wang; Yaru Yan; Heling Zhao
Journal:  Medicine (Baltimore)       Date:  2020-12-18       Impact factor: 1.817

5.  Cardiac output measurements via echocardiography versus thermodilution: A systematic review and meta-analysis.

Authors:  Yun Zhang; Yan Wang; Jing Shi; Zhiqiang Hua; Jinyu Xu
Journal:  PLoS One       Date:  2019-10-03       Impact factor: 3.240

6.  Comparison of cardiac output and cardiac index values measured by critical care echocardiography with the values measured by pulse index continuous cardiac output (PiCCO) in the pediatric intensive care unit:a preliminary study.

Authors:  Nagehan Aslan; Dincer Yildizdas; Ozden Ozgur Horoz; Yasemin Coban; Fadli Demir; Sevcan Erdem; Yasar Sertdemir
Journal:  Ital J Pediatr       Date:  2020-04-16       Impact factor: 2.638

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.