| Literature DB >> 28549338 |
Denis A Komarov1, Hiroshi Hirata2.
Abstract
In this paper, we introduce a procedure for the reconstruction of spectral-spatial EPR images using projections acquired with the constant sweep of a magnetic field. The application of a constant field-sweep and a predetermined data sampling rate simplifies the requirements for EPR imaging instrumentation and facilitates the backprojection-based reconstruction of spectral-spatial images. The proposed approach was applied to the reconstruction of a four-dimensional numerical phantom and to actual spectral-spatial EPR measurements. Image reconstruction using projections with a constant field-sweep was three times faster than the conventional approach with the application of a pseudo-angle and a scan range that depends on the applied field gradient. Spectral-spatial EPR imaging with a constant field-sweep for data acquisition only slightly reduces the signal-to-noise ratio or functional resolution of the resultant images and can be applied together with any common backprojection-based reconstruction algorithm.Keywords: 4D EPR imaging; Backprojection; Image reconstruction; Spectral-spatial EPR
Year: 2017 PMID: 28549338 DOI: 10.1016/j.jmr.2017.05.005
Source DB: PubMed Journal: J Magn Reson ISSN: 1090-7807 Impact factor: 2.229