Martina Ebner 1 , Siegfried Mariacher 1 , Kai Januschowski 1,2 , Katrin Boden 1 , Anna-Maria Seuthe 1 , Peter Szurman 1,2 , Karl Thomas Boden 1 . Show Affiliations »
Abstract
BACKGROUND: To evaluate intraocular pressure (IOP) using the application of a novel liquid patient interface for femtosecond laser-assisted cataract surgery with the FEMTO LDV Z8. METHODS: IOP was evaluated in enucleated porcine eyes prior, during and after the application of the Femto LDV Z8 liquid patient interface (Ziemer Ophthalmic Systems, Switzerland) using intracameral cannulation (n=20), intravitreal cannulation (n=20), rebound tonometry (n=20) and indentation tonometry (n=20). Pressure was assessed prior vacuum, during vacuum (30 s, 1 min, 2 min, 3 min) and after releasing the vacuum (1 min and 2 min). Two groups with different predefined vacuum levels (350 mbar, 420 mbar) were investigated. RESULTS: Mean intracameral pressure (±SD) increased during vacuum application from 20 mm Hg to 52.00 mm Hg (±6.35mm Hg; p=0.005) and 45.18 mmHg (±4.34 mm Hg; p=0.005) for the 420 mbar and the 350 mbar vacuum levels, respectively. Mean intravitreal pressure increased from 20 mm Hg to 25.60 mm Hg (±9.85 mm Hg; p=0.058) and 28.10 mm Hg (±2.54 mm Hg; p=0.059) for the 420 mbar and the 350 mbar vacuum levels, respectively. Pressure values from indentation and rebound tonometry were in between intracameral and intravitreal values. Mean intracameral IOP was 18.1% higher (p=0.019) in the 420 mbar group compared with the 350 mbar group. CONCLUSION: During vacuum application of the liquid patient interface of the Femto LDV Z8 for femtosecond laser-assisted cataract surgery, IOP values were higher in the anterior chamber compared with the intravitreal pressure measurements. The higher predefined vacuum level (350 mbar vs 420 mbar) resulted in significant higher intracameral IOP. © Article author(s) (or their employer(s) unless otherwise stated in the text of the article) 2017. All rights reserved. No commercial use is permitted unless otherwise expressly granted.
BACKGROUND: To evaluate intraocular pressure (IOP) using the application of a novel liquid patient interface for femtosecond laser-assisted cataract surgery with the FEMTO LDV Z8. METHODS: IOP was evaluated in enucleated porcine eyes prior, during and after the application of the Femto LDV Z8 liquid patient interface (Ziemer Ophthalmic Systems, Switzerland) using intracameral cannulation (n=20), intravitreal cannulation (n=20), rebound tonometry (n=20) and indentation tonometry (n=20). Pressure was assessed prior vacuum, during vacuum (30 s, 1 min, 2 min, 3 min) and after releasing the vacuum (1 min and 2 min). Two groups with different predefined vacuum levels (350 mbar, 420 mbar) were investigated. RESULTS: Mean intracameral pressure (±SD) increased during vacuum application from 20 mm Hg to 52.00 mm Hg (±6.35mm Hg; p=0.005) and 45.18 mmHg (±4.34 mm Hg; p=0.005) for the 420 mbar and the 350 mbar vacuum levels, respectively. Mean intravitreal pressure increased from 20 mm Hg to 25.60 mm Hg (±9.85 mm Hg; p=0.058) and 28.10 mm Hg (±2.54 mm Hg; p=0.059) for the 420 mbar and the 350 mbar vacuum levels, respectively. Pressure values from indentation and rebound tonometry were in between intracameral and intravitreal values. Mean intracameral IOP was 18.1% higher (p=0.019) in the 420 mbar group compared with the 350 mbar group. CONCLUSION: During vacuum application of the liquid patient interface of the Femto LDV Z8 for femtosecond laser-assisted cataract surgery, IOP values were higher in the anterior chamber compared with the intravitreal pressure measurements. The higher predefined vacuum level (350 mbar vs 420 mbar) resulted in significant higher intracameral IOP. © Article author(s) (or their employer(s) unless otherwise stated in the text of the article) 2017. All rights reserved. No commercial use is permitted unless otherwise expressly granted.
Entities: Disease
Species
Keywords:
experimental – laboratory; intraocular pressure; lens and zonules; treatment lasers
Mesh: See more »
Substances: See more »
Year: 2017
PMID: 28546151 DOI: 10.1136/bjophthalmol-2016-309890
Source DB: PubMed Journal: Br J Ophthalmol ISSN: 0007-1161 Impact factor: 4.638