Patient doses and occupational exposure in a hybrid operating room.

C Andrés1, H Pérez-García2, M Agulla2, R Torres2, D Miguel2, A Del Castillo2, C M Flota3, D Alonso2, J de Frutos2, C Vaquero3.   

Abstract

PURPOSE: This study aimed to characterize the radiation exposure to patients and workers in a new vascular hybrid operating room during X-ray-guided procedures.
METHODS: During one year, data from 260 interventions performed in a hybrid operating room equipped with a Siemens Artis Zeego angiography system were monitored. The patient doses were analysed using the following parameters: radiation time, kerma-area product, patient entrance reference point dose and peak skin dose. Staff radiation exposure and ambient dose equivalent were also measured using direct reading dosimeters and thermoluminescent dosimeters.
RESULTS: The radiation time, kerma-area product, patient entrance reference point dose and peak skin dose were, on average, 19:15min, 67Gy·cm2, 0.41Gy and 0.23Gy, respectively. Although the contribution of the acquisition mode was smaller than 5% in terms of the radiation time, this mode accounted for more than 60% of the effective dose per patient. All of the worker dose measurements remained below the limits established by law.
CONCLUSIONS: The working conditions in the hybrid operating room HOR are safe in terms of patient and staff radiation protection. Nevertheless, doses are highly dependent on the workload; thus, further research is necessary to evaluate any possible radiological deviation of the daily working conditions in the HOR.
Copyright © 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Artis Zeego; Hybrid operating room; Occupational exposure; Patient dose

Mesh:

Year:  2017        PMID: 28535913     DOI: 10.1016/j.ejmp.2017.04.006

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  5 in total

Review 1.  [Hybrid operating theater].

Authors:  T Fuchs-Buder; N Settembre; D Schmartz
Journal:  Anaesthesist       Date:  2018-07       Impact factor: 1.041

2.  Evaluation of peak skin dose during percutaneous coronary intervention procedures: relationship with fluoroscopic pulse rate and target vessel.

Authors:  Takuro Tanaka; Kosuke Matsubara; Satoshi Kobayashi
Journal:  Radiol Phys Technol       Date:  2021-01-05

3.  Curved periacetabular osteotomy using intraoperative real-time 3-dimensional computed tomography with a robotic C-arm system: A case report.

Authors:  Ariha Goshi; Shigeo Fukunishi; Shohei Okahisa; Taishi Okada; Shinichi Yoshiya
Journal:  Medicine (Baltimore)       Date:  2018-11       Impact factor: 1.817

4.  Team composition and staff roles in a hybrid operating room: A prospective study using video observations.

Authors:  May Bazzi; Ingegerd Bergbom; Mikael Hellström; Isabell Fridh; Karin Ahlberg; Solveig M Lundgren
Journal:  Nurs Open       Date:  2019-07-03

5.  Intraoperative modulation of arterial blood flow in a hybrid operating room: A report of three cases.

Authors:  Yoshikatsu Saitoh; Yasuyuki Hara; Shigehito Miyagi; Chikashi Nakanishi; Wataru Nakanishi; Ryuichi Nishimura; Daijirou Akamatsu; Hitoshi Goto; Michiaki Unno; Takashi Kamei
Journal:  Clin Case Rep       Date:  2019-08-15
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.