Toric outcomes: Computer-assisted registration versus intraoperative aberrometry.

Jonathan D Solomon1, John Ladas2.   

Abstract

PURPOSE: To compare refractive outcomes of intraoperative computer-assisted registration and intraoperative aberrometry for the reduction of cylinder during toric intraocular lens (IOL) placement.
SETTING: Bowie Vision Institute, Bowie, Maryland, USA.
DESIGN: Prospective randomized case series.
METHOD: The patients were divided into 2 groups that had toric IOL implantation after phacoemulsification. The intraoperative computer-assisted registration group (Group 1) had preoperative toric calculations. The aberrometry group (Group 2) was guided by a vergence formula and intraoperative pseudophakic cylindrical measurements to determine the final IOL power and intended orientation. The primary outcome measure was the mean postoperative remaining refractive astigmatism, and it was compared with the predicted amount of cylindrical correction with the IOL.
RESULTS: Fifty-two patients (104 eyes) had sequential cataract surgery. The mean amount of cylinder correction was 1.60 diopters (D) ± 0.70 (SD) (range 0.75 to 3.08 D) in Group 1 and 1.74 ± 0.79 D (range 0.72 to 3.08 D) in Group 2. The mean remaining refractive astigmatism was -0.29 ± 0.22 D in Group 1 and -0.46 ± 0.25 D in Group 2 (P = .0003). A difference vector of 0.1 @ 87 degrees (0.31 D arithmetic mean) was calculated in Group 1 and 0.0 @ 82 degrees (0.44 D arithmetic mean) in Group 2. The correction index was 1.03 in Group 1 and 0.95 in Group 2.
CONCLUSION: Intraoperative markerless computer-assisted registration and biometric guidance summarily yielded less remaining refractive cylinder than toric IOL placement guided by intraoperative aberrometry.
Copyright © 2017 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Year:  2017        PMID: 28532935     DOI: 10.1016/j.jcrs.2017.01.012

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  7 in total

1.  Predictive accuracy of an intraoperative aberrometry device for a new monofocal intraocular lens.

Authors:  Lindsay S Spekreijse; Noel J C Bauer; Frank J H M van den Biggelaar; Rob W P Simons; Claudette A Veldhuizen; Tos T J M Berendschot; Rudy M M A Nuijts
Journal:  J Cataract Refract Surg       Date:  2021-08-19       Impact factor: 3.528

2.  Clinical outcomes after aphakic versus aphakic/pseudophakic intraoperative aberrometry in cataract surgery with toric IOL implantation.

Authors:  Satish S Modi
Journal:  Int Ophthalmol       Date:  2020-08-08       Impact factor: 2.031

Review 3.  Optimizing outcomes with toric intraocular lenses.

Authors:  Manpreet Kaur; Farin Shaikh; Ruchita Falera; Jeewan S Titiyal
Journal:  Indian J Ophthalmol       Date:  2017-12       Impact factor: 1.848

4.  Results of Callisto Eye System in Toric Intraocular Lens Alignment.

Authors:  Bulent Kose; Sinan Albayrak
Journal:  Beyoglu Eye J       Date:  2020-02-14

5.  Comparison of Toric Intraocular Lens Alignment Using Image-Guided System and Manual Marking Technique.

Authors:  Bulent Kose; Hakika Erdogan
Journal:  Beyoglu Eye J       Date:  2020-07-29

6.  Comparison of Preoperative Measurements with Intraoperative Aberrometry in Predicting Need for Correction in Eyes with Low Astigmatism Undergoing Cataract Surgery.

Authors:  John A Hovanesian
Journal:  Clin Ophthalmol       Date:  2021-05-26

7.  Clinical evaluation of toric intraocular lens implantation based on iTrace wavefront keratometric astigmatism.

Authors:  Zhe Zhang; Hui Li; Jing Zhou; Yaqin Zhang; Suhua Zhang
Journal:  BMC Ophthalmol       Date:  2020-11-16       Impact factor: 2.209

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.