Fig. 1Multichannel vaginal cylinder with 35 mm in diameter and Syed-Neblett obturator with 22 mm
Fig. 2Translational shift of 6.5 mm to the bladder (yellow and orange) anteriorly and rectum (brown and green) posteriorly, with no tissue compression
Target characteristics of the 6 patients
| CC Length (mm) | Circumferential shape | FL above cylinder tip (mm) | Max. Thickness (mm) | Max. Thickness Direction | Min. Bladder distance (mm) | Min. Rectum distance (mm) |
|---|---|---|---|---|---|---|
| 42 | Circumferential | 5 | 12,4 | Posterior-left | 2 | Abutting |
| 68 | Circumferential | 5 | 17 | Posterior-right | 2 | Abutting |
| 50 | Circumferential | 10 | 20 | Posterior-left | Abutting | Abutting |
| 70 | Circumferential | 0 | 12 | left | Abutting | Abutting |
| 55 | Circumferential | 0 | 10 | Anterior-right | Abutting | Abutting |
| 30 | 1 to 7 h | 7 | 12,8 | left | 2 | Abutting |
| 67 | 6 to 1 h | 7 | 19 | right | 3 | Abutting |
| 24 | 3 to 8 h | 6 | 15 | posterior | 8 | Abutting |
| 31 | 7 to 1 h | 5 | 14 | right | Abutting | 9,4 |
| 50 | Circumferential | 2 | 11 | Posterior-left | Abutting | Abutting |
FL Free length
Dosimetric comparison between ISBT and MCVC plans in HR-CTV and IR-CTV
| HRCTV (cGy) |
| IRCTV (cGy) |
| |||
|---|---|---|---|---|---|---|
| ISBT | MCVC | ISBT | MCVC | |||
| Bladder D2cc | 338 | 556 | 0.0008 | 413 | 537 | 0.0001 |
| Rectum D2cc | 336 | 468 | 0.023 | 434 | 520 | 0.006 |
| Vagina mucosa D0.5 cc | 699 | 1134 | 0.003 | 1850 | 1534 | 0.09 |
| Vagina mucosa D1cc | 543 | 1009 | 0.0007 | 1448 | 1343 | 0.13 |
| Vagina mucosa D2cc | 383 | 864 | 0.0002 | 1161 | 1178 | 0.73 |
Dose relation between vaginal mucosa thickness and BT technique in IRCTV targets
| Vaginal Mucosa | IR-CTV | ||||||
|---|---|---|---|---|---|---|---|
| ISBT (cGy) | MCVC (cGy) | ||||||
| Thickness | D0.5 cc | D1cc | D2cc | D0.5 cc | D1cc | D2cc |
|
| 2 mm | 1767 | 1352 | 1061 | 1494 | 1289 | 1135 | NS |
| 3 mm | 1850 | 1448 | 1161 | 1534 | 1343 | 1178 | NS |
| 4 mm | 1915* | 1501 | 1212 | 1537* | 1370 | 1208 | NS |
*p = 0.06
Exploratory analysis of vaginal mucosa dosimetry in ISBT, and MCVC plans after mucosa thinning
| Comparison groups | Vaginal parameter | Dose |
| |
|---|---|---|---|---|
| ISBT (cGy) | MCVC (cGy) | |||
| MCVC 2 mm vs ISBT 3 mm | D0.5 cc | 1850 | 1494 | 0.04 |
| D1cc | 1448 | 1289 | NS | |
| D2cc | 1161 | 1135 | NS | |
| MCVC 3 mm vs ISBT 4 mm | D0.5 cc | 1915 | 1534 | 0.03 |
| D1cc | 1501 | 1343 | NS | |
| D2cc | 1212 | 1178 | NS | |